I-125 seed calibration using the SeedSelectron(®) afterloader: a practical solution to fulfill AAPM-ESTRO recommendations
PURPOSE: SeedSelectron(®) v1.26b (Nucletron BV, The Netherlands) is an afterloader system used in prostate interstitial permanent brachytherapy with I-125 selectSeed seeds. It contains a diode array to assay all implanted seeds. Only one or two seeds can be extracted during the surgical procedure an...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Termedia Publishing House
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3551372/ https://www.ncbi.nlm.nih.gov/pubmed/23346136 http://dx.doi.org/10.5114/jcb.2012.27948 |
_version_ | 1782256550114492416 |
---|---|
author | Perez-Calatayud, Jose Richart, Jose Guirado, Damián Pérez-García, Jordi Rodríguez, Silvia Santos, Manuel Ballester, Facundo |
author_facet | Perez-Calatayud, Jose Richart, Jose Guirado, Damián Pérez-García, Jordi Rodríguez, Silvia Santos, Manuel Ballester, Facundo |
author_sort | Perez-Calatayud, Jose |
collection | PubMed |
description | PURPOSE: SeedSelectron(®) v1.26b (Nucletron BV, The Netherlands) is an afterloader system used in prostate interstitial permanent brachytherapy with I-125 selectSeed seeds. It contains a diode array to assay all implanted seeds. Only one or two seeds can be extracted during the surgical procedure and assayed using a well chamber to check the manufacturer air-kerma strength (S(K)) and to calibrate the diode array. Therefore, it is not feasible to assay 5–10% seeds as required by the AAPM-ESTRO. In this study, we present a practical solution of the SeedSelectron(®) users to fulfill the AAPM- ESTRO recommendations. MATERIAL AND METHODS: The method is based on: a) the SourceCheck(®) well ionization chamber (PTW, Germany) provided with a PTW insert; b) n = 10 selectSeed from the same batch and class as the seeds for the implant; c) the Nucletron insert to accommodate the n = 10 seeds on the SourceCheck(®) and to measure their averaged S(K). Results for 56 implants have been studied comparing the S(K) value from the manufacturer with the one obtained with the n = 10 seeds using the Nucletron insert prior to the implant and with the S(K) of just one seed measured with the PTW insert during the implant. RESULTS: We are faced with S(K) deviation for individual seeds up to 7.8%. However, in the majority of cases S(K) is in agreement with the manufacturer value. With the method proposed using the Nucletron insert, the large deviations of S(K) are reduced and for 56 implants studied no deviation outside the range of the class were found. CONCLUSIONS: The new Nucletron insert and the proposed procedure allow to evaluate the S(K) of the n = 10 seeds prior to the implant, fulfilling the AAPM-ESTRO recommendations. It has been adopted by Nucletron to be extended to seedSelectron(®) users under request. |
format | Online Article Text |
id | pubmed-3551372 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Termedia Publishing House |
record_format | MEDLINE/PubMed |
spelling | pubmed-35513722013-01-23 I-125 seed calibration using the SeedSelectron(®) afterloader: a practical solution to fulfill AAPM-ESTRO recommendations Perez-Calatayud, Jose Richart, Jose Guirado, Damián Pérez-García, Jordi Rodríguez, Silvia Santos, Manuel Ballester, Facundo J Contemp Brachytherapy Original Article PURPOSE: SeedSelectron(®) v1.26b (Nucletron BV, The Netherlands) is an afterloader system used in prostate interstitial permanent brachytherapy with I-125 selectSeed seeds. It contains a diode array to assay all implanted seeds. Only one or two seeds can be extracted during the surgical procedure and assayed using a well chamber to check the manufacturer air-kerma strength (S(K)) and to calibrate the diode array. Therefore, it is not feasible to assay 5–10% seeds as required by the AAPM-ESTRO. In this study, we present a practical solution of the SeedSelectron(®) users to fulfill the AAPM- ESTRO recommendations. MATERIAL AND METHODS: The method is based on: a) the SourceCheck(®) well ionization chamber (PTW, Germany) provided with a PTW insert; b) n = 10 selectSeed from the same batch and class as the seeds for the implant; c) the Nucletron insert to accommodate the n = 10 seeds on the SourceCheck(®) and to measure their averaged S(K). Results for 56 implants have been studied comparing the S(K) value from the manufacturer with the one obtained with the n = 10 seeds using the Nucletron insert prior to the implant and with the S(K) of just one seed measured with the PTW insert during the implant. RESULTS: We are faced with S(K) deviation for individual seeds up to 7.8%. However, in the majority of cases S(K) is in agreement with the manufacturer value. With the method proposed using the Nucletron insert, the large deviations of S(K) are reduced and for 56 implants studied no deviation outside the range of the class were found. CONCLUSIONS: The new Nucletron insert and the proposed procedure allow to evaluate the S(K) of the n = 10 seeds prior to the implant, fulfilling the AAPM-ESTRO recommendations. It has been adopted by Nucletron to be extended to seedSelectron(®) users under request. Termedia Publishing House 2012-03-30 2012-03 /pmc/articles/PMC3551372/ /pubmed/23346136 http://dx.doi.org/10.5114/jcb.2012.27948 Text en Copyright © 2012 Termedia http://creativecommons.org/licenses/by-nc-nd/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-Noncommercial 3.0 Unported License, permitting all non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Perez-Calatayud, Jose Richart, Jose Guirado, Damián Pérez-García, Jordi Rodríguez, Silvia Santos, Manuel Ballester, Facundo I-125 seed calibration using the SeedSelectron(®) afterloader: a practical solution to fulfill AAPM-ESTRO recommendations |
title | I-125 seed calibration using the SeedSelectron(®) afterloader: a practical solution to fulfill AAPM-ESTRO recommendations |
title_full | I-125 seed calibration using the SeedSelectron(®) afterloader: a practical solution to fulfill AAPM-ESTRO recommendations |
title_fullStr | I-125 seed calibration using the SeedSelectron(®) afterloader: a practical solution to fulfill AAPM-ESTRO recommendations |
title_full_unstemmed | I-125 seed calibration using the SeedSelectron(®) afterloader: a practical solution to fulfill AAPM-ESTRO recommendations |
title_short | I-125 seed calibration using the SeedSelectron(®) afterloader: a practical solution to fulfill AAPM-ESTRO recommendations |
title_sort | i-125 seed calibration using the seedselectron(®) afterloader: a practical solution to fulfill aapm-estro recommendations |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3551372/ https://www.ncbi.nlm.nih.gov/pubmed/23346136 http://dx.doi.org/10.5114/jcb.2012.27948 |
work_keys_str_mv | AT perezcalatayudjose i125seedcalibrationusingtheseedselectronafterloaderapracticalsolutiontofulfillaapmestrorecommendations AT richartjose i125seedcalibrationusingtheseedselectronafterloaderapracticalsolutiontofulfillaapmestrorecommendations AT guiradodamian i125seedcalibrationusingtheseedselectronafterloaderapracticalsolutiontofulfillaapmestrorecommendations AT perezgarciajordi i125seedcalibrationusingtheseedselectronafterloaderapracticalsolutiontofulfillaapmestrorecommendations AT rodriguezsilvia i125seedcalibrationusingtheseedselectronafterloaderapracticalsolutiontofulfillaapmestrorecommendations AT santosmanuel i125seedcalibrationusingtheseedselectronafterloaderapracticalsolutiontofulfillaapmestrorecommendations AT ballesterfacundo i125seedcalibrationusingtheseedselectronafterloaderapracticalsolutiontofulfillaapmestrorecommendations |