Cargando…
Ventilation distribution in rats: Part 2 – A comparison of electrical impedance tomography and hyperpolarised helium magnetic resonance imaging
BACKGROUND: Hyperpolarised helium MRI (He3 MRI) is a new technique that enables imaging of the air distribution within the lungs. This allows accurate determination of the ventilation distribution in vivo. The technique has the disadvantages of requiring an expensive helium isotope, complex apparatu...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3497860/ https://www.ncbi.nlm.nih.gov/pubmed/22966835 http://dx.doi.org/10.1186/1475-925X-11-68 |
_version_ | 1782249766289145856 |
---|---|
author | Dunster, Kimble R Friese, Marlies EJ Fraser, John F Galloway, Graham J Cowin, Gary J Schibler, Andreas |
author_facet | Dunster, Kimble R Friese, Marlies EJ Fraser, John F Galloway, Graham J Cowin, Gary J Schibler, Andreas |
author_sort | Dunster, Kimble R |
collection | PubMed |
description | BACKGROUND: Hyperpolarised helium MRI (He3 MRI) is a new technique that enables imaging of the air distribution within the lungs. This allows accurate determination of the ventilation distribution in vivo. The technique has the disadvantages of requiring an expensive helium isotope, complex apparatus and moving the patient to a compatible MRI scanner. Electrical impedance tomography (EIT) a non-invasive bedside technique that allows constant monitoring of lung impedance, which is dependent on changes in air space capacity in the lung. We have used He3MRI measurements of ventilation distribution as the gold standard for assessment of EIT. METHODS: Seven rats were ventilated in supine, prone, left and right lateral position with 70% helium/30% oxygen for EIT measurements and pure helium for He3 MRI. The same ventilator and settings were used for both measurements. Image dimensions, geometric centre and global in homogeneity index were calculated. RESULTS: EIT images were smaller and of lower resolution and contained less anatomical detail than those from He3 MRI. However, both methods could measure positional induced changes in lung ventilation, as assessed by the geometric centre. The global in homogeneity index were comparable between the techniques. CONCLUSION: EIT is a suitable technique for monitoring ventilation distribution and inhomgeneity as assessed by comparison with He3 MRI. |
format | Online Article Text |
id | pubmed-3497860 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-34978602012-11-15 Ventilation distribution in rats: Part 2 – A comparison of electrical impedance tomography and hyperpolarised helium magnetic resonance imaging Dunster, Kimble R Friese, Marlies EJ Fraser, John F Galloway, Graham J Cowin, Gary J Schibler, Andreas Biomed Eng Online Research BACKGROUND: Hyperpolarised helium MRI (He3 MRI) is a new technique that enables imaging of the air distribution within the lungs. This allows accurate determination of the ventilation distribution in vivo. The technique has the disadvantages of requiring an expensive helium isotope, complex apparatus and moving the patient to a compatible MRI scanner. Electrical impedance tomography (EIT) a non-invasive bedside technique that allows constant monitoring of lung impedance, which is dependent on changes in air space capacity in the lung. We have used He3MRI measurements of ventilation distribution as the gold standard for assessment of EIT. METHODS: Seven rats were ventilated in supine, prone, left and right lateral position with 70% helium/30% oxygen for EIT measurements and pure helium for He3 MRI. The same ventilator and settings were used for both measurements. Image dimensions, geometric centre and global in homogeneity index were calculated. RESULTS: EIT images were smaller and of lower resolution and contained less anatomical detail than those from He3 MRI. However, both methods could measure positional induced changes in lung ventilation, as assessed by the geometric centre. The global in homogeneity index were comparable between the techniques. CONCLUSION: EIT is a suitable technique for monitoring ventilation distribution and inhomgeneity as assessed by comparison with He3 MRI. BioMed Central 2012-09-11 /pmc/articles/PMC3497860/ /pubmed/22966835 http://dx.doi.org/10.1186/1475-925X-11-68 Text en Copyright ©2012 Dunster et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Dunster, Kimble R Friese, Marlies EJ Fraser, John F Galloway, Graham J Cowin, Gary J Schibler, Andreas Ventilation distribution in rats: Part 2 – A comparison of electrical impedance tomography and hyperpolarised helium magnetic resonance imaging |
title | Ventilation distribution in rats: Part 2 – A comparison of electrical impedance tomography and hyperpolarised helium magnetic resonance imaging |
title_full | Ventilation distribution in rats: Part 2 – A comparison of electrical impedance tomography and hyperpolarised helium magnetic resonance imaging |
title_fullStr | Ventilation distribution in rats: Part 2 – A comparison of electrical impedance tomography and hyperpolarised helium magnetic resonance imaging |
title_full_unstemmed | Ventilation distribution in rats: Part 2 – A comparison of electrical impedance tomography and hyperpolarised helium magnetic resonance imaging |
title_short | Ventilation distribution in rats: Part 2 – A comparison of electrical impedance tomography and hyperpolarised helium magnetic resonance imaging |
title_sort | ventilation distribution in rats: part 2 – a comparison of electrical impedance tomography and hyperpolarised helium magnetic resonance imaging |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3497860/ https://www.ncbi.nlm.nih.gov/pubmed/22966835 http://dx.doi.org/10.1186/1475-925X-11-68 |
work_keys_str_mv | AT dunsterkimbler ventilationdistributioninratspart2acomparisonofelectricalimpedancetomographyandhyperpolarisedheliummagneticresonanceimaging AT friesemarliesej ventilationdistributioninratspart2acomparisonofelectricalimpedancetomographyandhyperpolarisedheliummagneticresonanceimaging AT fraserjohnf ventilationdistributioninratspart2acomparisonofelectricalimpedancetomographyandhyperpolarisedheliummagneticresonanceimaging AT gallowaygrahamj ventilationdistributioninratspart2acomparisonofelectricalimpedancetomographyandhyperpolarisedheliummagneticresonanceimaging AT cowingaryj ventilationdistributioninratspart2acomparisonofelectricalimpedancetomographyandhyperpolarisedheliummagneticresonanceimaging AT schiblerandreas ventilationdistributioninratspart2acomparisonofelectricalimpedancetomographyandhyperpolarisedheliummagneticresonanceimaging |