The effect of breast shielding during lumbar spine radiography

BACKGROUND: The aim of the study was to determine the influence of lead shielding on the dose to female breasts in conventional x-ray lumbar spine imaging. The correlation between the body mass index and the dose received by the breast was also investigated. MATERIALS AND METHODS: Breast surface dos...

Descripción completa

Detalles Bibliográficos
Autores principales: Mekis, Nejc, Zontar, Dejan, Skrk, Damijan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Versita, Warsaw 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3573831/
https://www.ncbi.nlm.nih.gov/pubmed/23450158
http://dx.doi.org/10.2478/raon-2013-0004
_version_ 1782259512176017408
author Mekis, Nejc
Zontar, Dejan
Skrk, Damijan
author_facet Mekis, Nejc
Zontar, Dejan
Skrk, Damijan
author_sort Mekis, Nejc
collection PubMed
description BACKGROUND: The aim of the study was to determine the influence of lead shielding on the dose to female breasts in conventional x-ray lumbar spine imaging. The correlation between the body mass index and the dose received by the breast was also investigated. MATERIALS AND METHODS: Breast surface dose was measured by thermoluminescent dosimeters (TLD). In the first phase measurements of breast dose with and without shielding from lumbar spine imaging in two projections were conducted on an anthropomorphic phantom. In the second stage measurements were performed on 100 female patients, randomly divided into two groups of 50, with breast shielding only used in one group. RESULTS: On average, breast exposure dose in lumbar spine imaging in both projections (anteroposterior (AP) and lateral) was found reduced by approximately 80% (p < 0,001) when shielding with 0.5 mm lead equivalent was used (from 0.45±0.25 mGy to 0.09±0.07 mGy on the right and from 0.26±0.14 mGy to 0.06±0.04 mGy on the left breast). No correlation between the body mass index (BMI) and the breast surface radiation dose was observed. CONCLUSIONS: Although during the lumbar spine imaging breasts receive low-dose exposure even when shielding is not used, the dose can be reduced up to 80% by breast shielding with no influence on the image quality.
format Online
Article
Text
id pubmed-3573831
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Versita, Warsaw
record_format MEDLINE/PubMed
spelling pubmed-35738312013-03-01 The effect of breast shielding during lumbar spine radiography Mekis, Nejc Zontar, Dejan Skrk, Damijan Radiol Oncol Research Article BACKGROUND: The aim of the study was to determine the influence of lead shielding on the dose to female breasts in conventional x-ray lumbar spine imaging. The correlation between the body mass index and the dose received by the breast was also investigated. MATERIALS AND METHODS: Breast surface dose was measured by thermoluminescent dosimeters (TLD). In the first phase measurements of breast dose with and without shielding from lumbar spine imaging in two projections were conducted on an anthropomorphic phantom. In the second stage measurements were performed on 100 female patients, randomly divided into two groups of 50, with breast shielding only used in one group. RESULTS: On average, breast exposure dose in lumbar spine imaging in both projections (anteroposterior (AP) and lateral) was found reduced by approximately 80% (p < 0,001) when shielding with 0.5 mm lead equivalent was used (from 0.45±0.25 mGy to 0.09±0.07 mGy on the right and from 0.26±0.14 mGy to 0.06±0.04 mGy on the left breast). No correlation between the body mass index (BMI) and the breast surface radiation dose was observed. CONCLUSIONS: Although during the lumbar spine imaging breasts receive low-dose exposure even when shielding is not used, the dose can be reduced up to 80% by breast shielding with no influence on the image quality. Versita, Warsaw 2013-02-01 /pmc/articles/PMC3573831/ /pubmed/23450158 http://dx.doi.org/10.2478/raon-2013-0004 Text en Copyright © by Association of Radiology & Oncology http://creativecommons.org/licenses/by/3.0/ This article is an open-access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).
spellingShingle Research Article
Mekis, Nejc
Zontar, Dejan
Skrk, Damijan
The effect of breast shielding during lumbar spine radiography
title The effect of breast shielding during lumbar spine radiography
title_full The effect of breast shielding during lumbar spine radiography
title_fullStr The effect of breast shielding during lumbar spine radiography
title_full_unstemmed The effect of breast shielding during lumbar spine radiography
title_short The effect of breast shielding during lumbar spine radiography
title_sort effect of breast shielding during lumbar spine radiography
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3573831/
https://www.ncbi.nlm.nih.gov/pubmed/23450158
http://dx.doi.org/10.2478/raon-2013-0004
work_keys_str_mv AT mekisnejc theeffectofbreastshieldingduringlumbarspineradiography
AT zontardejan theeffectofbreastshieldingduringlumbarspineradiography
AT skrkdamijan theeffectofbreastshieldingduringlumbarspineradiography
AT mekisnejc effectofbreastshieldingduringlumbarspineradiography
AT zontardejan effectofbreastshieldingduringlumbarspineradiography
AT skrkdamijan effectofbreastshieldingduringlumbarspineradiography