Cargando…

Distortion-free freehand-scanning OCT implemented with real-time scanning speed variance correction

Hand-held OCT systems that offer physicians greater freedom to access imaging sites of interest could be useful for many clinical applications. In this study, by incorporating the theoretical speckle model into the decorrelation function, we have explicitly correlated the cross-correlation coefficie...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Xuan, Huang, Yong, Kang, Jin U.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Optical Society of America 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3601635/
http://dx.doi.org/10.1364/OE.20.016567
Descripción
Sumario:Hand-held OCT systems that offer physicians greater freedom to access imaging sites of interest could be useful for many clinical applications. In this study, by incorporating the theoretical speckle model into the decorrelation function, we have explicitly correlated the cross-correlation coefficient to the lateral displacement between adjacent A-scans. We used this model to develop and study a freehand-scanning OCT system capable of real-time scanning speed correction and distortion-free imaging—for the first time to the best our knowledge. To validate our model and the system, we performed a series of calibration experiments. Experimental results show that our method can extract lateral scanning distance. In addition, using the manually scanned hand-held OCT system, we obtained OCT images from various samples by freehand manual scanning, including images obtained from human in vivo.