Cargando…

Braile Vena Cava Filter and Greenfield Filter in Terms of Centralization

The aim of this study was to evaluate complications experienced during implantation of the Braile Vena Cava filter (VCF) and the efficacy of the centralization mechanism of the filter. This retrospective cohort study evaluated all Braile Biomédica VCFs implanted from 2004 to 2009 in Hospital de Base...

Descripción completa

Detalles Bibliográficos
Autores principales: de Godoy, José Maria Pereira, Menezes da Silva, Adinaldo A, Reis, Luis Fernando, Miquelin, Daniel, Torati, José Luis Simon
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Bentham Open 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3582014/
https://www.ncbi.nlm.nih.gov/pubmed/23459189
http://dx.doi.org/10.2174/1874192401307010009
Descripción
Sumario:The aim of this study was to evaluate complications experienced during implantation of the Braile Vena Cava filter (VCF) and the efficacy of the centralization mechanism of the filter. This retrospective cohort study evaluated all Braile Biomédica VCFs implanted from 2004 to 2009 in Hospital de Base Medicine School in São José do Rio Preto, Brazil. Of particular concern was the filter’s symmetry during implantation and complications experienced during the procedure. All the angiographic examinations performed during the implantation of the filters were analyzed in respect to the following parameters: migration of the filter, non-opening or difficulties in the implantation and centralization of the filter. A total of 112 Braile CVFs were implanted and there were no reports of filter opening difficulties or in respect to migration. Asymmetry was observed in 1/112 (0.9%) cases. A statistically significant difference was seen on comparing historical data on decentralization of the Greenfield filter with the data of this study. The Braile Biomédico filter is an evolution of the Greenfield filter providing improved embolus capture and better implantation symmetry.