Cargando…
Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source
PURPOSE: Recently, a new HDR (60)Co brachytherapy source, Flexisource Co-60, has been developed (Nucletron B.V. Veenendaal, The Netherlands). This study aims to obtain dosimetric data for this source for its use in clinical practice as required by AAPM and ESTRO. MATERIAL AND METHODS: Two Monte Carl...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Termedia Publishing House
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3551374/ https://www.ncbi.nlm.nih.gov/pubmed/23346138 http://dx.doi.org/10.5114/jcb.2012.27950 |
_version_ | 1782256550592643072 |
---|---|
author | Vijande, Javier Granero, Domingo Perez-Calatayud, Jose Ballester, Facundo |
author_facet | Vijande, Javier Granero, Domingo Perez-Calatayud, Jose Ballester, Facundo |
author_sort | Vijande, Javier |
collection | PubMed |
description | PURPOSE: Recently, a new HDR (60)Co brachytherapy source, Flexisource Co-60, has been developed (Nucletron B.V. Veenendaal, The Netherlands). This study aims to obtain dosimetric data for this source for its use in clinical practice as required by AAPM and ESTRO. MATERIAL AND METHODS: Two Monte Carlo radiation transport codes were used: Penelope2008 and GEANT4. The source was centrally-positioned in a 100 cm radius water phantom. Absorbed dose and collisional kerma were obtained using 0.01 cm (close) and 0.1 cm (far) sized voxels to provide high-resolution dosimetry near (far from) the source. Dose rate distributions obtained with the two Monte Carlo codes were compared. RESULTS AND DISCUSSION: Simulations performed with those two radiation transport codes showed an agreement typically within 0.2% for r > 0.8 cm and up to 2% closer to the source. Detailed results of dose distributions are being made available. CONCLUSIONS: Dosimetric data are provided for the new Flexisource Co-60 source. These data are meant to be used in treatment planning systems in clinical practice. |
format | Online Article Text |
id | pubmed-3551374 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Termedia Publishing House |
record_format | MEDLINE/PubMed |
spelling | pubmed-35513742013-01-23 Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source Vijande, Javier Granero, Domingo Perez-Calatayud, Jose Ballester, Facundo J Contemp Brachytherapy Original Article PURPOSE: Recently, a new HDR (60)Co brachytherapy source, Flexisource Co-60, has been developed (Nucletron B.V. Veenendaal, The Netherlands). This study aims to obtain dosimetric data for this source for its use in clinical practice as required by AAPM and ESTRO. MATERIAL AND METHODS: Two Monte Carlo radiation transport codes were used: Penelope2008 and GEANT4. The source was centrally-positioned in a 100 cm radius water phantom. Absorbed dose and collisional kerma were obtained using 0.01 cm (close) and 0.1 cm (far) sized voxels to provide high-resolution dosimetry near (far from) the source. Dose rate distributions obtained with the two Monte Carlo codes were compared. RESULTS AND DISCUSSION: Simulations performed with those two radiation transport codes showed an agreement typically within 0.2% for r > 0.8 cm and up to 2% closer to the source. Detailed results of dose distributions are being made available. CONCLUSIONS: Dosimetric data are provided for the new Flexisource Co-60 source. These data are meant to be used in treatment planning systems in clinical practice. Termedia Publishing House 2012-03-30 2012-03 /pmc/articles/PMC3551374/ /pubmed/23346138 http://dx.doi.org/10.5114/jcb.2012.27950 Text en Copyright © 2012 Termedia http://creativecommons.org/licenses/by-nc-nd/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-Noncommercial 3.0 Unported License, permitting all non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Vijande, Javier Granero, Domingo Perez-Calatayud, Jose Ballester, Facundo Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source |
title | Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source |
title_full | Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source |
title_fullStr | Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source |
title_full_unstemmed | Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source |
title_short | Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source |
title_sort | monte carlo dosimetric study of the flexisource co-60 high dose rate source |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3551374/ https://www.ncbi.nlm.nih.gov/pubmed/23346138 http://dx.doi.org/10.5114/jcb.2012.27950 |
work_keys_str_mv | AT vijandejavier montecarlodosimetricstudyoftheflexisourceco60highdoseratesource AT granerodomingo montecarlodosimetricstudyoftheflexisourceco60highdoseratesource AT perezcalatayudjose montecarlodosimetricstudyoftheflexisourceco60highdoseratesource AT ballesterfacundo montecarlodosimetricstudyoftheflexisourceco60highdoseratesource |