Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters

This work investigated the suitability of passive dosimeters for reference dosimetry in small fields with acceptable accuracy. Absorbed dose to water rate was determined in nine small radiation fields with diameters between 4 and 35 mm in a Leksell Gamma Knife (LGK) and a modified linear accelerator...

Descripción completa

Detalles Bibliográficos
Autores principales: Massillon-JL, Guerda, Cueva-Prócel, Diego, Díaz-Aguirre, Porfirio, Rodríguez-Ponce, Miguel, Herrera-Martínez, Flor
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3650064/
https://www.ncbi.nlm.nih.gov/pubmed/23671677
http://dx.doi.org/10.1371/journal.pone.0063418
_version_ 1782269069400997888
author Massillon-JL, Guerda
Cueva-Prócel, Diego
Díaz-Aguirre, Porfirio
Rodríguez-Ponce, Miguel
Herrera-Martínez, Flor
author_facet Massillon-JL, Guerda
Cueva-Prócel, Diego
Díaz-Aguirre, Porfirio
Rodríguez-Ponce, Miguel
Herrera-Martínez, Flor
author_sort Massillon-JL, Guerda
collection PubMed
description This work investigated the suitability of passive dosimeters for reference dosimetry in small fields with acceptable accuracy. Absorbed dose to water rate was determined in nine small radiation fields with diameters between 4 and 35 mm in a Leksell Gamma Knife (LGK) and a modified linear accelerator (linac) for stereotactic radiosurgery treatments. Measurements were made using Gafchromic film (MD-V2-55), alanine and thermoluminescent (TLD-100) dosimeters and compared with conventional dosimetry systems. Detectors were calibrated in terms of absorbed dose to water in (60)Co gamma-ray and 6 MV x-ray reference (10×10 cm(2)) fields using an ionization chamber calibrated at a standards laboratory. Absorbed dose to water rate computed with MD-V2-55 was higher than that obtained with the others dosimeters, possibly due to a smaller volume averaging effect. Ratio between the dose-rates determined with each dosimeter and those obtained with the film was evaluated for both treatment modalities. For the LGK, the ratio decreased as the dosimeter size increased and remained constant for collimator diameters larger than 8 mm. The same behaviour was observed for the linac and the ratio increased with field size, independent of the dosimeter used. These behaviours could be explained as an averaging volume effect due to dose gradient and lack of electronic equilibrium. Evaluation of the output factors for the LGK collimators indicated that, even when agreement was observed between Monte Carlo simulation and measurements with different dosimeters, this does not warrant that the absorbed dose to water rate in the field was properly known and thus, investigation of the reference dosimetry should be an important issue. These results indicated that alanine dosimeter provides a high degree of accuracy but cannot be used in fields smaller than 20 mm diameter. Gafchromic film can be considered as a suitable methodology for reference dosimetry. TLD dosimeters are not appropriate in fields smaller than 10 mm diameters.
format Online
Article
Text
id pubmed-3650064
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-36500642013-05-13 Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters Massillon-JL, Guerda Cueva-Prócel, Diego Díaz-Aguirre, Porfirio Rodríguez-Ponce, Miguel Herrera-Martínez, Flor PLoS One Research Article This work investigated the suitability of passive dosimeters for reference dosimetry in small fields with acceptable accuracy. Absorbed dose to water rate was determined in nine small radiation fields with diameters between 4 and 35 mm in a Leksell Gamma Knife (LGK) and a modified linear accelerator (linac) for stereotactic radiosurgery treatments. Measurements were made using Gafchromic film (MD-V2-55), alanine and thermoluminescent (TLD-100) dosimeters and compared with conventional dosimetry systems. Detectors were calibrated in terms of absorbed dose to water in (60)Co gamma-ray and 6 MV x-ray reference (10×10 cm(2)) fields using an ionization chamber calibrated at a standards laboratory. Absorbed dose to water rate computed with MD-V2-55 was higher than that obtained with the others dosimeters, possibly due to a smaller volume averaging effect. Ratio between the dose-rates determined with each dosimeter and those obtained with the film was evaluated for both treatment modalities. For the LGK, the ratio decreased as the dosimeter size increased and remained constant for collimator diameters larger than 8 mm. The same behaviour was observed for the linac and the ratio increased with field size, independent of the dosimeter used. These behaviours could be explained as an averaging volume effect due to dose gradient and lack of electronic equilibrium. Evaluation of the output factors for the LGK collimators indicated that, even when agreement was observed between Monte Carlo simulation and measurements with different dosimeters, this does not warrant that the absorbed dose to water rate in the field was properly known and thus, investigation of the reference dosimetry should be an important issue. These results indicated that alanine dosimeter provides a high degree of accuracy but cannot be used in fields smaller than 20 mm diameter. Gafchromic film can be considered as a suitable methodology for reference dosimetry. TLD dosimeters are not appropriate in fields smaller than 10 mm diameters. Public Library of Science 2013-05-09 /pmc/articles/PMC3650064/ /pubmed/23671677 http://dx.doi.org/10.1371/journal.pone.0063418 Text en © 2013 Massillon-JL et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Massillon-JL, Guerda
Cueva-Prócel, Diego
Díaz-Aguirre, Porfirio
Rodríguez-Ponce, Miguel
Herrera-Martínez, Flor
Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters
title Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters
title_full Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters
title_fullStr Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters
title_full_unstemmed Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters
title_short Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters
title_sort dosimetry for small fields in stereotactic radiosurgery using gafchromic md-v2-55 film, tld-100 and alanine dosimeters
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3650064/
https://www.ncbi.nlm.nih.gov/pubmed/23671677
http://dx.doi.org/10.1371/journal.pone.0063418
work_keys_str_mv AT massillonjlguerda dosimetryforsmallfieldsinstereotacticradiosurgeryusinggafchromicmdv255filmtld100andalaninedosimeters
AT cuevaproceldiego dosimetryforsmallfieldsinstereotacticradiosurgeryusinggafchromicmdv255filmtld100andalaninedosimeters
AT diazaguirreporfirio dosimetryforsmallfieldsinstereotacticradiosurgeryusinggafchromicmdv255filmtld100andalaninedosimeters
AT rodriguezponcemiguel dosimetryforsmallfieldsinstereotacticradiosurgeryusinggafchromicmdv255filmtld100andalaninedosimeters
AT herreramartinezflor dosimetryforsmallfieldsinstereotacticradiosurgeryusinggafchromicmdv255filmtld100andalaninedosimeters