Transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the Accurus vitrectomy platform to the Constellation Vision System for microincisional vitrectomy surgery
BACKGROUND: Microincisional vitrectomy surgery (MIVS) is the current standard surgical approach for pars plana vitrectomy. Historically, the most common surgical platform for vitrectomy surgery, since its introduction in 1997, has been the Accurus vitrectomy system. Recent introduction of the next g...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove Medical Press
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3581355/ https://www.ncbi.nlm.nih.gov/pubmed/23450081 http://dx.doi.org/10.2147/OPTH.S35603 |
_version_ | 1782260405439037440 |
---|---|
author | Murray, Timothy G Layton, Andrew J Tong, Kuo B Gittelman, Michael Latiff, Azeema Gologorsky, Daniel Vigoda, Michael M |
author_facet | Murray, Timothy G Layton, Andrew J Tong, Kuo B Gittelman, Michael Latiff, Azeema Gologorsky, Daniel Vigoda, Michael M |
author_sort | Murray, Timothy G |
collection | PubMed |
description | BACKGROUND: Microincisional vitrectomy surgery (MIVS) is the current standard surgical approach for pars plana vitrectomy. Historically, the most common surgical platform for vitrectomy surgery, since its introduction in 1997, has been the Accurus vitrectomy system. Recent introduction of the next generation of vitrectomy platforms has generated concerns associated with transitioning to new technology in the operating room environment. This study compared, in a matched fashion, surgical use of the Accurus vitrectomy system and the next generation Constellation Vision System to evaluate surgical efficiencies, complications, and user perceptions of this transition. METHODS: Electronic health records were abstracted as a hospital quality assurance activity and included all vitreoretinal surgical procedures at the Bascom Palmer Eye Institute, Anne Bates Leach Eye Hospital, during two discrete 12-month time periods. These two periods reflected dedicated usage of the Accurus (June 2008–May 2009) and Constellation Vision (July 2009–June 2010) systems. Data were limited to a single surgeon and evaluated for operating room (OR) total time usage/day, OR case time/case, and OR surgical time/case. Further analysis evaluated all patients undergoing combined MIVS and clear cornea phacoemulsification/intraocular lens (IOL) implantation during each individual time period to determine the impact of the instrumentation on these parameters. All records were evaluated for intraoperative complications. RESULTS: Five hundred and fourteen eligible patients underwent MIVS during the 2-year study windows, with 281 patients undergoing surgery with the Accurus system and 233 patients undergoing surgery with the Constellation system. Combined MIVS and phacoemulsification with IOL implantation was performed 141 times during this period with the Accurus and 158 times during the second study period with the Constellation. Total number of patients operated per day increased from 7.55 with Accurus to 8.53 with Constellation. Surgical room time decreased from 56 minutes with Accurus to 52 minutes with Constellation, and procedure time decreased from 35 minutes with Accurus to 31 minutes with Constellation (P < 0.004). Combined MIVS/phacoemulsification surgery saw similar declines in surgical room time and procedure time (P < 0.001). Subset analysis of procedures limited by case number per day (eg, four cases/day, five cases/day, six cases/day, and seven or more cases/day) showed similar outcomes with a decrease in surgical room time and procedure time. No increases in surgery-related complications were noted by quality assurance review during these time periods. DISCUSSION: Transitioning to advanced surgical technology is a complex issue for the surgeon, the hospital team, and the hospital administration. This study documents improvement in three significant measures of surgical efficiency: operative number of patients per day, operative room time, and surgical procedure time that reflect the positive impact of the novel, combined, integrated, posterior and anterior, ophthalmologic surgical platform of the Constellation Vision System. These data are imperative to evaluate the impact of transition from one surgical platform to another. During this transition, hospital quality assurance review and surgeon evaluation of operative complications showed no increased concerns for the shift from the Accurus to the Constellation Vision System surgical platform. Further, both operative staff and surgeons felt that the transition to the Constellation was not associated with increases in difficulty with setup, turnover, or use and that the Constellation decreased safety concerns for surgical usage. Ultimately, in this case, new technology benefited the surgeon, the patient, and the hospital. |
format | Online Article Text |
id | pubmed-3581355 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Dove Medical Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-35813552013-02-28 Transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the Accurus vitrectomy platform to the Constellation Vision System for microincisional vitrectomy surgery Murray, Timothy G Layton, Andrew J Tong, Kuo B Gittelman, Michael Latiff, Azeema Gologorsky, Daniel Vigoda, Michael M Clin Ophthalmol Original Research BACKGROUND: Microincisional vitrectomy surgery (MIVS) is the current standard surgical approach for pars plana vitrectomy. Historically, the most common surgical platform for vitrectomy surgery, since its introduction in 1997, has been the Accurus vitrectomy system. Recent introduction of the next generation of vitrectomy platforms has generated concerns associated with transitioning to new technology in the operating room environment. This study compared, in a matched fashion, surgical use of the Accurus vitrectomy system and the next generation Constellation Vision System to evaluate surgical efficiencies, complications, and user perceptions of this transition. METHODS: Electronic health records were abstracted as a hospital quality assurance activity and included all vitreoretinal surgical procedures at the Bascom Palmer Eye Institute, Anne Bates Leach Eye