Real-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective Study
Accurate and timely glucose monitoring is essential in intensive care units. Real-time continuous glucose monitoring system (CGMS) has been advocated for many years to improve glycemic management in critically ill patients. In order to determine the effect of calibration time on the accuracy of CGMS...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3610935/ https://www.ncbi.nlm.nih.gov/pubmed/23555886 http://dx.doi.org/10.1371/journal.pone.0060070 |
_version_ | 1782264523069063168 |
---|---|
author | Yue, Xiao-Yan Zheng, Yi Cai, Ye-Hua Yin, Ning-Ning Zhou, Jian-Xin |
author_facet | Yue, Xiao-Yan Zheng, Yi Cai, Ye-Hua Yin, Ning-Ning Zhou, Jian-Xin |
author_sort | Yue, Xiao-Yan |
collection | PubMed |
description | Accurate and timely glucose monitoring is essential in intensive care units. Real-time continuous glucose monitoring system (CGMS) has been advocated for many years to improve glycemic management in critically ill patients. In order to determine the effect of calibration time on the accuracy of CGMS, real-time subcutaneous CGMS was used in 18 critically ill patients. CGMS sensor was calibrated with blood glucose measurements by blood gas/glucose analyzer every 12 hours. Venous blood was sampled every 2 to 4 hours, and glucose concentration was measured by standard central laboratory device (CLD) and by blood gas/glucose analyzer. With CLD measurement as reference, relative absolute difference (mean±SD) in CGMS and blood gas/glucose analyzer were 14.4%±12.2% and 6.5%±6.2%, respectively. The percentage of matched points in Clarke error grid zone A was 74.8% in CGMS, and 98.4% in blood gas/glucose analyzer. The relative absolute difference of CGMS obtained within 6 hours after sensor calibration (8.8%±7.2%) was significantly less than that between 6 to 12 hours after calibration (20.1%±13.5%, p<0.0001). The percentage of matched points in Clarke error grid zone A was also significantly higher in data sets within 6 hours after calibration (92.4% versus 57.1%, p<0.0001). In conclusion, real-time subcutaneous CGMS is accurate in glucose monitoring in critically ill patients. CGMS sensor should be calibrated less than 6 hours, no matter what time interval recommended by manufacturer. |
format | Online Article Text |
id | pubmed-3610935 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-36109352013-04-03 Real-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective Study Yue, Xiao-Yan Zheng, Yi Cai, Ye-Hua Yin, Ning-Ning Zhou, Jian-Xin PLoS One Research Article Accurate and timely glucose monitoring is essential in intensive care units. Real-time continuous glucose monitoring system (CGMS) has been advocated for many years to improve glycemic management in critically ill patients. In order to determine the effect of calibration time on the accuracy of CGMS, real-time subcutaneous CGMS was used in 18 critically ill patients. CGMS sensor was calibrated with blood glucose measurements by blood gas/glucose analyzer every 12 hours. Venous blood was sampled every 2 to 4 hours, and glucose concentration was measured by standard central laboratory device (CLD) and by blood gas/glucose analyzer. With CLD measurement as reference, relative absolute difference (mean±SD) in CGMS and blood gas/glucose analyzer were 14.4%±12.2% and 6.5%±6.2%, respectively. The percentage of matched points in Clarke error grid zone A was 74.8% in CGMS, and 98.4% in blood gas/glucose analyzer. The relative absolute difference of CGMS obtained within 6 hours after sensor calibration (8.8%±7.2%) was significantly less than that between 6 to 12 hours after calibration (20.1%±13.5%, p<0.0001). The percentage of matched points in Clarke error grid zone A was also significantly higher in data sets within 6 hours after calibration (92.4% versus 57.1%, p<0.0001). In conclusion, real-time subcutaneous CGMS is accurate in glucose monitoring in critically ill patients. CGMS sensor should be calibrated less than 6 hours, no matter what time interval recommended by manufacturer. Public Library of Science 2013-03-28 /pmc/articles/PMC3610935/ /pubmed/23555886 http://dx.doi.org/10.1371/journal.pone.0060070 Text en © 2013 Yue et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Yue, Xiao-Yan Zheng, Yi Cai, Ye-Hua Yin, Ning-Ning Zhou, Jian-Xin Real-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective Study |
title | Real-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective Study |
title_full | Real-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective Study |
title_fullStr | Real-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective Study |
title_full_unstemmed | Real-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective Study |
title_short | Real-Time Continuous Glucose Monitoring Shows High Accuracy within 6 Hours after Sensor Calibration: A Prospective Study |
title_sort | real-time continuous glucose monitoring shows high accuracy within 6 hours after sensor calibration: a prospective study |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3610935/ https://www.ncbi.nlm.nih.gov/pubmed/23555886 http://dx.doi.org/10.1371/journal.pone.0060070 |
work_keys_str_mv | AT yuexiaoyan realtimecontinuousglucosemonitoringshowshighaccuracywithin6hoursaftersensorcalibrationaprospectivestudy AT zhengyi realtimecontinuousglucosemonitoringshowshighaccuracywithin6hoursaftersensorcalibrationaprospectivestudy AT caiyehua realtimecontinuousglucosemonitoringshowshighaccuracywithin6hoursaftersensorcalibrationaprospectivestudy AT yinningning realtimecontinuousglucosemonitoringshowshighaccuracywithin6hoursaftersensorcalibrationaprospectivestudy AT zhoujianxin realtimecontinuousglucosemonitoringshowshighaccuracywithin6hoursaftersensorcalibrationaprospectivestudy |