Cargando…

Central Corneal Thickness Measurement by Ultrasound versus Orbscan II

PURPOSE: To compare Orbscan II and ultrasonic pachymetry for measurement of central corneal thickness (CCT) in eyes scheduled for keratorefractive surgery. METHODS: CCT was measured using Orbscan II (Bausch & Lomb, USA) and then by ultrasonic pachymetry (Tomey SP-3000, Tomey Ltd, Japan) in 100 e...

Descripción completa

Detalles Bibliográficos
Autores principales: Faramarzi, Amir, Ziai, Hossein
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Ophthalmic Research Center 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3589223/
https://www.ncbi.nlm.nih.gov/pubmed/23479527
Descripción
Sumario:PURPOSE: To compare Orbscan II and ultrasonic pachymetry for measurement of central corneal thickness (CCT) in eyes scheduled for keratorefractive surgery. METHODS: CCT was measured using Orbscan II (Bausch & Lomb, USA) and then by ultrasonic pachymetry (Tomey SP-3000, Tomey Ltd, Japan) in 100 eyes of 100 patients with no history of ocular surgery scheduled for excimer laser refractive surgery. RESULTS: Mean CCT was 544.7±35.5 (range 453–637) μm by ultrasonic pachymetry versus 546.9±41.6 (range 435–648) μm measured by Orbscan II applying an acoustic factor of 0.92 (P=0.14). The standard deviation of measurements was greater with Orbscan pachymetry but the difference was not statistically significant. CONCLUSION: CCT measurements by Orbscan II (applying an acoustic factor) and by ultrasonic pachymetry are not significantly different; however, when CCT readings by Orbscan II are in the lower range, it is advisable to recheck the measurements using ultrasonic pachymetry.