Cargando…
The Magnitude of Errors Associated in Measuring the Loads Applied on an Assistive Device While Walking
Measurement of the loads exerted on the limb is a fundamental part of designing of an assistive device, which has been done by using strain gauges or a transducer. Although calculation of loads applied on an orthosis coefficients achieved from calibration is a standard way, most of researchers deter...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2012
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3662106/ https://www.ncbi.nlm.nih.gov/pubmed/23724373 |
_version_ | 1782270800617799680 |
---|---|
author | Karimi, Mohammad Taghi Jamshidi, Nima |
author_facet | Karimi, Mohammad Taghi Jamshidi, Nima |
author_sort | Karimi, Mohammad Taghi |
collection | PubMed |
description | Measurement of the loads exerted on the limb is a fundamental part of designing of an assistive device, which has been done by using strain gauges or a transducer. Although calculation of loads applied on an orthosis coefficients achieved from calibration is a standard way, most of researchers determined the loads based on available equations. Therefore, the aim of this research is finding the accuracy of this method with respect to calibration. Some strain gauges were attached on the lateral bar of a reciprocal gait orthosis. It was calibrated for axial force and the anteroposterior and mediolateral moments. The outputs of strain gauge were changed to force, and moments based on the coefficients were achieved from calibration using available equations, while 5 normal subjects were walking with the orthosis. There was a significant difference between the force and moments exerted on the orthosis during walking based on the two methods (mostly extension moment and compression force). The results of this research indicated that the output of strain gauge varies based on the methods of calculation. As calibration the gauge is a standard method, it is recommended to calibrate it before use. |
format | Online Article Text |
id | pubmed-3662106 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2012 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-36621062013-05-30 The Magnitude of Errors Associated in Measuring the Loads Applied on an Assistive Device While Walking Karimi, Mohammad Taghi Jamshidi, Nima J Med Signals Sens Original Article Measurement of the loads exerted on the limb is a fundamental part of designing of an assistive device, which has been done by using strain gauges or a transducer. Although calculation of loads applied on an orthosis coefficients achieved from calibration is a standard way, most of researchers determined the loads based on available equations. Therefore, the aim of this research is finding the accuracy of this method with respect to calibration. Some strain gauges were attached on the lateral bar of a reciprocal gait orthosis. It was calibrated for axial force and the anteroposterior and mediolateral moments. The outputs of strain gauge were changed to force, and moments based on the coefficients were achieved from calibration using available equations, while 5 normal subjects were walking with the orthosis. There was a significant difference between the force and moments exerted on the orthosis during walking based on the two methods (mostly extension moment and compression force). The results of this research indicated that the output of strain gauge varies based on the methods of calculation. As calibration the gauge is a standard method, it is recommended to calibrate it before use. Medknow Publications & Media Pvt Ltd 2012 /pmc/articles/PMC3662106/ /pubmed/23724373 Text en Copyright: © Journal of Medical Signals and Sensors http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Karimi, Mohammad Taghi Jamshidi, Nima The Magnitude of Errors Associated in Measuring the Loads Applied on an Assistive Device While Walking |
title | The Magnitude of Errors Associated in Measuring the Loads Applied on an Assistive Device While Walking |
title_full | The Magnitude of Errors Associated in Measuring the Loads Applied on an Assistive Device While Walking |
title_fullStr | The Magnitude of Errors Associated in Measuring the Loads Applied on an Assistive Device While Walking |
title_full_unstemmed | The Magnitude of Errors Associated in Measuring the Loads Applied on an Assistive Device While Walking |
title_short | The Magnitude of Errors Associated in Measuring the Loads Applied on an Assistive Device While Walking |
title_sort | magnitude of errors associated in measuring the loads applied on an assistive device while walking |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3662106/ https://www.ncbi.nlm.nih.gov/pubmed/23724373 |
work_keys_str_mv | AT karimimohammadtaghi themagnitudeoferrorsassociatedinmeasuringtheloadsappliedonanassistivedevicewhilewalking AT jamshidinima themagnitudeoferrorsassociatedinmeasuringtheloadsappliedonanassistivedevicewhilewalking AT karimimohammadtaghi magnitudeoferrorsassociatedinmeasuringtheloadsappliedonanassistivedevicewhilewalking AT jamshidinima magnitudeoferrorsassociatedinmeasuringtheloadsappliedonanassistivedevicewhilewalking |