Comparison of Measurements of 12 Analytes in Equine Blood Samples Using the In-Practice Falcor 350 and the Reference KoneLab 30i Analysers

Falcor 350 is a wet-reagent biochemistry analyser that is available for in-house use. The aim of this study was to compare the results produced by this analyser with those obtained by the KoneLab 30i that served as the reference instrument. Blood samples from 60 clinical cases were analysed for urea...

Descripción completa

Detalles Bibliográficos
Autores principales: Papasouliotis, K., Tennant, K. V., Dodkin, S., Mason, J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: International Scholarly Research Network 2012
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3671722/
https://www.ncbi.nlm.nih.gov/pubmed/23762583
http://dx.doi.org/10.5402/2012/475419
_version_ 1782272028254928896
author Papasouliotis, K.
Tennant, K. V.
Dodkin, S.
Mason, J.
author_facet Papasouliotis, K.
Tennant, K. V.
Dodkin, S.
Mason, J.
author_sort Papasouliotis, K.
collection PubMed
description Falcor 350 is a wet-reagent biochemistry analyser that is available for in-house use. The aim of this study was to compare the results produced by this analyser with those obtained by the KoneLab 30i that served as the reference instrument. Blood samples from 60 clinical cases were analysed for urea, creatinine, total proteins, albumin, creatine kinase, aspartate aminotransferase, alkaline phosphatase, total bilirubin, total calcium, phosphate, sodium, and potassium using both instruments. Good to excellent correlations (r (s) value) value) were identified for creatinine (0.88), total proteins (0.92), albumin (0.93), creatine kinase (0.98), aspartate aminotransferase (0.98), alkaline phosphatase (0.94), total bilirubin (0.98), phosphate (0.95), and potassium (0.97). The correlations for total calcium (0.71), sodium (0.68), and urea (0.64) were fair. For albumin, aspartate aminotransferase, creatine kinase, phosphate, potassium, total bilirubin, creatinine, and total proteins, the two instruments produce values that are closely related to each other and are sufficiently similar to allow them to be used interchangeably without the need for additional correction factor computations. Because of differences in the methodologies, the Falcor results for alkaline phosphatase, total calcium, and sodium cannot be used interchangeably and should be interpreted using reference intervals established from the Falcor analyser.
format Online
Article
Text
id pubmed-3671722
institution National Center for Biotechnology Information
language English
publishDate 2012
publisher International Scholarly Research Network
record_format MEDLINE/PubMed
spelling pubmed-36717222013-06-12 Comparison of Measurements of 12 Analytes in Equine Blood Samples Using the In-Practice Falcor 350 and the Reference KoneLab 30i Analysers Papasouliotis, K. Tennant, K. V. Dodkin, S. Mason, J. ISRN Vet Sci Research Article Falcor 350 is a wet-reagent biochemistry analyser that is available for in-house use. The aim of this study was to compare the results produced by this analyser with those obtained by the KoneLab 30i that served as the reference instrument. Blood samples from 60 clinical cases were analysed for urea, creatinine, total proteins, albumin, creatine kinase, aspartate aminotransferase, alkaline phosphatase, total bilirubin, total calcium, phosphate, sodium, and potassium using both instruments. Good to excellent correlations (r (s) value) value) were identified for creatinine (0.88), total proteins (0.92), albumin (0.93), creatine kinase (0.98), aspartate aminotransferase (0.98), alkaline phosphatase (0.94), total bilirubin (0.98), phosphate (0.95), and potassium (0.97). The correlations for total calcium (0.71), sodium (0.68), and urea (0.64) were fair. For albumin, aspartate aminotransferase, creatine kinase, phosphate, potassium, total bilirubin, creatinine, and total proteins, the two instruments produce values that are closely related to each other and are sufficiently similar to allow them to be used interchangeably without the need for additional correction factor computations. Because of differences in the methodologies, the Falcor results for alkaline phosphatase, total calcium, and sodium cannot be used interchangeably and should be interpreted using reference intervals established from the Falcor analyser. International Scholarly Research Network 2012-09-27 /pmc/articles/PMC3671722/ /pubmed/23762583 http://dx.doi.org/10.5402/2012/475419 Text en Copyright © 2012 K. Papasouliotis et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Papasouliotis, K.
Tennant, K. V.
Dodkin, S.
Mason, J.
Comparison of Measurements of 12 Analytes in Equine Blood Samples Using the In-Practice Falcor 350 and the Reference KoneLab 30i Analysers
title Comparison of Measurements of 12 Analytes in Equine Blood Samples Using the In-Practice Falcor 350 and the Reference KoneLab 30i Analysers
title_full Comparison of Measurements of 12 Analytes in Equine Blood Samples Using the In-Practice Falcor 350 and the Reference KoneLab 30i Analysers
title_fullStr Comparison of Measurements of 12 Analytes in Equine Blood Samples Using the In-Practice Falcor 350 and the Reference KoneLab 30i Analysers
title_full_unstemmed Comparison of Measurements of 12 Analytes in Equine Blood Samples Using the In-Practice Falcor 350 and the Reference KoneLab 30i Analysers
title_short Comparison of Measurements of 12 Analytes in Equine Blood Samples Using the In-Practice Falcor 350 and the Reference KoneLab 30i Analysers
title_sort comparison of measurements of 12 analytes in equine blood samples using the in-practice falcor 350 and the reference konelab 30i analysers
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3671722/
https://www.ncbi.nlm.nih.gov/pubmed/23762583
http://dx.doi.org/10.5402/2012/475419
work_keys_str_mv AT papasouliotisk comparisonofmeasurementsof12analytesinequinebloodsamplesusingtheinpracticefalcor350andthereferencekonelab30ianalysers
AT tennantkv comparisonofmeasurementsof12analytesinequinebloodsamplesusingtheinpracticefalcor350andthereferencekonelab30ianalysers
AT dodkins comparisonofmeasurementsof12analytesinequinebloodsamplesusingtheinpracticefalcor350andthereferencekonelab30ianalysers
AT masonj comparisonofmeasurementsof12analytesinequinebloodsamplesusingtheinpracticefalcor350andthereferencekonelab30ianalysers