Long-term bladeless LASIK outcomes with the FS200 Femtosecond and EX500 Excimer Laser workstation: the Refractive Suite

PURPOSE: The evaluation of the safety, efficacy, and long-term stability of LASIK procedures utilizing novel platform comprising a femtosecond and excimer laser and multiple networked diagnostics. SETTING: Private clinical ophthalmology practice. PATIENTS AND METHODS: In consecutive cases of myopic...

Descripción completa

Detalles Bibliográficos
Autores principales: Kanellopoulos, A John, Asimellis, George
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3583408/
https://www.ncbi.nlm.nih.gov/pubmed/23459015
http://dx.doi.org/10.2147/OPTH.S40454
_version_ 1782475421398335488
author Kanellopoulos, A John
Asimellis, George
author_facet Kanellopoulos, A John
Asimellis, George
author_sort Kanellopoulos, A John
collection PubMed
description PURPOSE: The evaluation of the safety, efficacy, and long-term stability of LASIK procedures utilizing novel platform comprising a femtosecond and excimer laser and multiple networked diagnostics. SETTING: Private clinical ophthalmology practice. PATIENTS AND METHODS: In consecutive cases of myopic LASIK procedure with a novel refractive platform (FS200 Femtosecond and EX500 Excimer Laser), 190 eyes (from 109 different patients) were evaluated pre- and postoperatively for the following parameters: refractive error, best corrected distance visual acuity, uncorrected distance visual acuity, topography (Placido-disc based) and tomography (Scheimpflug-image based), wavefront analysis, pupillometry, and contrast sensitivity. Follow-up visits were conducted for at least 12 months. RESULTS: The change from pre- to postoperative mean refractive error was from −5.29 ± 2.39 diopters (D) (range −8.0 to −0.50 D) to −0.27 ± 0.09 D at the 3-month visit, −0.27 ± 0.10 D at the 6-month visit, and −0.39 ± 0.08 D at the 1-year visit. The change from pre- to postoperative refractive astigmatism was −1.07 ± 0.91 D (range −4.25 to 0 D) to −0.14 ± 0.04 D at 3 months, −0.15 ± 0.04 at 6 months, and −0.16 ± 0.04 at the 1-year visit. The proportion of the eyes with postoperative astigmatism within 0.5 D ranged between 95.6% and 99%. The proportion of eyes achieving uncorrected distance visual acuity of 1.0 (decimal) was 93.0%. CONCLUSION: The myopic LASIK clinical results with the FS200 Femtosecond Laser and EX500 Excimer Laser showed outstanding efficacy, great safety, and long-term stability.
format Online
Article
Text
id pubmed-3583408
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-35834082013-03-04 Long-term bladeless LASIK outcomes with the FS200 Femtosecond and EX500 Excimer Laser workstation: the Refractive Suite Kanellopoulos, A John Asimellis, George Clin Ophthalmol Original Research PURPOSE: The evaluation of the safety, efficacy, and long-term stability of LASIK procedures utilizing novel platform comprising a femtosecond and excimer laser and multiple networked diagnostics. SETTING: Private clinical ophthalmology practice. PATIENTS AND METHODS: In consecutive cases of myopic LASIK procedure with a novel refractive platform (FS200 Femtosecond and EX500 Excimer Laser), 190 eyes (from 109 different patients) were evaluated pre- and postoperatively for the following parameters: refractive error, best corrected distance visual acuity, uncorrected distance visual acuity, topography (Placido-disc based) and tomography (Scheimpflug-image based), wavefront analysis, pupillometry, and contrast sensitivity. Follow-up visits were conducted for at least 12 months. RESULTS: The change from pre- to postoperative mean refractive error was from −5.29 ± 2.39 diopters (D) (range −8.0 to −0.50 D) to −0.27 ± 0.09 D at the 3-month visit, −0.27 ± 0.10 D at the 6-month visit, and −0.39 ± 0.08 D at the 1-year visit. The change from pre- to postoperative refractive astigmatism was −1.07 ± 0.91 D (range −4.25 to 0 D) to −0.14 ± 0.04 D at 3 months, −0.15 ± 0.04 at 6 months, and −0.16 ± 0.04 at the 1-year visit. The proportion of the eyes with postoperative astigmatism within 0.5 D ranged between 95.6% and 99%. The proportion of eyes achieving uncorrected distance visual acuity of 1.0 (decimal) was 93.0%. CONCLUSION: The myopic LASIK clinical results with the FS200 Femtosecond Laser and EX500 Excimer Laser showed outstanding efficacy, great safety, and long-term stability. Dove Medical Press 2013 2013-02-21 /pmc/articles/PMC3583408/ /pubmed/23459015 http://dx.doi.org/10.2147/OPTH.S40454 Text en © 2013 Kanellopoulos and Asimellis, publisher and licensee Dove Medical Press Ltd. This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited.
spellingShingle Original Research
Kanellopoulos, A John
Asimellis, George
Long-term bladeless LASIK outcomes with the FS200 Femtosecond and EX500 Excimer Laser workstation: the Refractive Suite
title Long-term bladeless LASIK outcomes with the FS200 Femtosecond and EX500 Excimer Laser workstation: the Refractive Suite
title_full Long-term bladeless LASIK outcomes with the FS200 Femtosecond and EX500 Excimer Laser workstation: the Refractive Suite
title_fullStr Long-term bladeless LASIK outcomes with the FS200 Femtosecond and EX500 Excimer Laser workstation: the Refractive Suite
title_full_unstemmed Long-term bladeless LASIK outcomes with the FS200 Femtosecond and EX500 Excimer Laser workstation: the Refractive Suite
title_short Long-term bladeless LASIK outcomes with the FS200 Femtosecond and EX500 Excimer Laser workstation: the Refractive Suite
title_sort long-term bladeless lasik outcomes with the fs200 femtosecond and ex500 excimer laser workstation: the refractive suite
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3583408/
https://www.ncbi.nlm.nih.gov/pubmed/23459015
http://dx.doi.org/10.2147/OPTH.S40454
work_keys_str_mv AT kanellopoulosajohn longtermbladelesslasikoutcomeswiththefs200femtosecondandex500excimerlaserworkstationtherefractivesuite
AT asimellisgeorge longtermbladelesslasikoutcomeswiththefs200femtosecondandex500excimerlaserworkstationtherefractivesuite