A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle(3) and BrainLAB iPlan RT Dose
BACKGROUND: Carbon fiber (CF) is now the material of choice for radiation therapy couch tops. Initial designs included side metal bars for rigidity; however, with the advent of IGRT, involving on board imaging, new thicker CF couch tops without metal bars have been developed. The new design allows f...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3549905/ https://www.ncbi.nlm.nih.gov/pubmed/23140425 http://dx.doi.org/10.1186/1748-717X-7-190 |
_version_ | 1782256495534014464 |
---|---|
author | Njeh, Christopher F Parker, Jason Spurgin, Joseph Rhoe, Elizabeth |
author_facet | Njeh, Christopher F Parker, Jason Spurgin, Joseph Rhoe, Elizabeth |
author_sort | Njeh, Christopher F |
collection | PubMed |
description | BACKGROUND: Carbon fiber (CF) is now the material of choice for radiation therapy couch tops. Initial designs included side metal bars for rigidity; however, with the advent of IGRT, involving on board imaging, new thicker CF couch tops without metal bars have been developed. The new design allows for excellent imaging at the expense of potentially unacceptable dose attenuation and perturbation. OBJECTIVES: We set out to model the BrainLAB imaging couch top (ICT) in Philips Pinnacle(3) treatment planning system (TPS), to validate the already modeled ICT in BrainLAB iPlan RT Dose treatment planning system and to compute the magnitude of the loss in skin sparing. RESULTS: Using CF density of 0.55 g/cm(3) and foam density of 0.03 g/cm(3), we demonstrated an excellent agreement between measured dose and Pinnacle(3) TPS computed dose using 6 MV beam. The agreement was within 1% for all gantry angle measured except for 120(o), which was 1.8%. The measured and iPlan RT Dose TPS computed dose agreed to within 1% for all gantry angles and field sizes measured except for 100(o) where the agreement was 1.4% for 10 cm × 10 cm field size. Predicted attenuation through the couch by iPlan RT Dose TPS (3.4% - 9.5%) and Pinnacle(3) TPS (2% - 6.6%) were within the same magnitude and similar to previously reported in the literature. Pinnacle(3) TPS estimated an 8% to 20% increase in skin dose with increase in field size. With the introduction of the CF couch top, it estimated an increase in skin dose by approximately 46 - 90%. The clinical impact of omitting the couch in treatment planning will be dependent on the beam arrangement, the percentage of the beams intersecting the couch and their angles of incidence. CONCLUSION: We have successfully modeled the ICT in Pinnacle(3) TPS and validated the modeled ICT in iPlan RT Dose. It is recommended that the ICT be included in treatment planning for all treatments that involve posteriors beams. There is a significant increase in skin dose that is dependent on the percentage of the beam passing through the couch and the angle of incidence. |
format | Online Article Text |
id | pubmed-3549905 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-35499052013-01-24 A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle(3) and BrainLAB iPlan RT Dose Njeh, Christopher F Parker, Jason Spurgin, Joseph Rhoe, Elizabeth Radiat Oncol Research BACKGROUND: Carbon fiber (CF) is now the material of choice for radiation therapy couch tops. Initial designs included side metal bars for rigidity; however, with the advent of IGRT, involving on board imaging, new thicker CF couch tops without metal bars have been developed. The new design allows for excellent imaging at the expense of potentially unacceptable dose attenuation and perturbation. OBJECTIVES: We set out to model the BrainLAB imaging couch top (ICT) in Philips Pinnacle(3) treatment planning system (TPS), to validate the already modeled ICT in BrainLAB iPlan RT Dose treatment planning system and to compute the magnitude of the loss in skin sparing. RESULTS: Using CF density of 0.55 g/cm(3) and foam density of 0.03 g/cm(3), we demonstrated an excellent agreement between measured dose and Pinnacle(3) TPS computed dose using 6 MV beam. The agreement was within 1% for all gantry angle measured except for 120(o), which was 1.8%. The measured and iPlan RT Dose TPS computed dose agreed to within 1% for all gantry angles and field sizes measured except for 100(o) where the agreement was 1.4% for 10 cm × 10 cm field size. Predicted attenuation through the couch by iPlan RT Dose TPS (3.4% - 9.5%) and Pinnacle(3) TPS (2% - 6.6%) were within the same magnitude and similar to previously reported in the literature. Pinnacle(3) TPS estimated an 8% to 20% increase in skin dose with increase in field size. With the introduction of the CF couch top, it estimated an increase in skin dose by approximately 46 - 90%. The clinical impact of omitting the couch in treatment planning will be dependent on the beam arrangement, the percentage of the beams intersecting the couch and their angles of incidence. CONCLUSION: We have successfully modeled the ICT in Pinnacle(3) TPS and validated the modeled ICT in iPlan RT Dose. It is recommended that the ICT be included in treatment planning for all treatments that involve posteriors beams. There is a significant increase in skin dose that is dependent on the percentage of the beam passing through the couch and the angle of incidence. BioMed Central 2012-11-09 /pmc/articles/PMC3549905/ /pubmed/23140425 http://dx.doi.org/10.1186/1748-717X-7-190 Text en Copyright ©2012 Njeh et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Njeh, Christopher F Parker, Jason Spurgin, Joseph Rhoe, Elizabeth A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle(3) and BrainLAB iPlan RT Dose |
title | A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle(3) and BrainLAB iPlan RT Dose |
title_full | A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle(3) and BrainLAB iPlan RT Dose |
title_fullStr | A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle(3) and BrainLAB iPlan RT Dose |
title_full_unstemmed | A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle(3) and BrainLAB iPlan RT Dose |
title_short | A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle(3) and BrainLAB iPlan RT Dose |
title_sort | validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: philips pinnacle(3) and brainlab iplan rt dose |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3549905/ https://www.ncbi.nlm.nih.gov/pubmed/23140425 http://dx.doi.org/10.1186/1748-717X-7-190 |
work_keys_str_mv | AT njehchristopherf avalidationofcarbonfiberimagingcouchtopmodelingintworadiationtherapytreatmentplanningsystemsphilipspinnacle3andbrainlabiplanrtdose AT parkerjason avalidationofcarbonfiberimagingcouchtopmodelingintworadiationtherapytreatmentplanningsystemsphilipspinnacle3andbrainlabiplanrtdose AT spurginjoseph avalidationofcarbonfiberimagingcouchtopmodelingintworadiationtherapytreatmentplanningsystemsphilipspinnacle3andbrainlabiplanrtdose AT rhoeelizabeth avalidationofcarbonfiberimagingcouchtopmodelingintworadiationtherapytreatmentplanningsystemsphilipspinnacle3andbrainlabiplanrtdose AT njehchristopherf validationofcarbonfiberimagingcouchtopmodelingintworadiationtherapytreatmentplanningsystemsphilipspinnacle3andbrainlabiplanrtdose AT parkerjason validationofcarbonfiberimagingcouchtopmodelingintworadiationtherapytreatmentplanningsystemsphilipspinnacle3andbrainlabiplanrtdose AT spurginjoseph validationofcarbonfiberimagingcouchtopmodelingintworadiationtherapytreatmentplanningsystemsphilipspinnacle3andbrainlabiplanrtdose AT rhoeelizabeth validationofcarbonfiberimagingcouchtopmodelingintworadiationtherapytreatmentplanningsystemsphilipspinnacle3andbrainlabiplanrtdose |