The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing Cardiac Resynchronisation Therapy with Conventional Bipolar Leads

BACKGROUND: Phrenic nerve stimulation (PNS) and high pacing thresholds (HPT) hinder biventricular stimulation in patients (pts) undergoing cardiac resynchronization therapy (CRT). A new quadripolar left ventricular (LV) lead (Quartet 1458Q, St. Jude Medical) with increased number of pacing configura...

Descripción completa

Detalles Bibliográficos
Autores principales: Ohlow, Marc-Alexander, Lauer, Bernward, Brunelli, Michele, Daralammouri, Yunis, Geller, Christoph
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Indian Heart Rhythm Society 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3594899/
https://www.ncbi.nlm.nih.gov/pubmed/23573059
_version_ 1782262356743553024
author Ohlow, Marc-Alexander
Lauer, Bernward
Brunelli, Michele
Daralammouri, Yunis
Geller, Christoph
author_facet Ohlow, Marc-Alexander
Lauer, Bernward
Brunelli, Michele
Daralammouri, Yunis
Geller, Christoph
author_sort Ohlow, Marc-Alexander
collection PubMed
description BACKGROUND: Phrenic nerve stimulation (PNS) and high pacing thresholds (HPT) hinder biventricular stimulation in patients (pts) undergoing cardiac resynchronization therapy (CRT). A new quadripolar left ventricular (LV) lead (Quartet 1458Q, St. Jude Medical) with increased number of pacing configuration, might overcome this problem. METHODS: All consecutive pts in whom a standard bipolar lead intraoperatively resulted in PNS and/or HPT (≥4.00V/1mV), received, during the same implant, a quadripolar LV lead. Aim of the study was to evaluate acute and short term outcome. RESULTS: 26 pts [24 (92%) male, mean age 74±6 years)] with PNS (22 pts; 85%) and HPT (4 pts; 15%) were included. Permanent right ventricular pacing was the reason for broad QRS complex in 4 (15%) pts, whereas all other pts had a left bundle branch block. Severely symptomatic (NYHA Class ≥3) heart failure with reduced ejection fraction (EF 31±9%) was mostly caused by ischemic heart disease (14 pts; 54%). Idiopathic dilated cardiomyopathy and valvular heart disease were diagnosed in 6 (23%) pts each. In most (24/26, 92%) pts the use of the Quartet lead led to successful biventricular pacing due to a significant reduction in intraoperative pacing threshold (5.2V/1.0ms vs. 1.4V/0.8ms; p=0.03), which was maintained (1.2V/0.7ms) at follow-up. PNS never represented reason for failed LV pacing, neither acutely nor during follow-up. CONCLUSION: Excessively HPT and/or PNS are frequently encountered when conventional bipolar leads are used for CRT. A new quadripolar LV lead increases the rate of successful biventricular stimulation. Lower pacing threshold and freedom from PNS are maintained at follow-up.
format Online
Article
Text
id pubmed-3594899
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Indian Heart Rhythm Society
record_format MEDLINE/PubMed
spelling pubmed-35948992013-04-09 The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing Cardiac Resynchronisation Therapy with Conventional Bipolar Leads Ohlow, Marc-Alexander Lauer, Bernward Brunelli, Michele Daralammouri, Yunis Geller, Christoph Indian Pacing Electrophysiol J Original Article BACKGROUND: Phrenic nerve stimulation (PNS) and high pacing thresholds (HPT) hinder biventricular stimulation in patients (pts) undergoing cardiac resynchronization therapy (CRT). A new quadripolar left ventricular (LV) lead (Quartet 1458Q, St. Jude Medical) with increased number of pacing configuration, might overcome this problem. METHODS: All consecutive pts in whom a standard bipolar lead intraoperatively resulted in PNS and/or HPT (≥4.00V/1mV), received, during the same implant, a quadripolar LV lead. Aim of the study was to evaluate acute and short term outcome. RESULTS: 26 pts [24 (92%) male, mean age 74±6 years)] with PNS (22 pts; 85%) and HPT (4 pts; 15%) were included. Permanent right ventricular pacing was the reason for broad QRS complex in 4 (15%) pts, whereas all other pts had a left bundle branch block. Severely symptomatic (NYHA Class ≥3) heart failure with reduced ejection fraction (EF 31±9%) was mostly caused by ischemic heart disease (14 pts; 54%). Idiopathic dilated cardiomyopathy and valvular heart disease were diagnosed in 6 (23%) pts each. In most (24/26, 92%) pts the use of the Quartet lead led to successful biventricular pacing due to a significant reduction in intraoperative pacing threshold (5.2V/1.0ms vs. 1.4V/0.8ms; p=0.03), which was maintained (1.2V/0.7ms) at follow-up. PNS never represented reason for failed LV pacing, neither acutely nor during follow-up. CONCLUSION: Excessively HPT and/or PNS are frequently encountered when conventional bipolar leads are used for CRT. A new quadripolar LV lead increases the rate of successful biventricular stimulation. Lower pacing threshold and freedom from PNS are maintained at follow-up. Indian Heart Rhythm Society 2013-03-07 /pmc/articles/PMC3594899/ /pubmed/23573059 Text en Copyright: © 2013 Ohlow et al. http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Ohlow, Marc-Alexander
Lauer, Bernward
Brunelli, Michele
Daralammouri, Yunis
Geller, Christoph
The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing Cardiac Resynchronisation Therapy with Conventional Bipolar Leads
title The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing Cardiac Resynchronisation Therapy with Conventional Bipolar Leads
title_full The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing Cardiac Resynchronisation Therapy with Conventional Bipolar Leads
title_fullStr The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing Cardiac Resynchronisation Therapy with Conventional Bipolar Leads
title_full_unstemmed The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing Cardiac Resynchronisation Therapy with Conventional Bipolar Leads
title_short The Use of a Quadripolar Left Ventricular Lead Increases Successful Implantation Rates in Patients with Phrenic Nerve Stimulation and/or High Pacing Thresholds Undergoing Cardiac Resynchronisation Therapy with Conventional Bipolar Leads
title_sort use of a quadripolar left ventricular lead increases successful implantation rates in patients with phrenic nerve stimulation and/or high pacing thresholds undergoing cardiac resynchronisation therapy with conventional bipolar leads
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3594899/
https://www.ncbi.nlm.nih.gov/pubmed/23573059
work_keys_str_mv AT ohlowmarcalexander theuseofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads
AT lauerbernward theuseofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads
AT brunellimichele theuseofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads
AT daralammouriyunis theuseofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads
AT gellerchristoph theuseofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads
AT ohlowmarcalexander useofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads
AT lauerbernward useofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads
AT brunellimichele useofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads
AT daralammouriyunis useofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads
AT gellerchristoph useofaquadripolarleftventricularleadincreasessuccessfulimplantationratesinpatientswithphrenicnervestimulationandorhighpacingthresholdsundergoingcardiacresynchronisationtherapywithconventionalbipolarleads