Cargando…

Reverse hybrid and cemented hip replacement compared using radiostereometry and dual-energy X-ray absorptiometry: 43 hips followed for 2 years in a prospective trial

BACKGROUND AND PURPOSE: Total hip replacement (THR) with a reverse hybrid (RH), a combination of a cemented polyethylene cup and a cementless femoral stem, has been increasingly used in Scandinavia. In a randomized trial, we compared an RH THR with a proximal hydroxyapatite- (HA-) coated stem to a c...

Descripción completa

Detalles Bibliográficos
Autores principales: Lindalen, Einar, Dahl, Jon, Nordsletten, Lars, Snorrason, Finnur, Høvik, Øystein, Röhrl, Stephan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Informa Healthcare 2012
Materias:
Hip
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3555444/
https://www.ncbi.nlm.nih.gov/pubmed/23116437
http://dx.doi.org/10.3109/17453674.2012.742393
Descripción
Sumario:BACKGROUND AND PURPOSE: Total hip replacement (THR) with a reverse hybrid (RH), a combination of a cemented polyethylene cup and a cementless femoral stem, has been increasingly used in Scandinavia. In a randomized trial, we compared an RH THR with a proximal hydroxyapatite- (HA-) coated stem to a conventional cemented THR. Both groups received the same polyethylene cup. PATIENTS AND METHODS: 51 patients (52 hips) were included. Radiostereometry (RSA) and dual-energy X-ray absorptiometry (DEXA) were performed postoperatively and after 6, 12, and 24 months. 42 patients (43 hips) were followed for 2 years. RESULTS: Mean cup rotation around the x-axis was 0.13° for the cemented group and –0.24° for the RH group (p = 0.03). Cup migration in the other axes, and stem migration and wear were similar between the 2 study groups. Bone remodeling around the cup was also similar between the groups. Bone loss in Gruen zone 1 was 18% for the cementless stems, as compared to an increase of 1.4% for the cemented ones (p < 0.001). Bone loss was similar in the other Gruen zones. Harris hip score and Oxford hip score were similar pre- and postoperatively in the 2 groups. INTERPRETATION: In the present study, RH THR with a cementless hydroxyapatite-coated stem and conventional cemented THR did not show any major differences regarding stem migration and bone loss after 2 years of follow-up.