Acoustic system for the estimation of the temporary blood chamber volume of the POLVAD heart supporting prosthesis

BACKGROUND: The paper presents a newly researched acoustic system for blood volume measurements for the developed family of Polish ventricular assist devices. The pneumatic heart-supporting devices are still the preferred solution in some cases, and monitoring of their operation, especially the temp...

Descripción completa

Detalles Bibliográficos
Autores principales: Konieczny, Grzegorz, Opilski, Zbigniew, Pustelny, Tadeusz, Gawlikowski, Maciej
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3502557/
https://www.ncbi.nlm.nih.gov/pubmed/22998766
http://dx.doi.org/10.1186/1475-925X-11-72
_version_ 1782250366711103488
author Konieczny, Grzegorz
Opilski, Zbigniew
Pustelny, Tadeusz
Gawlikowski, Maciej
author_facet Konieczny, Grzegorz
Opilski, Zbigniew
Pustelny, Tadeusz
Gawlikowski, Maciej
author_sort Konieczny, Grzegorz
collection PubMed
description BACKGROUND: The paper presents a newly researched acoustic system for blood volume measurements for the developed family of Polish ventricular assist devices. The pneumatic heart-supporting devices are still the preferred solution in some cases, and monitoring of their operation, especially the temporary blood volume, is yet to be solved. METHODS: The prototype of the POLVAD-EXT prosthesis developed by the Foundation of Cardiac Surgery Development, Zabrze, Poland, is equipped with the newly researched acoustic blood volume measurement system based on the principle of Helmholtz’s acoustic resonance. The results of static volume measurements acquired using the acoustic sensor were verified by measuring the volume of the liquid filling the prosthesis. Dynamic measurements were conducted on the hybrid model of the human cardiovascular system at the Foundation, with the Transonic T410 (11PLX transducer - 5% uncertainty) ultrasound flow rate sensor, used as the reference. RESULTS: The statistical analysis of a series of static tests have proved that the sensor solution provides blood volume measurement results with uncertainties (understood as a standard mean deviation) of less than 10%. Dynamic tests show a high correlation between the results of the acoustic system and those obtained by flow rate measurements using an ultrasound transit time type sensor. CONCLUSIONS: The results show that noninvasive, online temporary blood volume measurements in the POLVAD-EXT prosthesis, making use of the newly developed acoustic system, provides accurate static and dynamic measurements results. Conducted research provides the preliminary view on the possibility of reducing the additional sensor chamber volume in future.
format Online
Article
Text
id pubmed-3502557
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-35025572012-11-27 Acoustic system for the estimation of the temporary blood chamber volume of the POLVAD heart supporting prosthesis Konieczny, Grzegorz Opilski, Zbigniew Pustelny, Tadeusz Gawlikowski, Maciej Biomed Eng Online Research BACKGROUND: The paper presents a newly researched acoustic system for blood volume measurements for the developed family of Polish ventricular assist devices. The pneumatic heart-supporting devices are still the preferred solution in some cases, and monitoring of their operation, especially the temporary blood volume, is yet to be solved. METHODS: The prototype of the POLVAD-EXT prosthesis developed by the Foundation of Cardiac Surgery Development, Zabrze, Poland, is equipped with the newly researched acoustic blood volume measurement system based on the principle of Helmholtz’s acoustic resonance. The results of static volume measurements acquired using the acoustic sensor were verified by measuring the volume of the liquid filling the prosthesis. Dynamic measurements were conducted on the hybrid model of the human cardiovascular system at the Foundation, with the Transonic T410 (11PLX transducer - 5% uncertainty) ultrasound flow rate sensor, used as the reference. RESULTS: The statistical analysis of a series of static tests have proved that the sensor solution provides blood volume measurement results with uncertainties (understood as a standard mean deviation) of less than 10%. Dynamic tests show a high correlation between the results of the acoustic system and those obtained by flow rate measurements using an ultrasound transit time type sensor. CONCLUSIONS: The results show that noninvasive, online temporary blood volume measurements in the POLVAD-EXT prosthesis, making use of the newly developed acoustic system, provides accurate static and dynamic measurements results. Conducted research provides the preliminary view on the possibility of reducing the additional sensor chamber volume in future. BioMed Central 2012-09-21 /pmc/articles/PMC3502557/ /pubmed/22998766 http://dx.doi.org/10.1186/1475-925X-11-72 Text en Copyright ©2012 Konieczny et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Konieczny, Grzegorz
Opilski, Zbigniew
Pustelny, Tadeusz
Gawlikowski, Maciej
Acoustic system for the estimation of the temporary blood chamber volume of the POLVAD heart supporting prosthesis
title Acoustic system for the estimation of the temporary blood chamber volume of the POLVAD heart supporting prosthesis
title_full Acoustic system for the estimation of the temporary blood chamber volume of the POLVAD heart supporting prosthesis
title_fullStr Acoustic system for the estimation of the temporary blood chamber volume of the POLVAD heart supporting prosthesis
title_full_unstemmed Acoustic system for the estimation of the temporary blood chamber volume of the POLVAD heart supporting prosthesis
title_short Acoustic system for the estimation of the temporary blood chamber volume of the POLVAD heart supporting prosthesis
title_sort acoustic system for the estimation of the temporary blood chamber volume of the polvad heart supporting prosthesis
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3502557/
https://www.ncbi.nlm.nih.gov/pubmed/22998766
http://dx.doi.org/10.1186/1475-925X-11-72
work_keys_str_mv AT koniecznygrzegorz acousticsystemfortheestimationofthetemporarybloodchambervolumeofthepolvadheartsupportingprosthesis
AT opilskizbigniew acousticsystemfortheestimationofthetemporarybloodchambervolumeofthepolvadheartsupportingprosthesis
AT pustelnytadeusz acousticsystemfortheestimationofthetemporarybloodchambervolumeofthepolvadheartsupportingprosthesis
AT gawlikowskimaciej acousticsystemfortheestimationofthetemporarybloodchambervolumeofthepolvadheartsupportingprosthesis