Evaluation of the In Vivo Efficiency and Safety of Hepatic Radiofrequency Ablation Using a 15-G Octopus® in Pig Liver

OBJECTIVE: To determine in vivo efficacy of radiofrequency ablation (RFA) in porcine liver by using 15-gauge Octopus® (15-G Octopus®) electrodes to create a large coagulation. MATERIALS AND METHODS: A total of 18 coagulations were created by using a 180-W generator and 15-G Octopus® electrodes durin...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Eun Sun, Lee, Jeong Min, Kim, Kyung Won, Lee, In Joon, Han, Joon Koo, Choi, Byung Ihn
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Society of Radiology 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3590330/
https://www.ncbi.nlm.nih.gov/pubmed/23482995
http://dx.doi.org/10.3348/kjr.2013.14.2.194
_version_ 1782261845348843520
author Lee, Eun Sun
Lee, Jeong Min
Kim, Kyung Won
Lee, In Joon
Han, Joon Koo
Choi, Byung Ihn
author_facet Lee, Eun Sun
Lee, Jeong Min
Kim, Kyung Won
Lee, In Joon
Han, Joon Koo
Choi, Byung Ihn
author_sort Lee, Eun Sun
collection PubMed
description OBJECTIVE: To determine in vivo efficacy of radiofrequency ablation (RFA) in porcine liver by using 15-gauge Octopus® (15-G Octopus®) electrodes to create a large coagulation. MATERIALS AND METHODS: A total of 18 coagulations were created by using a 180-W generator and 15-G Octopus® electrodes during laparotomy, performed in 14 pigs. Coagulation necrosis was created in the pig livers by the use of one of three RFA protocols: 1) group A, monopolar RFA using a 15-G Octopus® electrode with a 5-mm inter-electrode distance (n = 4); 2) group B, monopolar RFA using a 15-G Octopus® electrode with a 10-mm inter-electrode distance (n = 6); and 3) group C, switching monopolar RFA using two 15-G Octopus® electrodes (n = 8). The energy efficiency, shape, maximum and minimum diameters (Dmx and Dmi), and the volume of the coagulation volume were measured in each group. The Summary statistics were obtained and Mann-Whitney test was were performed. RESULTS: The mean ablated volume of each group was 49.23 cm(3) in A, 64.11 cm(3) in B, and 72.35 cm(3) in C. The mean Dmx and Dmi values were 5.68 cm and 4.58 cm in A and 5.97 cm and 4.97 cm in B, respectively. In group C, the mean diameters of Dmx and Dmi were 6.80 cm and 5.11 cm, respectively. The mean ratios of Dmi/Dmx were 1.25, 1.20, and 1.35 in groups A, B, and C, respectively. There was one animal death during the RFA procedure, the cause of which could not be subsequently determined. However, there were no other significant, procedure-related complications during the seven-hour-delayed CT scans. CONCLUSION: RFA procedures using 15-G Octopus® electrodes are useful and safe for creating a large ablation in a single electrode model as well as in the multiple electrodes model.