Hospital, during two discrete 12-month time periods. These two periods reflected dedicated usage of the Accurus (June 2008–May 2009) and Constellation Vision (July 2009–June 2010) systems. Data were limited to a single surgeon and evaluated for operating room (OR) total time usage/day, OR case time/case, and OR surgical time/case. Further analysis evaluated all patients undergoing combined MIVS and clear cornea phacoemulsification/intraocular lens (IOL) implantation during each individual time period to determine the impact of the instrumentation on these parameters. All records were evaluated for intraoperative complications. RESULTS: Five hundred and fourteen eligible patients underwent MIVS during the 2-year study windows, with 281 patients undergoing surgery with the Accurus system and 233 patients undergoing surgery with the Constellation system. Combined MIVS and phacoemulsification with IOL implantation was performed 141 times during this period with the Accurus and 158 times during the second study period with the Constellation. Total number of patients operated per day increased from 7.55 with Accurus to 8.53 with Constellation. Surgical room time decreased from 56 minutes with Accurus to 52 minutes with Constellation, and procedure time decreased from 35 minutes with Accurus to 31 minutes with Constellation (P < 0.004). Combined MIVS/phacoemulsification surgery saw similar declines in surgical room time and procedure time (P < 0.001). Subset analysis of procedures limited by case number per day (eg, four cases/day, five cases/day, six cases/day, and seven or more cases/day) showed similar outcomes with a decrease in surgical room time and procedure time. No increases in surgery-related complications were noted by quality assurance review during these time periods. DISCUSSION: Transitioning to advanced surgical technology is a complex issue for the surgeon, the hospital team, and the hospital administration. This study documents improvement in three significant measures of surgical efficiency: operative number of patients per day, operative room time, and surgical procedure time that reflect the positive impact of the novel, combined, integrated, posterior and anterior, ophthalmologic surgical platform of the Constellation Vision System. These data are imperative to evaluate the impact of transition from one surgical platform to another. During this transition, hospital quality assurance review and surgeon evaluation of operative complications showed no increased concerns for the shift from the Accurus to the Constellation Vision System surgical platform. Further, both operative staff and surgeons felt that the transition to the Constellation was not associated with increases in difficulty with setup, turnover, or use and that the Constellation decreased safety concerns for surgical usage. Ultimately, in this case, new technology benefited the surgeon, the patient, and the hospital. Dove Medical Press 2013 2013-02-19 /pmc/articles/PMC3581355/ /pubmed/23450081 http://dx.doi.org/10.2147/OPTH.S35603 Text en © 2013 Murray et al, publisher and licensee Dove Medical Press Ltd. This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited. |
spellingShingle | Original Research Murray, Timothy G Layton, Andrew J Tong, Kuo B Gittelman, Michael Latiff, Azeema Gologorsky, Daniel Vigoda, Michael M Transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the Accurus vitrectomy platform to the Constellation Vision System for microincisional vitrectomy surgery |
title | Transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the Accurus vitrectomy platform to the Constellation Vision System for microincisional vitrectomy surgery |
title_full | Transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the Accurus vitrectomy platform to the Constellation Vision System for microincisional vitrectomy surgery |
title_fullStr | Transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the Accurus vitrectomy platform to the Constellation Vision System for microincisional vitrectomy surgery |
title_full_unstemmed | Transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the Accurus vitrectomy platform to the Constellation Vision System for microincisional vitrectomy surgery |
title_short | Transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the Accurus vitrectomy platform to the Constellation Vision System for microincisional vitrectomy surgery |
title_sort | transition to a novel advanced integrated vitrectomy platform: comparison of the surgical impact of moving from the accurus vitrectomy platform to the constellation vision system for microincisional vitrectomy surgery |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3581355/ https://www.ncbi.nlm.nih.gov/pubmed/23450081 http://dx.doi.org/10.2147/OPTH.S35603 |
work_keys_str_mv | AT murraytimothyg transitiontoanoveladvancedintegratedvitrectomyplatformcomparisonofthesurgicalimpactofmovingfromtheaccurusvitrectomyplatformtotheconstellationvisionsystemformicroincisionalvitrectomysurgery AT laytonandrewj transitiontoanoveladvancedintegratedvitrectomyplatformcomparisonofthesurgicalimpactofmovingfromtheaccurusvitrectomyplatformtotheconstellationvisionsystemformicroincisionalvitrectomysurgery AT tongkuob transitiontoanoveladvancedintegratedvitrectomyplatformcomparisonofthesurgicalimpactofmovingfromtheaccurusvitrectomyplatformtotheconstellationvisionsystemformicroincisionalvitrectomysurgery AT gittelmanmichael transitiontoanoveladvancedintegratedvitrectomyplatformcomparisonofthesurgicalimpactofmovingfromtheaccurusvitrectomyplatformtotheconstellationvisionsystemformicroincisionalvitrectomysurgery AT latiffazeema transitiontoanoveladvancedintegratedvitrectomyplatformcomparisonofthesurgicalimpactofmovingfromtheaccurusvitrectomyplatformtotheconstellationvisionsystemformicroincisionalvitrectomysurgery AT gologorskydaniel transitiontoanoveladvancedintegratedvitrectomyplatformcomparisonofthesurgicalimpactofmovingfromtheaccurusvitrectomyplatformtotheconstellationvisionsystemformicroincisionalvitrectomysurgery AT vigodamichaelm transitiontoanoveladvancedintegratedvitrectomyplatformcomparisonofthesurgicalimpactofmovingfromtheaccurusvitrectomyplatformtotheconstellationvisionsystemformicroincisionalvitrectomysurgery |