Contrast-enhanced small-animal PET/CT in cancer research: strong improvement of diagnostic accuracy without significant alteration of quantitative accuracy and NEMA NU 4–2008 image quality parameters

BACKGROUND: The use of iodinated contrast media in small-animal positron emission tomography (PET)/computed tomography (CT) could improve anatomic referencing and tumor delineation but may introduce inaccuracies in the attenuation correction of the PET images. This study evaluated the diagnostic per...

Descripción completa

Detalles Bibliográficos
Autores principales: Lasnon, Charline, Quak, Elske, Briand, Mélanie, Gu, Zheng, Louis, Marie-Hélène, Aide, Nicolas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3563455/
https://www.ncbi.nlm.nih.gov/pubmed/23327687
http://dx.doi.org/10.1186/2191-219X-3-5
_version_ 1782258187475353600
author Lasnon, Charline
Quak, Elske
Briand, Mélanie
Gu, Zheng
Louis, Marie-Hélène
Aide, Nicolas
author_facet Lasnon, Charline
Quak, Elske
Briand, Mélanie
Gu, Zheng
Louis, Marie-Hélène
Aide, Nicolas
author_sort Lasnon, Charline
collection PubMed
description BACKGROUND: The use of iodinated contrast media in small-animal positron emission tomography (PET)/computed tomography (CT) could improve anatomic referencing and tumor delineation but may introduce inaccuracies in the attenuation correction of the PET images. This study evaluated the diagnostic performance and accuracy of quantitative values in contrast-enhanced small-animal PET/CT ((CE)PET/CT) as compared to unenhanced small animal PET/CT ((UE)PET/CT). METHODS: Firstly, a NEMA NU 4–2008 phantom (filled with (18)F-FDG or (18)F-FDG plus contrast media) and a homemade phantom, mimicking an abdominal tumor surrounded by water or contrast media, were used to evaluate the impact of iodinated contrast media on the image quality parameters and accuracy of quantitative values for a pertinent-sized target. Secondly, two studies in 22 abdominal tumor-bearing mice and rats were performed. The first animal experiment studied the impact of a dual-contrast media protocol, comprising the intravenous injection of a long-lasting contrast agent mixed with (18)F-FDG and the intraperitoneal injection of contrast media, on tumor delineation and the accuracy of quantitative values. The second animal experiment compared the diagnostic performance and quantitative values of (CE)PET/CT versus (UE)PET/CT by sacrificing the animals after the tracer uptake period and imaging them before and after intraperitoneal injection of contrast media. RESULTS: There was minimal impact on IQ parameters (%SD(unif) and spillover ratios in air and water) when the NEMA NU 4–2008 phantom was filled with (18)F-FDG plus contrast media. In the homemade phantom, measured activity was similar to true activity (−0.02%) and overestimated by 10.30% when vials were surrounded by water or by an iodine solution, respectively. The first animal experiment showed excellent tumor delineation and a good correlation between small-animal (SA)-PET and ex vivo quantification (r(2) = 0.87, P < 0.0001). The second animal experiment showed a good correlation between (CE)PET/CT and (UE)PET/CT quantitative values (r(2) = 0.99, P < 0.0001). Receiver operating characteristic analysis demonstrated better diagnostic accuracy of (CE)PET/CT versus (UE)PET/CT (senior researcher, area under the curve (AUC) 0.96 versus 0.77, P = 0.004; junior researcher, AUC 0.78 versus 0.58, P = 0.004). CONCLUSIONS: The use of iodinated contrast media for small-animal PET imaging significantly improves tumor delineation and diagnostic performance, without significant alteration of SA-PET quantitative accuracy and NEMA NU 4–2008 IQ parameters.
format Online
Article
Text
id pubmed-3563455
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Springer
record_format MEDLINE/PubMed
spelling pubmed-35634552013-02-05 Contrast-enhanced small-animal PET/CT in cancer research: strong improvement of diagnostic accuracy without significant alteration of quantitative accuracy and NEMA NU 4–2008 image quality parameters Lasnon, Charline Quak, Elske Briand, Mélanie Gu, Zheng Louis, Marie-Hélène Aide, Nicolas EJNMMI Res Original Research BACKGROUND: The use of iodinated contrast media in small-animal positron emission tomography (PET)/computed tomography (CT) could improve anatomic referencing and tumor delineation but may introduce inaccuracies in the attenuation correction of the PET images. This study evaluated the diagnostic performance and accuracy of quantitative values in contrast-enhanced small-animal PET/CT ((CE)PET/CT) as compared to unenhanced small animal PET/CT ((UE)PET/CT). METHODS: Firstly, a NEMA NU 4–2008 phantom (filled with (18)F-FDG or (18)F-FDG plus contrast media) and a homemade phantom, mimicking an abdominal tumor surrounded by water or contrast media, were used to evaluate the impact of iodinated contrast media on the image quality parameters