Prospective ECG-Triggered Coronary CT Angiography: Clinical Value of Noise-Based Tube Current Reduction Method with Iterative Reconstruction
OBJECTIVES: To evaluate the clinical value of noise-based tube current reduction method with iterative reconstruction for obtaining consistent image quality with dose optimization in prospective electrocardiogram (ECG)-triggered coronary CT angiography (CCTA). MATERIALS AND METHODS: We performed a p...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3669389/ https://www.ncbi.nlm.nih.gov/pubmed/23741444 http://dx.doi.org/10.1371/journal.pone.0065025 |
_version_ | 1782271748194959360 |
---|---|
author | Shen, Junlin Du, Xiangying Guo, Daode Cao, Lizhen Gao, Yan Yang, Qi Li, Pengyu Liu, Jiabin Li, Kuncheng |
author_facet | Shen, Junlin Du, Xiangying Guo, Daode Cao, Lizhen Gao, Yan Yang, Qi Li, Pengyu Liu, Jiabin Li, Kuncheng |
author_sort | Shen, Junlin |
collection | PubMed |
description | OBJECTIVES: To evaluate the clinical value of noise-based tube current reduction method with iterative reconstruction for obtaining consistent image quality with dose optimization in prospective electrocardiogram (ECG)-triggered coronary CT angiography (CCTA). MATERIALS AND METHODS: We performed a prospective randomized study evaluating 338 patients undergoing CCTA with prospective ECG-triggering. Patients were randomly assigned to fixed tube current with filtered back projection (Group 1, n = 113), noise-based tube current with filtered back projection (Group 2, n = 109) or with iterative reconstruction (Group 3, n = 116). Tube voltage was fixed at 120 kV. Qualitative image quality was rated on a 5-point scale (1 = impaired, to 5 = excellent, with 3–5 defined as diagnostic). Image noise and signal intensity were measured; signal-to-noise ratio was calculated; radiation dose parameters were recorded. Statistical analyses included one-way analysis of variance, chi-square test, Kruskal-Wallis test and multivariable linear regression. RESULTS: Image noise was maintained at the target value of 35HU with small interquartile range for Group 2 (35.00–35.03HU) and Group 3 (34.99–35.02HU), while from 28.73 to 37.87HU for Group 1. All images in the three groups were acceptable for diagnosis. A relative 20% and 51% reduction in effective dose for Group 2 (2.9 mSv) and Group 3 (1.8 mSv) were achieved compared with Group 1 (3.7 mSv). After adjustment for scan characteristics, iterative reconstruction was associated with 26% reduction in effective dose. CONCLUSION: Noise-based tube current reduction method with iterative reconstruction maintains image noise precisely at the desired level and achieves consistent image quality. Meanwhile, effective dose can be reduced by more than 50%. |
format | Online Article Text |
id | pubmed-3669389 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-36693892013-06-05 Prospective ECG-Triggered Coronary CT Angiography: Clinical Value of Noise-Based Tube Current Reduction Method with Iterative Reconstruction Shen, Junlin Du, Xiangying Guo, Daode Cao, Lizhen Gao, Yan Yang, Qi Li, Pengyu Liu, Jiabin Li, Kuncheng PLoS One Research Article OBJECTIVES: To evaluate the clinical value of noise-based tube current reduction method with iterative reconstruction for obtaining consistent image quality with dose optimization in prospective electrocardiogram (ECG)-triggered coronary CT angiography (CCTA). MATERIALS AND METHODS: We performed a prospective randomized study evaluating 338 patients undergoing CCTA with prospective ECG-triggering. Patients were randomly assigned to fixed tube current with filtered back projection (Group 1, n = 113), noise-based tube current with filtered back projection (Group 2, n = 109) or with iterative reconstruction (Group 3, n = 116). Tube voltage was fixed at 120 kV. Qualitative image quality was rated on a 5-point scale (1 = impaired, to 5 = excellent, with 3–5 defined as diagnostic). Image noise and signal intensity were measured; signal-to-noise ratio was calculated; radiation dose parameters were recorded. Statistical analyses included one-way analysis of variance, chi-square test, Kruskal-Wallis test and multivariable linear regression. RESULTS: Image noise was maintained at the target value of 35HU with small interquartile range for Group 2 (35.00–35.03HU) and Group 3 (34.99–35.02HU), while from 28.73 to 37.87HU for Group 1. All images in the three groups were acceptable for diagnosis. A relative 20% and 51% reduction in effective dose for Group 2 (2.9 mSv) and Group 3 (1.8 mSv) were achieved compared with Group 1 (3.7 mSv). After adjustment for scan characteristics, iterative reconstruction was associated with 26% reduction in effective dose. CONCLUSION: Noise-based tube current reduction method with iterative reconstruction maintains image noise precisely at the desired level and achieves consistent image quality. Meanwhile, effective dose can be reduced by more than 50%. Public Library of Science 2013-05-31 /pmc/articles/PMC3669389/ /pubmed/23741444 http://dx.doi.org/10.1371/journal.pone.0065025 Text en © 2013 Shen et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Shen, Junlin Du, Xiangying Guo, Daode Cao, Lizhen Gao, Yan Yang, Qi Li, Pengyu Liu, Jiabin Li, Kuncheng Prospective ECG-Triggered Coronary CT Angiography: Clinical Value of Noise-Based Tube Current Reduction Method with Iterative Reconstruction |
title | Prospective ECG-Triggered Coronary CT Angiography: Clinical Value of Noise-Based Tube Current Reduction Method with Iterative Reconstruction |
title_full | Prospective ECG-Triggered Coronary CT Angiography: Clinical Value of Noise-Based Tube Current Reduction Method with Iterative Reconstruction |
title_fullStr | Prospective ECG-Triggered Coronary CT Angiography: Clinical Value of Noise-Based Tube Current Reduction Method with Iterative Reconstruction |
title_full_unstemmed | Prospective ECG-Triggered Coronary CT Angiography: Clinical Value of Noise-Based Tube Current Reduction Method with Iterative Reconstruction |
title_short | Prospective ECG-Triggered Coronary CT Angiography: Clinical Value of Noise-Based Tube Current Reduction Method with Iterative Reconstruction |
title_sort | prospective ecg-triggered coronary ct angiography: clinical value of noise-based tube current reduction method with iterative reconstruction |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3669389/ https://www.ncbi.nlm.nih.gov/pubmed/23741444 http://dx.doi.org/10.1371/journal.pone.0065025 |
work_keys_str_mv | AT shenjunlin prospectiveecgtriggeredcoronaryctangiographyclinicalvalueofnoisebasedtubecurrentreductionmethodwithiterativereconstruction AT duxiangying prospectiveecgtriggeredcoronaryctangiographyclinicalvalueofnoisebasedtubecurrentreductionmethodwithiterativereconstruction AT guodaode prospectiveecgtriggeredcoronaryctangiographyclinicalvalueofnoisebasedtubecurrentreductionmethodwithiterativereconstruction AT caolizhen prospectiveecgtriggeredcoronaryctangiographyclinicalvalueofnoisebasedtubecurrentreductionmethodwithiterativereconstruction AT gaoyan prospectiveecgtriggeredcoronaryctangiographyclinicalvalueofnoisebasedtubecurrentreductionmethodwithiterativereconstruction AT yangqi prospectiveecgtriggeredcoronaryctangiographyclinicalvalueofnoisebasedtubecurrentreductionmethodwithiterativereconstruction AT lipengyu prospectiveecgtriggeredcoronaryctangiographyclinicalvalueofnoisebasedtubecurrentreductionmethodwithiterativereconstruction AT liujiabin prospectiveecgtriggeredcoronaryctangiographyclinicalvalueofnoisebasedtubecurrentreductionmethodwithiterativereconstruction AT likuncheng prospectiveecgtriggeredcoronaryctangiographyclinicalvalueofnoisebasedtubecurrentreductionmethodwithiterativereconstruction |