Application of Dispersive Liquid-Liquid Microextraction with Graphite Furnace Atomic Absorption Spectrometry for Determination of Trace Amounts of Zinc in Water Samples

A selective and simple method for separation and preconcentration of zinc ions was developed by using dispersive liquid-liquid microextraction. Parameters that have an effect on the microextraction efficiency such as volume of extraction and disperser solvent, extraction time, and adding salt were i...

Descripción completa

Detalles Bibliográficos
Autor principal: Mazloomifar, Ali
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3666227/
https://www.ncbi.nlm.nih.gov/pubmed/23737791
http://dx.doi.org/10.1155/2013/743792
Descripción
Sumario:A selective and simple method for separation and preconcentration of zinc ions was developed by using dispersive liquid-liquid microextraction. Parameters that have an effect on the microextraction efficiency such as volume of extraction and disperser solvent, extraction time, and adding salt were investigated. Under optimum conditions, a preconcentration factor of 250 was obtained. The limit of detection (LOD) obtained under the optimal conditions was 0.09 ng mL(−1). The linearity of method was obtained in range of 0.2–50 ng mL(−1) with a correlation coefficient (r) of 0.9974. The relative standard deviation for 10 replicate determinations at 1.0 ng mL(−1) of zinc was 2.53%. The proposed method was successfully applied to the analysis of zinc in water sample.