format Online
Article
Text
id pubmed-3590330
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher The Korean Society of Radiology
record_format MEDLINE/PubMed
spelling pubmed-35903302013-03-12 Evaluation of the In Vivo Efficiency and Safety of Hepatic Radiofrequency Ablation Using a 15-G Octopus® in Pig Liver Lee, Eun Sun Lee, Jeong Min Kim, Kyung Won Lee, In Joon Han, Joon Koo Choi, Byung Ihn Korean J Radiol Experimental and Others OBJECTIVE: To determine in vivo efficacy of radiofrequency ablation (RFA) in porcine liver by using 15-gauge Octopus® (15-G Octopus®) electrodes to create a large coagulation. MATERIALS AND METHODS: A total of 18 coagulations were created by using a 180-W generator and 15-G Octopus® electrodes during laparotomy, performed in 14 pigs. Coagulation necrosis was created in the pig livers by the use of one of three RFA protocols: 1) group A, monopolar RFA using a 15-G Octopus® electrode with a 5-mm inter-electrode distance (n = 4); 2) group B, monopolar RFA using a 15-G Octopus® electrode with a 10-mm inter-electrode distance (n = 6); and 3) group C, switching monopolar RFA using two 15-G Octopus® electrodes (n = 8). The energy efficiency, shape, maximum and minimum diameters (Dmx and Dmi), and the volume of the coagulation volume were measured in each group. The Summary statistics were obtained and Mann-Whitney test was were performed. RESULTS: The mean ablated volume of each group was 49.23 cm(3) in A, 64.11 cm(3) in B, and 72.35 cm(3) in C. The mean Dmx and Dmi values were 5.68 cm and 4.58 cm in A and 5.97 cm and 4.97 cm in B, respectively. In group C, the mean diameters of Dmx and Dmi were 6.80 cm and 5.11 cm, respectively. The mean ratios of Dmi/Dmx were 1.25, 1.20, and 1.35 in groups A, B, and C, respectively. There was one animal death during the RFA procedure, the cause of which could not be subsequently determined. However, there were no other significant, procedure-related complications during the seven-hour-delayed CT scans. CONCLUSION: RFA procedures using 15-G Octopus® electrodes are useful and safe for creating a large ablation in a single electrode model as well as in the multiple electrodes model. The Korean Society of Radiology 2013 2013-02-22 /pmc/articles/PMC3590330/ /pubmed/23482995 http://dx.doi.org/10.3348/kjr.2013.14.2.194 Text en Copyright © 2013 The Korean Society of Radiology http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Experimental and Others
Lee, Eun Sun
Lee, Jeong Min
Kim, Kyung Won
Lee, In Joon
Han, Joon Koo
Choi, Byung Ihn
Evaluation of the In Vivo Efficiency and Safety of Hepatic Radiofrequency Ablation Using a 15-G Octopus® in Pig Liver
title Evaluation of the In Vivo Efficiency and Safety of Hepatic Radiofrequency Ablation Using a 15-G Octopus® in Pig Liver
title_full Evaluation of the In Vivo Efficiency and Safety of Hepatic Radiofrequency Ablation Using a 15-G Octopus® in Pig Liver
title_fullStr Evaluation of the In Vivo Efficiency and Safety of Hepatic Radiofrequency Ablation Using a 15-G Octopus® in Pig Liver
title_full_unstemmed Evaluation of the In Vivo Efficiency and Safety of Hepatic Radiofrequency Ablation Using a 15-G Octopus® in Pig Liver
title_short Evaluation of the In Vivo Efficiency and Safety of Hepatic Radiofrequency Ablation Using a 15-G Octopus® in Pig Liver
title_sort evaluation of the in vivo efficiency and safety of hepatic radiofrequency ablation using a 15-g octopus® in pig liver
topic Experimental and Others
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3590330/
https://www.ncbi.nlm.nih.gov/pubmed/23482995
http://dx.doi.org/10.3348/kjr.2013.14.2.194
work_keys_str_mv AT leeeunsun evaluationoftheinvivoefficiencyandsafetyofhepaticradiofrequencyablationusinga15goctopusinpigliver
AT leejeongmin evaluationoftheinvivoefficiencyandsafetyofhepaticradiofrequencyablationusinga15goctopusinpigliver
AT kimkyungwon evaluationoftheinvivoefficiencyandsafetyofhepaticradiofrequencyablationusinga15goctopusinpigliver
AT leeinjoon evaluationoftheinvivoefficiencyandsafetyofhepaticradiofrequencyablationusinga15goctopusinpigliver
AT hanjoonkoo evaluationoftheinvivoefficiencyandsafetyofhepaticradiofrequencyablationusinga15goctopusinpigliver
AT choibyungihn evaluationoftheinvivoefficiencyandsafetyofhepaticradiofrequencyablationusinga15goctopusinpigliver