Cargando…

Numerical Study of Magnetoacoustic Signal Generation with Magnetic Induction Based on Inhomogeneous Conductivity Anisotropy

Magnetoacoustic tomography with magnetic induction (MAT-MI) is a noninvasive imaging modality for generating electrical conductivity images of biological tissues with high spatial resolution. In this paper, we create a numerical model, including a permanent magnet, a coil, and a two-layer coaxial cy...

Descripción completa

Detalles Bibliográficos
Autores principales: Li, Xun, Hu, Sanqing, Li, Lihua, Zhu, Shanan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3622383/
https://www.ncbi.nlm.nih.gov/pubmed/23606896
http://dx.doi.org/10.1155/2013/161357
_version_ 1782265826832809984
author Li, Xun
Hu, Sanqing
Li, Lihua
Zhu, Shanan
author_facet Li, Xun
Hu, Sanqing
Li, Lihua
Zhu, Shanan
author_sort Li, Xun
collection PubMed
description Magnetoacoustic tomography with magnetic induction (MAT-MI) is a noninvasive imaging modality for generating electrical conductivity images of biological tissues with high spatial resolution. In this paper, we create a numerical model, including a permanent magnet, a coil, and a two-layer coaxial cylinder with anisotropic electrical conductivities, for the MAT-MI forward problem. We analyze the MAT-MI sources in two cases, on a thin conductive boundary layer and in a homogeneous medium, and then develop a feasible numerical approach to solve the MAT-MI sound source densities in the anisotropic conductive model based on finite element analysis of electromagnetic field. Using the numerical finite element method, we then investigate the magnetoacoustic effect of anisotropic conductivity under the inhomogeneous static magnetic field and inhomogeneous magnetic field, quantitatively compute the boundary source densities in the conductive model, and calculate the sound pressure. The anisotropic conductivity contributes to the distribution of the eddy current density, Lorentz force density, and acoustic signal. The proposed models and approaches provide a more realistic simulation environment for MAT-MI.
format Online
Article
Text
id pubmed-3622383
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-36223832013-04-19 Numerical Study of Magnetoacoustic Signal Generation with Magnetic Induction Based on Inhomogeneous Conductivity Anisotropy Li, Xun Hu, Sanqing Li, Lihua Zhu, Shanan Comput Math Methods Med Research Article Magnetoacoustic tomography with magnetic induction (MAT-MI) is a noninvasive imaging modality for generating electrical conductivity images of biological tissues with high spatial resolution. In this paper, we create a numerical model, including a permanent magnet, a coil, and a two-layer coaxial cylinder with anisotropic electrical conductivities, for the MAT-MI forward problem. We analyze the MAT-MI sources in two cases, on a thin conductive boundary layer and in a homogeneous medium, and then develop a feasible numerical approach to solve the MAT-MI sound source densities in the anisotropic conductive model based on finite element analysis of electromagnetic field. Using the numerical finite element method, we then investigate the magnetoacoustic effect of anisotropic conductivity under the inhomogeneous static magnetic field and inhomogeneous magnetic field, quantitatively compute the boundary source densities in the conductive model, and calculate the sound pressure. The anisotropic conductivity contributes to the distribution of the eddy current density, Lorentz force density, and acoustic signal. The proposed models and approaches provide a more realistic simulation environment for MAT-MI. Hindawi Publishing Corporation 2013 2013-03-26 /pmc/articles/PMC3622383/ /pubmed/23606896 http://dx.doi.org/10.1155/2013/161357 Text en Copyright © 2013 Xun Li et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Li, Xun
Hu, Sanqing
Li, Lihua
Zhu, Shanan
Numerical Study of Magnetoacoustic Signal Generation with Magnetic Induction Based on Inhomogeneous Conductivity Anisotropy
title Numerical Study of Magnetoacoustic Signal Generation with Magnetic Induction Based on Inhomogeneous Conductivity Anisotropy
title_full Numerical Study of Magnetoacoustic Signal Generation with Magnetic Induction Based on Inhomogeneous Conductivity Anisotropy
title_fullStr Numerical Study of Magnetoacoustic Signal Generation with Magnetic Induction Based on Inhomogeneous Conductivity Anisotropy
title_full_unstemmed Numerical Study of Magnetoacoustic Signal Generation with Magnetic Induction Based on Inhomogeneous Conductivity Anisotropy
title_short Numerical Study of Magnetoacoustic Signal Generation with Magnetic Induction Based on Inhomogeneous Conductivity Anisotropy
title_sort numerical study of magnetoacoustic signal generation with magnetic induction based on inhomogeneous conductivity anisotropy
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3622383/
https://www.ncbi.nlm.nih.gov/pubmed/23606896
http://dx.doi.org/10.1155/2013/161357
work_keys_str_mv AT lixun numericalstudyofmagnetoacousticsignalgenerationwithmagneticinductionbasedoninhomogeneousconductivityanisotropy
AT husanqing numericalstudyofmagnetoacousticsignalgenerationwithmagneticinductionbasedoninhomogeneousconductivityanisotropy
AT lilihua numericalstudyofmagnetoacousticsignalgenerationwithmagneticinductionbasedoninhomogeneousconductivityanisotropy
AT zhushanan numericalstudyofmagnetoacousticsignalgenerationwithmagneticinductionbasedoninhomogeneousconductivityanisotropy