Cargando…
RP-UPLC method development and validation for the simultaneous estimation of ibuprofen and famotidine in pharmaceutical dosage form
AIM AND BACKRGOUND: A stability-indicating LC method was developed for the simultaneous determination of Ibuprofen and Famotidine in pharmaceutical dosage forms. MATERIALS AND METHODS: The chromatographic separation was achieved on Acquity UPLC BEH C-18,50 mm × 2.1 mm and 1.7 μm column with gradient...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3658083/ https://www.ncbi.nlm.nih.gov/pubmed/23781479 http://dx.doi.org/10.4103/2229-4708.103873 |
Sumario: | AIM AND BACKRGOUND: A stability-indicating LC method was developed for the simultaneous determination of Ibuprofen and Famotidine in pharmaceutical dosage forms. MATERIALS AND METHODS: The chromatographic separation was achieved on Acquity UPLC BEH C-18,50 mm × 2.1 mm and 1.7 μm column with gradient elution. The mobile phase A contains a mixture of 50 mM sodium acetate buffer (pH 5.5): methanol (85:15, v/v), and the mobile phase B contains a mixture of 50 mM sodium acetate buffer (pH 5.5): methanol (25:75, v/v). The flow rate was 0.3 mL min(-1), and the detection wavelength was 260 nm. RESULTS: The limit of detection for Ibuprofen and Famotidine was 1.6 and 1.2 μg mL(-1), respectively. The limit of quantification (LOQ) for Ibuprofen and Famotidine was 5.1 and 4.3 μg mL(-1), respectively. CONCLUSION: This method was validated for accuracy, precision, and linearity. The method was also found to be stability indicating. |
---|