Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010

BACKGROUND: Allowing patients to measure their blood pressure at home is recognized as being of clinical value. However, it is not known how often these measurements are taken correctly. Blood pressure monitors for home use fall into two types based on the position of the cuff, ie, at the upper arm...

Descripción completa

Detalles Bibliográficos
Autores principales: Takahashi, Hakuo, Yoshika, Masamichi, Yokoi, Toyohiko
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3671796/
https://www.ncbi.nlm.nih.gov/pubmed/23745050
http://dx.doi.org/10.2147/VHRM.S44569
_version_ 1782272040595619840
author Takahashi, Hakuo
Yoshika, Masamichi
Yokoi, Toyohiko
author_facet Takahashi, Hakuo
Yoshika, Masamichi
Yokoi, Toyohiko
author_sort Takahashi, Hakuo
collection PubMed
description BACKGROUND: Allowing patients to measure their blood pressure at home is recognized as being of clinical value. However, it is not known how often these measurements are taken correctly. Blood pressure monitors for home use fall into two types based on the position of the cuff, ie, at the upper arm or the wrist. The latter is particularly convenient, as measurements can be taken fully clothed. This study aimed to evaluate the performance of the wrist-type blood pressure monitors Omron RS8 (HEM-6310F-E), Omron RS6 (HEM-6221-E), and Omron RS3 (HEM-6130-E). METHODS: A team of three trained doctors validated the performance of these devices by comparing the measurements obtained from these devices with those taken using a standard mercury sphygmomanometer. All the devices met the validation requirements of the European Society of Hypertension International Protocol revision 2010. RESULTS: The difference in blood pressure readings between the tested device and the standard mercury sphygmomanometer was within 3 mmHg, which is acceptable according to the European Society of Hypertension guidelines. CONCLUSION: All the home devices tested were found to be suitable for measuring blood pressure at home because their performance fulfilled the requirement of the guidelines.
format Online
Article
Text
id pubmed-3671796
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-36717962013-06-06 Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010 Takahashi, Hakuo Yoshika, Masamichi Yokoi, Toyohiko Vasc Health Risk Manag Methodology BACKGROUND: Allowing patients to measure their blood pressure at home is recognized as being of clinical value. However, it is not known how often these measurements are taken correctly. Blood pressure monitors for home use fall into two types based on the position of the cuff, ie, at the upper arm or the wrist. The latter is particularly convenient, as measurements can be taken fully clothed. This study aimed to evaluate the performance of the wrist-type blood pressure monitors Omron RS8 (HEM-6310F-E), Omron RS6 (HEM-6221-E), and Omron RS3 (HEM-6130-E). METHODS: A team of three trained doctors validated the performance of these devices by comparing the measurements obtained from these devices with those taken using a standard mercury sphygmomanometer. All the devices met the validation requirements of the European Society of Hypertension International Protocol revision 2010. RESULTS: The difference in blood pressure readings between the tested device and the standard mercury sphygmomanometer was within 3 mmHg, which is acceptable according to the European Society of Hypertension guidelines. CONCLUSION: All the home devices tested were found to be suitable for measuring blood pressure at home because their performance fulfilled the requirement of the guidelines. Dove Medical Press 2013 2013-05-28 /pmc/articles/PMC3671796/ /pubmed/23745050 http://dx.doi.org/10.2147/VHRM.S44569 Text en © 2013 Takahashi et al, publisher and licensee Dove Medical Press Ltd This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited.
spellingShingle Methodology
Takahashi, Hakuo
Yoshika, Masamichi
Yokoi, Toyohiko
Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010
title Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010
title_full Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010
title_fullStr Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010
title_full_unstemmed Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010
title_short Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010
title_sort validation of omron rs8, rs6, and rs3 home blood pressure monitoring devices, in accordance with the european society of hypertension international protocol revision 2010
topic Methodology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3671796/
https://www.ncbi.nlm.nih.gov/pubmed/23745050
http://dx.doi.org/10.2147/VHRM.S44569
work_keys_str_mv AT takahashihakuo validationofomronrs8rs6andrs3homebloodpressuremonitoringdevicesinaccordancewiththeeuropeansocietyofhypertensioninternationalprotocolrevision2010
AT yoshikamasamichi validationofomronrs8rs6andrs3homebloodpressuremonitoringdevicesinaccordancewiththeeuropeansocietyofhypertensioninternationalprotocolrevision2010
AT yokoitoyohiko validationofomronrs8rs6andrs3homebloodpressuremonitoringdevicesinaccordancewiththeeuropeansocietyofhypertensioninternationalprotocolrevision2010