Numerical investigations of MRI RF field induced heating for external fixation devices
BACKGROUND: The magnetic resonance imaging (MRI) radio frequency (RF) field induced heating on external fixation devices can be very high in the vicinity of device screws. Such induced RF heating is related to device constructs, device placements, as well as the device insertion depth into human sub...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3610272/ https://www.ncbi.nlm.nih.gov/pubmed/23394173 http://dx.doi.org/10.1186/1475-925X-12-12 |
_version_ | 1782264432947101696 |
---|---|
author | Liu, Yan Shen, Jianxiang Kainz, Wolfgang Qian, Songsong Wu, Wen Chen, Ji |
author_facet | Liu, Yan Shen, Jianxiang Kainz, Wolfgang Qian, Songsong Wu, Wen Chen, Ji |
author_sort | Liu, Yan |
collection | PubMed |
description | BACKGROUND: The magnetic resonance imaging (MRI) radio frequency (RF) field induced heating on external fixation devices can be very high in the vicinity of device screws. Such induced RF heating is related to device constructs, device placements, as well as the device insertion depth into human subjects. In this study, computational modeling is performed to determine factors associated with such induced heating. METHODS: Numerical modeling, based on the finite-difference time-domain (FDTD) method, is used to evaluate the temperature rises near external device screw tips inside the ASTM phantom for both 1.5-T and 3-T MRI systems. The modeling approach consists of 1) the development of RF coils for 1.5-T and 3-T, 2) the electromagnetic simulations of energy deposition near the screw tips of external fixation devices, and 3) the thermal simulations of temperature rises near the tips of these devices. RESULTS: It is found that changing insertion depth and screw spacing could largely affect the heating of these devices. In 1.5-T MRI system, smaller insertion depth and larger pin spacing will lead to higher temperature rise. However, for 3-T MRI system, the relation is not very clear when insertion depth is larger than 5 cm or when pin spacing became larger than 20 cm. The effect of connection bar material on device heating is also studied and the heating mechanism of the device is analysed. CONCLUSIONS: Numerical simulation is used to study RF heating for external fixation devices in both 1.5-T and 3-T MRI coils. Typically, shallower insertion depth and larger pin spacing with conductive bar lead to higher RF heating. The heating mechanism is explained using induced current along the device and power decay inside ASTM phantom. |
format | Online Article Text |
id | pubmed-3610272 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-36102722013-04-03 Numerical investigations of MRI RF field induced heating for external fixation devices Liu, Yan Shen, Jianxiang Kainz, Wolfgang Qian, Songsong Wu, Wen Chen, Ji Biomed Eng Online Research BACKGROUND: The magnetic resonance imaging (MRI) radio frequency (RF) field induced heating on external fixation devices can be very high in the vicinity of device screws. Such induced RF heating is related to device constructs, device placements, as well as the device insertion depth into human subjects. In this study, computational modeling is performed to determine factors associated with such induced heating. METHODS: Numerical modeling, based on the finite-difference time-domain (FDTD) method, is used to evaluate the temperature rises near external device screw tips inside the ASTM phantom for both 1.5-T and 3-T MRI systems. The modeling approach consists of 1) the development of RF coils for 1.5-T and 3-T, 2) the electromagnetic simulations of energy deposition near the screw tips of external fixation devices, and 3) the thermal simulations of temperature rises near the tips of these devices. RESULTS: It is found that changing insertion depth and screw spacing could largely affect the heating of these devices. In 1.5-T MRI system, smaller insertion depth and larger pin spacing will lead to higher temperature rise. However, for 3-T MRI system, the relation is not very clear when insertion depth is larger than 5 cm or when pin spacing became larger than 20 cm. The effect of connection bar material on device heating is also studied and the heating mechanism of the device is analysed. CONCLUSIONS: Numerical simulation is used to study RF heating for external fixation devices in both 1.5-T and 3-T MRI coils. Typically, shallower insertion depth and larger pin spacing with conductive bar lead to higher RF heating. The heating mechanism is explained using induced current along the device and power decay inside ASTM phantom. BioMed Central 2013-02-09 /pmc/articles/PMC3610272/ /pubmed/23394173 http://dx.doi.org/10.1186/1475-925X-12-12 Text en Copyright ©2013 Liu et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Liu, Yan Shen, Jianxiang Kainz, Wolfgang Qian, Songsong Wu, Wen Chen, Ji Numerical investigations of MRI RF field induced heating for external fixation devices |
title | Numerical investigations of MRI RF field induced heating for external fixation devices |
title_full | Numerical investigations of MRI RF field induced heating for external fixation devices |
title_fullStr | Numerical investigations of MRI RF field induced heating for external fixation devices |
title_full_unstemmed | Numerical investigations of MRI RF field induced heating for external fixation devices |
title_short | Numerical investigations of MRI RF field induced heating for external fixation devices |
title_sort | numerical investigations of mri rf field induced heating for external fixation devices |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3610272/ https://www.ncbi.nlm.nih.gov/pubmed/23394173 http://dx.doi.org/10.1186/1475-925X-12-12 |
work_keys_str_mv | AT liuyan numericalinvestigationsofmrirffieldinducedheatingforexternalfixationdevices AT shenjianxiang numericalinvestigationsofmrirffieldinducedheatingforexternalfixationdevices AT kainzwolfgang numericalinvestigationsofmrirffieldinducedheatingforexternalfixationdevices AT qiansongsong numericalinvestigationsofmrirffieldinducedheatingforexternalfixationdevices AT wuwen numericalinvestigationsofmrirffieldinducedheatingforexternalfixationdevices AT chenji numericalinvestigationsofmrirffieldinducedheatingforexternalfixationdevices |