Cargando…

Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models

PURPOSE: The aim of this study was to investigate the biomechanical stability provided by a novel, polylactic acid/nano-sized, β-tricalcium phosphate, bioabsorbable, self-retaining cervical fusion cage (BCFC). METHODS: Quasistatic nonconstraining torques (maximum 1.5 NM) induced flexion, extension,...

Descripción completa

Detalles Bibliográficos
Autores principales: Cao, Lu, Duan, Ping-Guo, Li, Xi-Lei, Yuan, Feng-Lai, Zhao, Ming-Dong, Che, Wu, Wang, Hui-Ren, Dong, Jian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3512541/
https://www.ncbi.nlm.nih.gov/pubmed/23226018
http://dx.doi.org/10.2147/IJN.S38288
_version_ 1782251749957959680
author Cao, Lu
Duan, Ping-Guo
Li, Xi-Lei
Yuan, Feng-Lai
Zhao, Ming-Dong
Che, Wu
Wang, Hui-Ren
Dong, Jian
author_facet Cao, Lu
Duan, Ping-Guo
Li, Xi-Lei
Yuan, Feng-Lai
Zhao, Ming-Dong
Che, Wu
Wang, Hui-Ren
Dong, Jian
author_sort Cao, Lu
collection PubMed
description PURPOSE: The aim of this study was to investigate the biomechanical stability provided by a novel, polylactic acid/nano-sized, β-tricalcium phosphate, bioabsorbable, self-retaining cervical fusion cage (BCFC). METHODS: Quasistatic nonconstraining torques (maximum 1.5 NM) induced flexion, extension, lateral bending (±1.5 NM), and axial rotation (±1.5 NM) on 32 sheep cervical spines (C2–C5). The motion segment C3–C4 was first tested intact; the following groups were tested after complete discectomy: autologous tricortical iliac crest bone graft, Medtronic–Wego polyetheretherketone (PEEK) cage, Solis PEEK cage, and BCFC. The autologous bone graft group was tested with an anterior plate. The mean range of motion (ROM) was calculated from the load-displacement curves. RESULTS: BCFC significantly decreased ROM in lateral bending and axial rotation compared to other implants, and no significant difference in ROM between two types of PEEK cages and BCFC could be observed in flexion and extension. Anterior cervical plate (ACP) significantly decreased ROM in flexion and extension, but no significant difference in ROM between BCFC and bone graft plus ACP could be determined in lateral bending and axial rotation. CONCLUSION: The BCFC device showed better stability to autologous tricortical iliac crest bone graft and PEEK cages in single-level anterior cervical discectomy and fusion models and thus may be a potential alternative to the current PEEK cages.
format Online
Article
Text
id pubmed-3512541
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-35125412012-12-05 Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models Cao, Lu Duan, Ping-Guo Li, Xi-Lei Yuan, Feng-Lai Zhao, Ming-Dong Che, Wu Wang, Hui-Ren Dong, Jian Int J Nanomedicine Original Research PURPOSE: The aim of this study was to investigate the biomechanical stability provided by a novel, polylactic acid/nano-sized, β-tricalcium phosphate, bioabsorbable, self-retaining cervical fusion cage (BCFC). METHODS: Quasistatic nonconstraining torques (maximum 1.5 NM) induced flexion, extension, lateral bending (±1.5 NM), and axial rotation (±1.5 NM) on 32 sheep cervical spines (C2–C5). The motion segment C3–C4 was first tested intact; the following groups were tested after complete discectomy: autologous tricortical iliac crest bone graft, Medtronic–Wego polyetheretherketone (PEEK) cage, Solis PEEK cage, and BCFC. The autologous bone graft group was tested with an anterior plate. The mean range of motion (ROM) was calculated from the load-displacement curves. RESULTS: BCFC significantly decreased ROM in lateral bending and axial rotation compared to other implants, and no significant difference in ROM between two types of PEEK cages and BCFC could be observed in flexion and extension. Anterior cervical plate (ACP) significantly decreased ROM in flexion and extension, but no significant difference in ROM between BCFC and bone graft plus ACP could be determined in lateral bending and axial rotation. CONCLUSION: The BCFC device showed better stability to autologous tricortical iliac crest bone graft and PEEK cages in single-level anterior cervical discectomy and fusion models and thus may be a potential alternative to the current PEEK cages. Dove Medical Press 2012 2012-11-27 /pmc/articles/PMC3512541/ /pubmed/23226018 http://dx.doi.org/10.2147/IJN.S38288 Text en © 2012 Cao et al, publisher and licensee Dove Medical Press Ltd. This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited.
spellingShingle Original Research
Cao, Lu
Duan, Ping-Guo
Li, Xi-Lei
Yuan, Feng-Lai
Zhao, Ming-Dong
Che, Wu
Wang, Hui-Ren
Dong, Jian
Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models
title Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models
title_full Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models
title_fullStr Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models
title_full_unstemmed Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models
title_short Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models
title_sort biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3512541/
https://www.ncbi.nlm.nih.gov/pubmed/23226018
http://dx.doi.org/10.2147/IJN.S38288
work_keys_str_mv AT caolu biomechanicalstabilityofabioabsorbableselfretainingpolylacticacidnanosizedbtricalciumphosphatecervicalspineinterbodyfusiondeviceinsinglelevelanteriorcervicaldiscectomyandfusionsheepmodels
AT duanpingguo biomechanicalstabilityofabioabsorbableselfretainingpolylacticacidnanosizedbtricalciumphosphatecervicalspineinterbodyfusiondeviceinsinglelevelanteriorcervicaldiscectomyandfusionsheepmodels
AT lixilei biomechanicalstabilityofabioabsorbableselfretainingpolylacticacidnanosizedbtricalciumphosphatecervicalspineinterbodyfusiondeviceinsinglelevelanteriorcervicaldiscectomyandfusionsheepmodels
AT yuanfenglai biomechanicalstabilityofabioabsorbableselfretainingpolylacticacidnanosizedbtricalciumphosphatecervicalspineinterbodyfusiondeviceinsinglelevelanteriorcervicaldiscectomyandfusionsheepmodels
AT zhaomingdong biomechanicalstabilityofabioabsorbableselfretainingpolylacticacidnanosizedbtricalciumphosphatecervicalspineinterbodyfusiondeviceinsinglelevelanteriorcervicaldiscectomyandfusionsheepmodels
AT chewu biomechanicalstabilityofabioabsorbableselfretainingpolylacticacidnanosizedbtricalciumphosphatecervicalspineinterbodyfusiondeviceinsinglelevelanteriorcervicaldiscectomyandfusionsheepmodels
AT wanghuiren biomechanicalstabilityofabioabsorbableselfretainingpolylacticacidnanosizedbtricalciumphosphatecervicalspineinterbodyfusiondeviceinsinglelevelanteriorcervicaldiscectomyandfusionsheepmodels
AT dongjian biomechanicalstabilityofabioabsorbableselfretainingpolylacticacidnanosizedbtricalciumphosphatecervicalspineinterbodyfusiondeviceinsinglelevelanteriorcervicaldiscectomyandfusionsheepmodels