Cargando…

Development and validation of an high-performance liquid chromatographic, and a ultraviolet spectrophotometric method for determination of Ambroxol hydrochloride in pharmaceutical preparations

A high-performance liquid chromatographic (HPLC) and ultraviolet (UV) methods were developed and validated for the quantitative determination of Ambroxol hydrochloride (AMH) in pharmaceutical dosage form. HPLC was carried out by reversed phase (RP) technique on an RP-18 column with a mobile phase co...

Descripción completa

Detalles Bibliográficos
Autores principales: Muralidharan, Selvadurai, Kumar, Jaya Raja, Dhanara, Sokkalingam Arumugam
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3645355/
https://www.ncbi.nlm.nih.gov/pubmed/23662284
http://dx.doi.org/10.4103/2231-4040.107503
_version_ 1782268503205609472
author Muralidharan, Selvadurai
Kumar, Jaya Raja
Dhanara, Sokkalingam Arumugam
author_facet Muralidharan, Selvadurai
Kumar, Jaya Raja
Dhanara, Sokkalingam Arumugam
author_sort Muralidharan, Selvadurai
collection PubMed
description A high-performance liquid chromatographic (HPLC) and ultraviolet (UV) methods were developed and validated for the quantitative determination of Ambroxol hydrochloride (AMH) in pharmaceutical dosage form. HPLC was carried out by reversed phase (RP) technique on an RP-18 column with a mobile phase composed of acetonitrile and water (pH 3.5 adjusted with orthophosphoric acid [60:40, v/v]). UV method was performed with the λ(max) at 250 nm. Both the methods showed good linearity, reproducibility, and precision. No spectral or chromatographic interferences from the tablet excipients were found in UV and HPLC. The method was successfully applied to commercial tablets. Validation parameters such as linearity, precision, accuracy, and specificity were determined. The HPLC Limit of detection (LOD) and Limit of quantification (LOQ) for Ambroxol were found to be 1 and 5 ng/ml, respectively. The UV LOD and LOQ for Ambroxol were found to be 1 and 4 μg/ml, respectively. The results were statistically compared using one-way analysis of variance. The proposed economical method could be applicable for routine analysis of AMH and monitoring of the quality of marketed drugs.
format Online
Article
Text
id pubmed-3645355
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-36453552013-05-09 Development and validation of an high-performance liquid chromatographic, and a ultraviolet spectrophotometric method for determination of Ambroxol hydrochloride in pharmaceutical preparations Muralidharan, Selvadurai Kumar, Jaya Raja Dhanara, Sokkalingam Arumugam J Adv Pharm Technol Res Short Communication A high-performance liquid chromatographic (HPLC) and ultraviolet (UV) methods were developed and validated for the quantitative determination of Ambroxol hydrochloride (AMH) in pharmaceutical dosage form. HPLC was carried out by reversed phase (RP) technique on an RP-18 column with a mobile phase composed of acetonitrile and water (pH 3.5 adjusted with orthophosphoric acid [60:40, v/v]). UV method was performed with the λ(max) at 250 nm. Both the methods showed good linearity, reproducibility, and precision. No spectral or chromatographic interferences from the tablet excipients were found in UV and HPLC. The method was successfully applied to commercial tablets. Validation parameters such as linearity, precision, accuracy, and specificity were determined. The HPLC Limit of detection (LOD) and Limit of quantification (LOQ) for Ambroxol were found to be 1 and 5 ng/ml, respectively. The UV LOD and LOQ for Ambroxol were found to be 1 and 4 μg/ml, respectively. The results were statistically compared using one-way analysis of variance. The proposed economical method could be applicable for routine analysis of AMH and monitoring of the quality of marketed drugs. Medknow Publications & Media Pvt Ltd 2013 /pmc/articles/PMC3645355/ /pubmed/23662284 http://dx.doi.org/10.4103/2231-4040.107503 Text en Copyright: © Journal of Advanced Pharmaceutical Technology & Research http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Short Communication
Muralidharan, Selvadurai
Kumar, Jaya Raja
Dhanara, Sokkalingam Arumugam
Development and validation of an high-performance liquid chromatographic, and a ultraviolet spectrophotometric method for determination of Ambroxol hydrochloride in pharmaceutical preparations
title Development and validation of an high-performance liquid chromatographic, and a ultraviolet spectrophotometric method for determination of Ambroxol hydrochloride in pharmaceutical preparations
title_full Development and validation of an high-performance liquid chromatographic, and a ultraviolet spectrophotometric method for determination of Ambroxol hydrochloride in pharmaceutical preparations
title_fullStr Development and validation of an high-performance liquid chromatographic, and a ultraviolet spectrophotometric method for determination of Ambroxol hydrochloride in pharmaceutical preparations
title_full_unstemmed Development and validation of an high-performance liquid chromatographic, and a ultraviolet spectrophotometric method for determination of Ambroxol hydrochloride in pharmaceutical preparations
title_short Development and validation of an high-performance liquid chromatographic, and a ultraviolet spectrophotometric method for determination of Ambroxol hydrochloride in pharmaceutical preparations
title_sort development and validation of an high-performance liquid chromatographic, and a ultraviolet spectrophotometric method for determination of ambroxol hydrochloride in pharmaceutical preparations
topic Short Communication
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3645355/
https://www.ncbi.nlm.nih.gov/pubmed/23662284
http://dx.doi.org/10.4103/2231-4040.107503
work_keys_str_mv AT muralidharanselvadurai developmentandvalidationofanhighperformanceliquidchromatographicandaultravioletspectrophotometricmethodfordeterminationofambroxolhydrochlorideinpharmaceuticalpreparations
AT kumarjayaraja developmentandvalidationofanhighperformanceliquidchromatographicandaultravioletspectrophotometricmethodfordeterminationofambroxolhydrochlorideinpharmaceuticalpreparations
AT dhanarasokkalingamarumugam developmentandvalidationofanhighperformanceliquidchromatographicandaultravioletspectrophotometricmethodfordeterminationofambroxolhydrochlorideinpharmaceuticalpreparations