Central Corneal Thickness Measurement by Ultrasound versus Orbscan II
PURPOSE: To compare Orbscan II and ultrasonic pachymetry for measurement of central corneal thickness (CCT) in eyes scheduled for keratorefractive surgery. METHODS: CCT was measured using Orbscan II (Bausch & Lomb, USA) and then by ultrasonic pachymetry (Tomey SP-3000, Tomey Ltd, Japan) in 100 e...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Ophthalmic Research Center
2008
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3589223/ https://www.ncbi.nlm.nih.gov/pubmed/23479527 |
_version_ | 1782261700962025472 |
---|---|
author | Faramarzi, Amir Ziai, Hossein |
author_facet | Faramarzi, Amir Ziai, Hossein |
author_sort | Faramarzi, Amir |
collection | PubMed |
description | PURPOSE: To compare Orbscan II and ultrasonic pachymetry for measurement of central corneal thickness (CCT) in eyes scheduled for keratorefractive surgery. METHODS: CCT was measured using Orbscan II (Bausch & Lomb, USA) and then by ultrasonic pachymetry (Tomey SP-3000, Tomey Ltd, Japan) in 100 eyes of 100 patients with no history of ocular surgery scheduled for excimer laser refractive surgery. RESULTS: Mean CCT was 544.7±35.5 (range 453–637) μm by ultrasonic pachymetry versus 546.9±41.6 (range 435–648) μm measured by Orbscan II applying an acoustic factor of 0.92 (P=0.14). The standard deviation of measurements was greater with Orbscan pachymetry but the difference was not statistically significant. CONCLUSION: CCT measurements by Orbscan II (applying an acoustic factor) and by ultrasonic pachymetry are not significantly different; however, when CCT readings by Orbscan II are in the lower range, it is advisable to recheck the measurements using ultrasonic pachymetry. |
format | Online Article Text |
id | pubmed-3589223 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2008 |
publisher | Ophthalmic Research Center |
record_format | MEDLINE/PubMed |
spelling | pubmed-35892232013-03-11 Central Corneal Thickness Measurement by Ultrasound versus Orbscan II Faramarzi, Amir Ziai, Hossein J Ophthalmic Vis Res Original Article PURPOSE: To compare Orbscan II and ultrasonic pachymetry for measurement of central corneal thickness (CCT) in eyes scheduled for keratorefractive surgery. METHODS: CCT was measured using Orbscan II (Bausch & Lomb, USA) and then by ultrasonic pachymetry (Tomey SP-3000, Tomey Ltd, Japan) in 100 eyes of 100 patients with no history of ocular surgery scheduled for excimer laser refractive surgery. RESULTS: Mean CCT was 544.7±35.5 (range 453–637) μm by ultrasonic pachymetry versus 546.9±41.6 (range 435–648) μm measured by Orbscan II applying an acoustic factor of 0.92 (P=0.14). The standard deviation of measurements was greater with Orbscan pachymetry but the difference was not statistically significant. CONCLUSION: CCT measurements by Orbscan II (applying an acoustic factor) and by ultrasonic pachymetry are not significantly different; however, when CCT readings by Orbscan II are in the lower range, it is advisable to recheck the measurements using ultrasonic pachymetry. Ophthalmic Research Center 2008-04 /pmc/articles/PMC3589223/ /pubmed/23479527 Text en © 2008 Ophthalmic Research Center, Shahid Beheshti University of Medical Sciences http://creativecommons.org/licenses/by-nc/3.0/ This work is licensed under a Creative Commons Attribution-NonCommercial 3.0 Unported License which allows users to read, copy, distribute and make derivative works for non-commercial purposes from the material, as long as the author of the original work is cited properly. |
spellingShingle | Original Article Faramarzi, Amir Ziai, Hossein Central Corneal Thickness Measurement by Ultrasound versus Orbscan II |
title | Central Corneal Thickness Measurement by Ultrasound versus Orbscan II |
title_full | Central Corneal Thickness Measurement by Ultrasound versus Orbscan II |
title_fullStr | Central Corneal Thickness Measurement by Ultrasound versus Orbscan II |
title_full_unstemmed | Central Corneal Thickness Measurement by Ultrasound versus Orbscan II |
title_short | Central Corneal Thickness Measurement by Ultrasound versus Orbscan II |
title_sort | central corneal thickness measurement by ultrasound versus orbscan ii |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3589223/ https://www.ncbi.nlm.nih.gov/pubmed/23479527 |
work_keys_str_mv | AT faramarziamir centralcornealthicknessmeasurementbyultrasoundversusorbscanii AT ziaihossein centralcornealthicknessmeasurementbyultrasoundversusorbscanii |