Intra- and Inter-Instrument Reliability of the Actiwatch 4 Accelerometer in a Mechanical Laboratory Setting
This study aimed to quantify the intra-and inter-instrument reliability of the Actiwatch 4 accelerometer (AW4) in a mechanical setting. Twenty seven AW4 were attached to an isokinetic dynamometer and subjected to angular acceleration for 30 min at 50 deg/sec representing moderate intensity (MPA cond...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Akademia Wychowania Fizycznego w Katowicach
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3588665/ https://www.ncbi.nlm.nih.gov/pubmed/23487398 http://dx.doi.org/10.2478/v10078-012-0002-z |
_version_ | 1782261608716697600 |
---|---|
author | Routen, Ashley C. Upton, Dominic Edwards, Martin G. Peters, Derek M. |
author_facet | Routen, Ashley C. Upton, Dominic Edwards, Martin G. Peters, Derek M. |
author_sort | Routen, Ashley C. |
collection | PubMed |
description | This study aimed to quantify the intra-and inter-instrument reliability of the Actiwatch 4 accelerometer (AW4) in a mechanical setting. Twenty seven AW4 were attached to an isokinetic dynamometer and subjected to angular acceleration for 30 min at 50 deg/sec representing moderate intensity (MPA condition) and 200 deg/sec representing vigorous intensity (VPA condition), with a repeat trial conducted. Reliability was assessed using coefficient of variation (CV), absolute percent error (APE), and intraclass correlation coefficients (ICC). Mean AW4 activity counts displayed acceptable reliability according to CV in both conditions (ConMPA: CVintra = 4.6%, APEintra = 6.6%, CVinter = 6.4%, APEinter = 5.2%; ConVPA: CVintra = 3.9%, APEintra = 5.6%, CVinter = 5.9%, APEinter = 4.7%). When counts were re-categorised into minutes of MPA and VPA, lower CV values were observed (ConMPA: CVintra = 3.2%, APEintra = 4.5%, CVinter = 4.3%, APEinter = 3.2%; ConVPA: CVintra = 0.0%, APEintra = 0.0%, CVinter = 0.0%, APEinter = 0.0%). When activity counts were re-categorised as minutes of MVPA, excellent reliability was observed (CVintra, APEintra, CVinter, and APEinter = 0.0%) in both conditions. Mean AW4 activity counts exhibit reliability statistics comparable to other accelerometers. Reliability is improved when activity counts are re-categorised as time spent in MPA and VPA, with greatest reliability obtained when counts are recategorised as time spent in MVPA. As MVPA is the subcomponent of physical activity most associated with health benefits it would appear that the AW4 is reliable for measuring time spent in this health enhancing intensity category, at least from testing in a mechanical laboratory setting. |
format | Online Article Text |
id | pubmed-3588665 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Akademia Wychowania Fizycznego w Katowicach |
record_format | MEDLINE/PubMed |
spelling | pubmed-35886652013-03-13 Intra- and Inter-Instrument Reliability of the Actiwatch 4 Accelerometer in a Mechanical Laboratory Setting Routen, Ashley C. Upton, Dominic Edwards, Martin G. Peters, Derek M. J Hum Kinet Research Article This study aimed to quantify the intra-and inter-instrument reliability of the Actiwatch 4 accelerometer (AW4) in a mechanical setting. Twenty seven AW4 were attached to an isokinetic dynamometer and subjected to angular acceleration for 30 min at 50 deg/sec representing moderate intensity (MPA condition) and 200 deg/sec representing vigorous intensity (VPA condition), with a repeat trial conducted. Reliability was assessed using coefficient of variation (CV), absolute percent error (APE), and intraclass correlation coefficients (ICC). Mean AW4 activity counts displayed acceptable reliability according to CV in both conditions (ConMPA: CVintra = 4.6%, APEintra = 6.6%, CVinter = 6.4%, APEinter = 5.2%; ConVPA: CVintra = 3.9%, APEintra = 5.6%, CVinter = 5.9%, APEinter = 4.7%). When counts were re-categorised into minutes of MPA and VPA, lower CV values were observed (ConMPA: CVintra = 3.2%, APEintra = 4.5%, CVinter = 4.3%, APEinter = 3.2%; ConVPA: CVintra = 0.0%, APEintra = 0.0%, CVinter = 0.0%, APEinter = 0.0%). When activity counts were re-categorised as minutes of MVPA, excellent reliability was observed (CVintra, APEintra, CVinter, and APEinter = 0.0%) in both conditions. Mean AW4 activity counts exhibit reliability statistics comparable to other accelerometers. Reliability is improved when activity counts are re-categorised as time spent in MPA and VPA, with greatest reliability obtained when counts are recategorised as time spent in MVPA. As MVPA is the subcomponent of physical activity most associated with health benefits it would appear that the AW4 is reliable for measuring time spent in this health enhancing intensity category, at least from testing in a mechanical laboratory setting. Akademia Wychowania Fizycznego w Katowicach 2012-04-03 /pmc/articles/PMC3588665/ /pubmed/23487398 http://dx.doi.org/10.2478/v10078-012-0002-z Text en © Editorial Committee of Journal of Human Kinetics http://creativecommons.org/licenses/by/3.0 This article is an open-access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/). |
spellingShingle | Research Article Routen, Ashley C. Upton, Dominic Edwards, Martin G. Peters, Derek M. Intra- and Inter-Instrument Reliability of the Actiwatch 4 Accelerometer in a Mechanical Laboratory Setting |
title | Intra- and Inter-Instrument Reliability of the Actiwatch 4 Accelerometer in a Mechanical Laboratory Setting |
title_full | Intra- and Inter-Instrument Reliability of the Actiwatch 4 Accelerometer in a Mechanical Laboratory Setting |
title_fullStr | Intra- and Inter-Instrument Reliability of the Actiwatch 4 Accelerometer in a Mechanical Laboratory Setting |
title_full_unstemmed | Intra- and Inter-Instrument Reliability of the Actiwatch 4 Accelerometer in a Mechanical Laboratory Setting |
title_short | Intra- and Inter-Instrument Reliability of the Actiwatch 4 Accelerometer in a Mechanical Laboratory Setting |
title_sort | intra- and inter-instrument reliability of the actiwatch 4 accelerometer in a mechanical laboratory setting |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3588665/ https://www.ncbi.nlm.nih.gov/pubmed/23487398 http://dx.doi.org/10.2478/v10078-012-0002-z |
work_keys_str_mv | AT routenashleyc intraandinterinstrumentreliabilityoftheactiwatch4accelerometerinamechanicallaboratorysetting AT uptondominic intraandinterinstrumentreliabilityoftheactiwatch4accelerometerinamechanicallaboratorysetting AT edwardsmarting intraandinterinstrumentreliabilityoftheactiwatch4accelerometerinamechanicallaboratorysetting AT petersderekm intraandinterinstrumentreliabilityoftheactiwatch4accelerometerinamechanicallaboratorysetting |