Effects of a Novel Push-through Technique Using the Implantable Collamer Lens Injector System for Graft Delivery during Endothelial Keratoplasty
PURPOSE: To investigate effects of a new push-through insertion method for donor lenticules using an injector system on endothelial viability ex vivo and in a clinical case series of endothelial keratoplasty. METHODS: An ex vivo delivery model was used with porcine corneoscleral rims. We compared th...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Ophthalmological Society
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3596624/ https://www.ncbi.nlm.nih.gov/pubmed/23543832 http://dx.doi.org/10.3341/kjo.2013.27.2.87 |
Sumario: | PURPOSE: To investigate effects of a new push-through insertion method for donor lenticules using an injector system on endothelial viability ex vivo and in a clinical case series of endothelial keratoplasty. METHODS: An ex vivo delivery model was used with porcine corneoscleral rims. We compared the endothelial viability in a new push-through insertion method using the Visian Implantable Collamer Lens (ICL) injector versus that of standard forceps-assisted insertion for lenticule delivery. Twenty porcine corneal lenticules were divided into four groups by insertion method and wound size. Vital dye staining was performed and devitalized areas were semi-quantitatively assessed by digital imaging. In the clinical case series, Descemet's stripping endothelial keratoplasty (DSEK) using the push-through method was performed in seven patients and endothelial outcome was determined six months postoperatively. RESULTS: Mean devitalized areas for the push-through method were significantly lower than for forceps-assisted insertion through 3.2 mm incision (23.99 ± 2.17% vs. 50.48 ± 5.07%, p = 0.009) in the ex vivo model. Average endothelial cell counts of donor tissues of patients who underwent DSEK were 26.4% lower six months postoperatively. CONCLUSIONS: Push-through delivery of donor lenticules using the Visian ICL injector system appears to be less harmful to endothelial cells than conventional forceps-assisted delivery. |
---|