and accuracy of quantitative values for a pertinent-sized target. Secondly, two studies in 22 abdominal tumor-bearing mice and rats were performed. The first animal experiment studied the impact of a dual-contrast media protocol, comprising the intravenous injection of a long-lasting contrast agent mixed with (18)F-FDG and the intraperitoneal injection of contrast media, on tumor delineation and the accuracy of quantitative values. The second animal experiment compared the diagnostic performance and quantitative values of (CE)PET/CT versus (UE)PET/CT by sacrificing the animals after the tracer uptake period and imaging them before and after intraperitoneal injection of contrast media. RESULTS: There was minimal impact on IQ parameters (%SD(unif) and spillover ratios in air and water) when the NEMA NU 4–2008 phantom was filled with (18)F-FDG plus contrast media. In the homemade phantom, measured activity was similar to true activity (−0.02%) and overestimated by 10.30% when vials were surrounded by water or by an iodine solution, respectively. The first animal experiment showed excellent tumor delineation and a good correlation between small-animal (SA)-PET and ex vivo quantification (r(2) = 0.87, P < 0.0001). The second animal experiment showed a good correlation between (CE)PET/CT and (UE)PET/CT quantitative values (r(2) = 0.99, P < 0.0001). Receiver operating characteristic analysis demonstrated better diagnostic accuracy of (CE)PET/CT versus (UE)PET/CT (senior researcher, area under the curve (AUC) 0.96 versus 0.77, P = 0.004; junior researcher, AUC 0.78 versus 0.58, P = 0.004). CONCLUSIONS: The use of iodinated contrast media for small-animal PET imaging significantly improves tumor delineation and diagnostic performance, without significant alteration of SA-PET quantitative accuracy and NEMA NU 4–2008 IQ parameters. Springer 2013-01-17 /pmc/articles/PMC3563455/ /pubmed/23327687 http://dx.doi.org/10.1186/2191-219X-3-5 Text en Copyright ©2013 Lasnon et al.; licensee Springer. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Research
Lasnon, Charline
Quak, Elske
Briand, Mélanie
Gu, Zheng
Louis, Marie-Hélène
Aide, Nicolas
Contrast-enhanced small-animal PET/CT in cancer research: strong improvement of diagnostic accuracy without significant alteration of quantitative accuracy and NEMA NU 4–2008 image quality parameters
title Contrast-enhanced small-animal PET/CT in cancer research: strong improvement of diagnostic accuracy without significant alteration of quantitative accuracy and NEMA NU 4–2008 image quality parameters
title_full Contrast-enhanced small-animal PET/CT in cancer research: strong improvement of diagnostic accuracy without significant alteration of quantitative accuracy and NEMA NU 4–2008 image quality parameters
title_fullStr Contrast-enhanced small-animal PET/CT in cancer research: strong improvement of diagnostic accuracy without significant alteration of quantitative accuracy and NEMA NU 4–2008 image quality parameters
title_full_unstemmed Contrast-enhanced small-animal PET/CT in cancer research: strong improvement of diagnostic accuracy without significant alteration of quantitative accuracy and NEMA NU 4–2008 image quality parameters
title_short Contrast-enhanced small-animal PET/CT in cancer research: strong improvement of diagnostic accuracy without significant alteration of quantitative accuracy and NEMA NU 4–2008 image quality parameters
title_sort contrast-enhanced small-animal pet/ct in cancer research: strong improvement of diagnostic accuracy without significant alteration of quantitative accuracy and nema nu 4–2008 image quality parameters
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3563455/
https://www.ncbi.nlm.nih.gov/pubmed/23327687
http://dx.doi.org/10.1186/2191-219X-3-5
work_keys_str_mv AT lasnoncharline contrastenhancedsmallanimalpetctincancerresearchstrongimprovementofdiagnosticaccuracywithoutsignificantalterationofquantitativeaccuracyandnemanu42008imagequalityparameters
AT quakelske contrastenhancedsmallanimalpetctincancerresearchstrongimprovementofdiagnosticaccuracywithoutsignificantalterationofquantitativeaccuracyandnemanu42008imagequalityparameters
AT briandmelanie contrastenhancedsmallanimalpetctincancerresearchstrongimprovementofdiagnosticaccuracywithoutsignificantalterationofquantitativeaccuracyandnemanu42008imagequalityparameters
AT guzheng contrastenhancedsmallanimalpetctincancerresearchstrongimprovementofdiagnosticaccuracywithoutsignificantalterationofquantitativeaccuracyandnemanu42008imagequalityparameters
AT louismariehelene contrastenhancedsmallanimalpetctincancerresearchstrongimprovementofdiagnosticaccuracywithoutsignificantalterationofquantitativeaccuracyandnemanu42008imagequalityparameters
AT aidenicolas contrastenhancedsmallanimalpetctincancerresearchstrongimprovementofdiagnosticaccuracywithoutsignificantalterationofquantitativeaccuracyandnemanu42008imagequalityparameters