Cargando…

Clinical outcomes between different stent designs with the same polymer and drug: comparison between the Taxus Express and Taxus Liberte stents

BACKGROUND/AIMS: The Taxus Liberte stent (Boston Scientific Co.) evolved from the Taxus Express stent, with enhanced stent deliverability and uniform drug delivery. This study was designed to compare angiographic and clinical outcomes in real-world practice between the Taxus Liberte and Taxus Expres...

Descripción completa

Detalles Bibliográficos
Autores principales: Son, Jang-Won, Kim, Ung, Park, Jong-Seon, Kim, Young-Jo, Jang, Jae-Sik, Yang, Tae-Hyun, Kim, Dong-Soo, Kim, Dong-Kie, Seol, Sang-Hoon, Kim, Doo-Il, Nam, Chang-Wook, Hur, Seung-Ho, Kim, Kwon-Bae
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Association of Internal Medicine 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3543963/
https://www.ncbi.nlm.nih.gov/pubmed/23345999
http://dx.doi.org/10.3904/kjim.2013.28.1.72
_version_ 1782255728967286784
author Son, Jang-Won
Kim, Ung
Park, Jong-Seon
Kim, Young-Jo
Jang, Jae-Sik
Yang, Tae-Hyun
Kim, Dong-Soo
Kim, Dong-Kie
Seol, Sang-Hoon
Kim, Doo-Il
Nam, Chang-Wook
Hur, Seung-Ho
Kim, Kwon-Bae
author_facet Son, Jang-Won
Kim, Ung
Park, Jong-Seon
Kim, Young-Jo
Jang, Jae-Sik
Yang, Tae-Hyun
Kim, Dong-Soo
Kim, Dong-Kie
Seol, Sang-Hoon
Kim, Doo-Il
Nam, Chang-Wook
Hur, Seung-Ho
Kim, Kwon-Bae
author_sort Son, Jang-Won
collection PubMed
description BACKGROUND/AIMS: The Taxus Liberte stent (Boston Scientific Co.) evolved from the Taxus Express stent, with enhanced stent deliverability and uniform drug delivery. This study was designed to compare angiographic and clinical outcomes in real-world practice between the Taxus Liberte and Taxus Express stents. METHODS: Between 2006 and 2008, 240 patients receiving the Taxus Liberte stent at three centers were registered and compared to historical control patients who had received the Taxus Express stent (n = 272). After propensity score matching, 173 patients treated with the Taxus Liberte stent and the same number of patients treated with the Taxus Express stent were selected. The primary outcome was a composite of major adverse cardiac events (MACE), including cardiac death, myocardial infarction (MI), ischemia driven target vessel revascularization (TVR), and stent thrombosis (ST) at 1 year. An additional angiographic assessment was conducted at 9 to 12 months. RESULTS: The study showed no significant difference between the Taxus Express and Taxus Liberte stents (death, 1.73% vs. 2.31%, p = 1.000; MI, 0% vs. 1.73%, p = 0.2478; TVR, 2.31% vs. 1.16%, p = 0.6848; and ST, 0% vs. 1.16%, p = 0.4986). The total MACE rate at 1 year did not differ between the groups (4.05% in Taxus Express vs. 4.05% in Taxus Liberte, p = 1.000). In addition, the binary restenosis rate did not differ (2.25% in Taxus Express vs. 1.80% in Taxus Liberte, p = 0.6848). CONCLUSIONS: In real-world experience with the two Taxus stent designs, both stents showed similarly good clinical and angiographic outcomes at 1 year. A long-term follow-up study is warranted.
format Online
Article
Text
id pubmed-3543963
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher The Korean Association of Internal Medicine
record_format MEDLINE/PubMed
spelling pubmed-35439632013-01-23 Clinical outcomes between different stent designs with the same polymer and drug: comparison between the Taxus Express and Taxus Liberte stents Son, Jang-Won Kim, Ung Park, Jong-Seon Kim, Young-Jo Jang, Jae-Sik Yang, Tae-Hyun Kim, Dong-Soo Kim, Dong-Kie Seol, Sang-Hoon Kim, Doo-Il Nam, Chang-Wook Hur, Seung-Ho Kim, Kwon-Bae Korean J Intern Med Original Article BACKGROUND/AIMS: The Taxus Liberte stent (Boston Scientific Co.) evolved from the Taxus Express stent, with enhanced stent deliverability and uniform drug delivery. This study was designed to compare angiographic and clinical outcomes in real-world practice between the Taxus Liberte and Taxus Express stents. METHODS: Between 2006 