Cargando…
Comparison of (60)Co and (192)Ir sources in HDR brachytherapy
This paper compares the isotopes (60)Co and (192)Ir as radiation sources for high-dose-rate (HDR) afterloading brachytherapy. The smaller size of (192)Ir sources made it the preferred radionuclide for temporary brachytherapy treatments. Recently also (60)Co sources have been made available with iden...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Termedia Publishing House
2011
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3551360/ https://www.ncbi.nlm.nih.gov/pubmed/23346129 http://dx.doi.org/10.5114/jcb.2011.26471 |
Sumario: | This paper compares the isotopes (60)Co and (192)Ir as radiation sources for high-dose-rate (HDR) afterloading brachytherapy. The smaller size of (192)Ir sources made it the preferred radionuclide for temporary brachytherapy treatments. Recently also (60)Co sources have been made available with identical geometrical dimensions. This paper compares the characteristics of both nuclides in different fields of brachytherapy based on scientific literature. In an additional part of this paper reports from medical physicists of several radiation therapy institutes are discussed. The purpose of this work is to investigate the advantages or disadvantages of both radionuclides for HDR brachytherapy due to their physical differences. The motivation is to provide useful information to support decision-making procedures in the selection of equipment for brachytherapy treatment rooms. The results of this work show that no advantages or disadvantages exist for (60)Co sources compared to (192)Ir sources with regard to clinical aspects. Nevertheless, there are potential logistical advantages of (60)Co sources due to its longer half-life (5.3 years vs. 74 days), making it an interesting alternative especially in developing countries. |
---|