Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation

We report a case of severe hemolysis and pulmonary valve insufficiency (PI) following right ventricular support using a paracorporeal pneumatic pump (Abiomed, Danvers, MA, USA). We speculate that the high velocity jet of blood emanating from the outflow cannula caused turbulence above the pulmonary...

Descripción completa

Detalles Bibliográficos
Autores principales: Schubert, Sarah A., Soleimani, Behzad, Pae, Walter E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3505954/
https://www.ncbi.nlm.nih.gov/pubmed/23213613
http://dx.doi.org/10.1155/2012/376384
_version_ 1782250834739855360
author Schubert, Sarah A.
Soleimani, Behzad
Pae, Walter E.
author_facet Schubert, Sarah A.
Soleimani, Behzad
Pae, Walter E.
author_sort Schubert, Sarah A.
collection PubMed
description We report a case of severe hemolysis and pulmonary valve insufficiency (PI) following right ventricular support using a paracorporeal pneumatic pump (Abiomed, Danvers, MA, USA). We speculate that the high velocity jet of blood emanating from the outflow cannula caused turbulence above the pulmonary valve, leading to PI and hemolysis. Despite the growing number of implanted ventricular assist devices, we could find no report in the literature describing pulmonary valve insufficiency secondary to right ventricular assist device (RVAD) placement. Fortunately, in this case, right ventricular function recovered sufficiently after seven days of support, allowing explantation of the device and resolution of PI and hemolysis.
format Online
Article
Text
id pubmed-3505954
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-35059542012-12-04 Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation Schubert, Sarah A. Soleimani, Behzad Pae, Walter E. Case Rep Transplant Case Report We report a case of severe hemolysis and pulmonary valve insufficiency (PI) following right ventricular support using a paracorporeal pneumatic pump (Abiomed, Danvers, MA, USA). We speculate that the high velocity jet of blood emanating from the outflow cannula caused turbulence above the pulmonary valve, leading to PI and hemolysis. Despite the growing number of implanted ventricular assist devices, we could find no report in the literature describing pulmonary valve insufficiency secondary to right ventricular assist device (RVAD) placement. Fortunately, in this case, right ventricular function recovered sufficiently after seven days of support, allowing explantation of the device and resolution of PI and hemolysis. Hindawi Publishing Corporation 2012 2012-09-26 /pmc/articles/PMC3505954/ /pubmed/23213613 http://dx.doi.org/10.1155/2012/376384 Text en Copyright © 2012 Sarah A. Schubert et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Schubert, Sarah A.
Soleimani, Behzad
Pae, Walter E.
Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation
title Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation
title_full Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation
title_fullStr Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation
title_full_unstemmed Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation
title_short Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation
title_sort hemolysis and pulmonary insufficiency following right ventricular assist device implantation
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3505954/
https://www.ncbi.nlm.nih.gov/pubmed/23213613
http://dx.doi.org/10.1155/2012/376384
work_keys_str_mv AT schubertsaraha hemolysisandpulmonaryinsufficiencyfollowingrightventricularassistdeviceimplantation
AT soleimanibehzad hemolysisandpulmonaryinsufficiencyfollowingrightventricularassistdeviceimplantation
AT paewaltere hemolysisandpulmonaryinsufficiencyfollowingrightventricularassistdeviceimplantation