Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound

BACKGROUND: The purpose of this study was to compare the accuracy of the new Sirius(®) Scheimpflug anterior segment examination device for measurement of central corneal thickness (CCT) and anterior chamber depth (ACD) with that of CCT measurements obtained by ultrasound pachymetry and ACD measureme...

Descripción completa

Detalles Bibliográficos
Autores principales: Jorge, J, Rosado, JL, Díaz-Rey, JA, González-Méijome, JM
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3587393/
https://www.ncbi.nlm.nih.gov/pubmed/23467857
http://dx.doi.org/10.2147/OPTH.S35121
_version_ 1782261394577555456
author Jorge, J
Rosado, JL
Díaz-Rey, JA
González-Méijome, JM
author_facet Jorge, J
Rosado, JL
Díaz-Rey, JA
González-Méijome, JM
author_sort Jorge, J
collection PubMed
description BACKGROUND: The purpose of this study was to compare the accuracy of the new Sirius(®) Scheimpflug anterior segment examination device for measurement of central corneal thickness (CCT) and anterior chamber depth (ACD) with that of CCT measurements obtained by ultrasound pachymetry and ACD measurements obtained by ultrasound biometry, respectively. METHODS: CCT and ACD was measured in 50 right eyes from 50 healthy subjects using a Sirius Scheimpflug camera, SP100 ultrasound pachymetry, and US800 ultrasound biometry. RESULTS: CCT measured with the Sirius was 546 ± 39 μm and 541 ± 35 μm with SP100 ultrasound pachymetry (P = 0.003). The difference was statistically significant (mean difference 4.68 ± 10.5 μm; limits of agreement −15.8 to 25.20 μm). ACD measured with the Sirius was 2.96 ± 0.3 mm compared with 3.36 ± 0.29 mm using US800 ultrasound biometry (P < 0.001). The difference was statistically significant (mean difference −0.40 ± 0.16 mm; limits of agreement −0.72 to 0.07 mm). When the ACD values obtained using ultrasound biometry were corrected according to the values for CCT measured by ultrasound, the agreement increased significantly between both technologies for ACD measurements (mean difference 0.15 ± 0.16 mm; limits of agreement −0.16 to 0.45 mm). CONCLUSION: CCT and ACD measured by Sirius and ultrasound methods showing good agreement between repeated measurements obtained in the same subjects (repeatability) with either instrument. However, CCT and ACD values, even after correcting ultrasound ACD by subtracting the CCT value obtained with either technology should not be used interchangeably.
format Online
Article
Text
id pubmed-3587393
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-35873932013-03-06 Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound Jorge, J Rosado, JL Díaz-Rey, JA González-Méijome, JM Clin Ophthalmol Original Research BACKGROUND: The purpose of this study was to compare the accuracy of the new Sirius(®) Scheimpflug anterior segment examination device for measurement of central corneal thickness (CCT) and anterior chamber depth (ACD) with that of CCT measurements obtained by ultrasound pachymetry and ACD measurements obtained by ultrasound biometry, respectively. METHODS: CCT and ACD was measured in 50 right eyes from 50 healthy subjects using a Sirius Scheimpflug camera, SP100 ultrasound pachymetry, and US800 ultrasound biometry. RESULTS: CCT measured with the Sirius was 546 ± 39 μm and 541 ± 35 μm with SP100 ultrasound pachymetry (P = 0.003). The difference was statistically significant (mean difference 4.68 ± 10.5 μm; limits of agreement −15.8 to 25.20 μm). ACD measured with the Sirius was 2.96 ± 0.3 mm compared with 3.36 ± 0.29 mm using US800 ultrasound biometry (P < 0.001). The difference was statistically significant (mean difference −0.40 ± 0.16 mm; limits of agreement −0.72 to 0.07 mm). When the ACD values obtained using ultrasound biometry were corrected according to the values for CCT measured by ultrasound, the agreement increased significantly between both technologies for ACD measurements (mean difference 0.15 ± 0.16 mm; limits of agreement −0.16 to 0.45 mm). CONCLUSION: CCT and ACD measured by Sirius and ultrasound methods showing good agreement between repeated measurements obtained in the same subjects (repeatability) with either instrument. However, CCT and ACD values, even after correcting ultrasound ACD by subtracting the CCT value obtained with either technology should not be used interchangeably. Dove Medical Press 2013 2013-02-26 /pmc/articles/PMC3587393/ /pubmed/23467857 http://dx.doi.org/10.2147/OPTH.S35121 Text en © 2013 Jorge et al, publisher and licensee Dove Medical Press Ltd. This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited.
spellingShingle Original Research
Jorge, J
Rosado, JL
Díaz-Rey, JA
González-Méijome, JM
Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound
title Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound
title_full Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound
title_fullStr Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound
title_full_unstemmed Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound
title_short Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound
title_sort central corneal thickness and anterior chamber depth measurement by sirius(®) scheimpflug tomography and ultrasound
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3587393/
https://www.ncbi.nlm.nih.gov/pubmed/23467857
http://dx.doi.org/10.2147/OPTH.S35121
work_keys_str_mv AT jorgej centralcornealthicknessandanteriorchamberdepthmeasurementbysiriusscheimpflugtomographyandultrasound
AT rosadojl centralcornealthicknessandanteriorchamberdepthmeasurementbysiriusscheimpflugtomographyandultrasound
AT diazreyja centralcornealthicknessandanteriorchamberdepthmeasurementbysiriusscheimpflugtomographyandultrasound
AT gonzalezmeijomejm centralcornealthicknessandanteriorchamberdepthmeasurementbysiriusscheimpflugtomographyandultrasound