and 2008, 240 patients receiving the Taxus Liberte stent at three centers were registered and compared to historical control patients who had received the Taxus Express stent (n = 272). After propensity score matching, 173 patients treated with the Taxus Liberte stent and the same number of patients treated with the Taxus Express stent were selected. The primary outcome was a composite of major adverse cardiac events (MACE), including cardiac death, myocardial infarction (MI), ischemia driven target vessel revascularization (TVR), and stent thrombosis (ST) at 1 year. An additional angiographic assessment was conducted at 9 to 12 months. RESULTS: The study showed no significant difference between the Taxus Express and Taxus Liberte stents (death, 1.73% vs. 2.31%, p = 1.000; MI, 0% vs. 1.73%, p = 0.2478; TVR, 2.31% vs. 1.16%, p = 0.6848; and ST, 0% vs. 1.16%, p = 0.4986). The total MACE rate at 1 year did not differ between the groups (4.05% in Taxus Express vs. 4.05% in Taxus Liberte, p = 1.000). In addition, the binary restenosis rate did not differ (2.25% in Taxus Express vs. 1.80% in Taxus Liberte, p = 0.6848). CONCLUSIONS: In real-world experience with the two Taxus stent designs, both stents showed similarly good clinical and angiographic outcomes at 1 year. A long-term follow-up study is warranted. The Korean Association of Internal Medicine 2013-01 2012-12-28 /pmc/articles/PMC3543963/ /pubmed/23345999 http://dx.doi.org/10.3904/kjim.2013.28.1.72 Text en Copyright © 2013 The Korean Association of Internal Medicine http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Son, Jang-Won
Kim, Ung
Park, Jong-Seon
Kim, Young-Jo
Jang, Jae-Sik
Yang, Tae-Hyun
Kim, Dong-Soo
Kim, Dong-Kie
Seol, Sang-Hoon
Kim, Doo-Il
Nam, Chang-Wook
Hur, Seung-Ho
Kim, Kwon-Bae
Clinical outcomes between different stent designs with the same polymer and drug: comparison between the Taxus Express and Taxus Liberte stents
title Clinical outcomes between different stent designs with the same polymer and drug: comparison between the Taxus Express and Taxus Liberte stents
title_full Clinical outcomes between different stent designs with the same polymer and drug: comparison between the Taxus Express and Taxus Liberte stents
title_fullStr Clinical outcomes between different stent designs with the same polymer and drug: comparison between the Taxus Express and Taxus Liberte stents
title_full_unstemmed Clinical outcomes between different stent designs with the same polymer and drug: comparison between the Taxus Express and Taxus Liberte stents
title_short Clinical outcomes between different stent designs with the same polymer and drug: comparison between the Taxus Express and Taxus Liberte stents
title_sort clinical outcomes between different stent designs with the same polymer and drug: comparison between the taxus express and taxus liberte stents
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3543963/
https://www.ncbi.nlm.nih.gov/pubmed/23345999
http://dx.doi.org/10.3904/kjim.2013.28.1.72
work_keys_str_mv AT sonjangwon clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT kimung clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT parkjongseon clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT kimyoungjo clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT jangjaesik clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT yangtaehyun clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT kimdongsoo clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT kimdongkie clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT seolsanghoon clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT kimdooil clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT namchangwook clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT hurseungho clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents
AT kimkwonbae clinicaloutcomesbetweendifferentstentdesignswiththesamepolymeranddrugcomparisonbetweenthetaxusexpressandtaxuslibertestents