Differences in Blood Pressure Measurements Obtained Using an Automatic Oscillometric Sphygmomanometer Depending on Clothes-Wearing Status

BACKGROUND: According to the current guidelines for blood pressure monitoring, clinicians are recommended to measure blood pressure by completely exposing the upper arm. However, it is a common practice that blood pressure is measured with the cuff placed over the sleeve or with the sleeve rolled up...

Descripción completa

Detalles Bibliográficos
Autores principales: Ki, Ji Hoon, Oh, Mi Kyeong, Lee, Soo Hee
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Family Medicine 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3611103/
https://www.ncbi.nlm.nih.gov/pubmed/23560214
http://dx.doi.org/10.4082/kjfm.2013.34.2.145
_version_ 1782264542813749248
author Ki, Ji Hoon
Oh, Mi Kyeong
Lee, Soo Hee
author_facet Ki, Ji Hoon
Oh, Mi Kyeong
Lee, Soo Hee
author_sort Ki, Ji Hoon
collection PubMed
description BACKGROUND: According to the current guidelines for blood pressure monitoring, clinicians are recommended to measure blood pressure by completely exposing the upper arm. However, it is a common practice that blood pressure is measured with the cuff placed over the sleeve or with the sleeve rolled up. We therefore conducted this study to examine whether there are any differences in blood pressure measurements among the three different settings: the sleeve group, the rolled sleeve group, and the bare arm group. METHODS: We conducted the current study in 141 male and female adult patients who visited our clinical department. In these patients, we took repeatedly blood pressure measurements using the same automatic oscillometric device on three different settings. Then, we analyzed the results with the use of randomized block design analysis of variance. RESULTS: The mean values of systolic blood pressure (SBP) between the first reading and those of the second reading were 128.5 ± 10.6 mm Hg in the sleeve group, 128.3 ± 10.8 mm Hg in the rolled sleeve group, and 128.3 ± 10.7 mm Hg in the bare arm group. These results indicate that there were no significant differences among the three groups (P = 0.32). In addition, the mean values of diastolic blood pressure (DBP) between the first reading and those of the second reading were 80.7 ± 6.1 mm Hg in the sleeve group, 80.7 ± 6.1 mm Hg in the rolled sleeve group, and 80.6 ± 5.9 mm Hg in the bare arm group. These results indicate that there were no significant differences among the three groups (P = 0.77). In addition, based on the age, sex, past or current history of hypertension or diabetes mellitus, the thickness of sleeve, weight, a drinking history, and a smoking history, there were no significant differences in SBP and DBP among the three groups. CONCLUSION: There were no significant differences in blood pressure measurements between the three different settings (the sleeve group, the rolled sleeve group, and the bare arm group).
format Online
Article
Text
id pubmed-3611103
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher The Korean Academy of Family Medicine
record_format MEDLINE/PubMed
spelling pubmed-36111032013-04-04 Differences in Blood Pressure Measurements Obtained Using an Automatic Oscillometric Sphygmomanometer Depending on Clothes-Wearing Status Ki, Ji Hoon Oh, Mi Kyeong Lee, Soo Hee Korean J Fam Med Original Article BACKGROUND: According to the current guidelines for blood pressure monitoring, clinicians are recommended to measure blood pressure by completely exposing the upper arm. However, it is a common practice that blood pressure is measured with the cuff placed over the sleeve or with the sleeve rolled up. We therefore conducted this study to examine whether there are any differences in blood pressure measurements among the three different settings: the sleeve group, the rolled sleeve group, and the bare arm group. METHODS: We conducted the current study in 141 male and female adult patients who visited our clinical department. In these patients, we took repeatedly blood pressure measurements using the same automatic oscillometric device on three different settings. Then, we analyzed the results with the use of randomized block design analysis of variance. RESULTS: The mean values of systolic blood pressure (SBP) between the first reading and those of the second reading were 128.5 ± 10.6 mm Hg in the sleeve group, 128.3 ± 10.8 mm Hg in the rolled sleeve group, and 128.3 ± 10.7 mm Hg in the bare arm group. These results indicate that there were no significant differences among the three groups (P = 0.32). In addition, the mean values of diastolic blood pressure (DBP) between the first reading and those of the second reading were 80.7 ± 6.1 mm Hg in the sleeve group, 80.7 ± 6.1 mm Hg in the rolled sleeve group, and 80.6 ± 5.9 mm Hg in the bare arm group. These results indicate that there were no significant differences among the three groups (P = 0.77). In addition, based on the age, sex, past or current history of hypertension or diabetes mellitus, the thickness of sleeve, weight, a drinking history, and a smoking history, there were no significant differences in SBP and DBP among the three groups. CONCLUSION: There were no significant differences in blood pressure measurements between the three different settings (the sleeve group, the rolled sleeve group, and the bare arm group). The Korean Academy of Family Medicine 2013-03 2013-03-20 /pmc/articles/PMC3611103/ /pubmed/23560214 http://dx.doi.org/10.4082/kjfm.2013.34.2.145 Text en Copyright © 2013 The Korean Academy of Family Medicine http://creativecommons.org/licenses/by-nc/3.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Ki, Ji Hoon
Oh, Mi Kyeong
Lee, Soo Hee
Differences in Blood Pressure Measurements Obtained Using an Automatic Oscillometric Sphygmomanometer Depending on Clothes-Wearing Status
title Differences in Blood Pressure Measurements Obtained Using an Automatic Oscillometric Sphygmomanometer Depending on Clothes-Wearing Status
title_full Differences in Blood Pressure Measurements Obtained Using an Automatic Oscillometric Sphygmomanometer Depending on Clothes-Wearing Status
title_fullStr Differences in Blood Pressure Measurements Obtained Using an Automatic Oscillometric Sphygmomanometer Depending on Clothes-Wearing Status
title_full_unstemmed Differences in Blood Pressure Measurements Obtained Using an Automatic Oscillometric Sphygmomanometer Depending on Clothes-Wearing Status
title_short Differences in Blood Pressure Measurements Obtained Using an Automatic Oscillometric Sphygmomanometer Depending on Clothes-Wearing Status
title_sort differences in blood pressure measurements obtained using an automatic oscillometric sphygmomanometer depending on clothes-wearing status
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3611103/
https://www.ncbi.nlm.nih.gov/pubmed/23560214
http://dx.doi.org/10.4082/kjfm.2013.34.2.145
work_keys_str_mv AT kijihoon differencesinbloodpressuremeasurementsobtainedusinganautomaticoscillometricsphygmomanometerdependingonclotheswearingstatus
AT ohmikyeong differencesinbloodpressuremeasurementsobtainedusinganautomaticoscillometricsphygmomanometerdependingonclotheswearingstatus
AT leesoohee differencesinbloodpressuremeasurementsobtainedusinganautomaticoscillometricsphygmomanometerdependingonclotheswearingstatus