The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study

BACKGROUND: The broad use of force plates in clinical settings for postural control assessment suggests the need for instruments that are easy to use, affordable and readily available. In addition, these instruments of measurement should be reliable and valid as adequate reliability and validity are...

Descripción completa

Detalles Bibliográficos
Autores principales: Golriz, Samira, Hebert, Jeffrey J, Foreman, K Bo, Walker, Bruce F
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3502125/
https://www.ncbi.nlm.nih.gov/pubmed/22620857
http://dx.doi.org/10.1186/2045-709X-20-15
_version_ 1782250271010717696
author Golriz, Samira
Hebert, Jeffrey J
Foreman, K Bo
Walker, Bruce F
author_facet Golriz, Samira
Hebert, Jeffrey J
Foreman, K Bo
Walker, Bruce F
author_sort Golriz, Samira
collection PubMed
description BACKGROUND: The broad use of force plates in clinical settings for postural control assessment suggests the need for instruments that are easy to use, affordable and readily available. In addition, these instruments of measurement should be reliable and valid as adequate reliability and validity are prerequisites to making correct inferences. The aim of this study was to examine the concurrent validity of postural control measures obtained with a clinical force plate. METHODS: Thirty-one healthy adults were recruited. Participants completed 1 set of 5 trials on each force plate. Postural control measures (centre of pressure [COP] average velocity and sway area) were collected and compared using the Midot Posture Scale Analyzer (clinical force plate) and the Accugait force plate (criterion measure). Intra class correlation coefficient (ICC), standard error of measurement , and paired t-tests were calculated and Bland-Altman plots were constructed to compare the force plates and assess consistency of measurement and agreement between them. RESULTS: The ICC values (ICC = 0.14-0.60) between the two force plates were lower than the acceptable value for both COP average velocity and sway area. There was significant difference (p > 0.05) in COP average velocity and sway area between the force plates. Examination of the plots revealed that there is less difference between the force plates in lower magnitudes of COP for average velocity and sway area however, the greater the average velocity and sway area, the greater the difference between the measures obtained from the two force plates. CONCLUSION: Findings of this study showed poor concurrent validity of the clinical force plate. This clinical force plate cannot be a replacement for known reliable and valid force plates and consequently measures obtained from this force plate should be treated with caution especially in a clinical population.
format Online
Article
Text
id pubmed-3502125
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-35021252012-11-21 The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study Golriz, Samira Hebert, Jeffrey J Foreman, K Bo Walker, Bruce F Chiropr Man Therap Research BACKGROUND: The broad use of force plates in clinical settings for postural control assessment suggests the need for instruments that are easy to use, affordable and readily available. In addition, these instruments of measurement should be reliable and valid as adequate reliability and validity are prerequisites to making correct inferences. The aim of this study was to examine the concurrent validity of postural control measures obtained with a clinical force plate. METHODS: Thirty-one healthy adults were recruited. Participants completed 1 set of 5 trials on each force plate. Postural control measures (centre of pressure [COP] average velocity and sway area) were collected and compared using the Midot Posture Scale Analyzer (clinical force plate) and the Accugait force plate (criterion measure). Intra class correlation coefficient (ICC), standard error of measurement , and paired t-tests were calculated and Bland-Altman plots were constructed to compare the force plates and assess consistency of measurement and agreement between them. RESULTS: The ICC values (ICC = 0.14-0.60) between the two force plates were lower than the acceptable value for both COP average velocity and sway area. There was significant difference (p > 0.05) in COP average velocity and sway area between the force plates. Examination of the plots revealed that there is less difference between the force plates in lower magnitudes of COP for average velocity and sway area however, the greater the average velocity and sway area, the greater the difference between the measures obtained from the two force plates. CONCLUSION: Findings of this study showed poor concurrent validity of the clinical force plate. This clinical force plate cannot be a replacement for known reliable and valid force plates and consequently measures obtained from this force plate should be treated with caution especially in a clinical population. BioMed Central 2012-05-23 /pmc/articles/PMC3502125/ /pubmed/22620857 http://dx.doi.org/10.1186/2045-709X-20-15 Text en Copyright ©2012 Golriz et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Golriz, Samira
Hebert, Jeffrey J
Foreman, K Bo
Walker, Bruce F
The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study
title The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study
title_full The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study
title_fullStr The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study
title_full_unstemmed The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study
title_short The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study
title_sort validity of a portable clinical force plate in assessment of static postural control: concurrent validity study
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3502125/
https://www.ncbi.nlm.nih.gov/pubmed/22620857
http://dx.doi.org/10.1186/2045-709X-20-15
work_keys_str_mv AT golrizsamira thevalidityofaportableclinicalforceplateinassessmentofstaticposturalcontrolconcurrentvaliditystudy
AT hebertjeffreyj thevalidityofaportableclinicalforceplateinassessmentofstaticposturalcontrolconcurrentvaliditystudy
AT foremankbo thevalidityofaportableclinicalforceplateinassessmentofstaticposturalcontrolconcurrentvaliditystudy
AT walkerbrucef thevalidityofaportableclinicalforceplateinassessmentofstaticposturalcontrolconcurrentvaliditystudy
AT golrizsamira validityofaportableclinicalforceplateinassessmentofstaticposturalcontrolconcurrentvaliditystudy
AT hebertjeffreyj validityofaportableclinicalforceplateinassessmentofstaticposturalcontrolconcurrentvaliditystudy
AT foremankbo validityofaportableclinicalforceplateinassessmentofstaticposturalcontrolconcurrentvaliditystudy
AT walkerbrucef validityofaportableclinicalforceplateinassessmentofstaticposturalcontrolconcurrentvaliditystudy