Cargando…

Performance Evaluation of a Modular Detector Unit for X-Ray Computed Tomography

A research prototype CT scanner is currently under development in our lab. One of the key components in this project is the CT detector. This paper describes the design and performance evaluation of the modular CT detector unit for our proposed scanner. It consists of a Photodiode Array Assembly whi...

Descripción completa

Detalles Bibliográficos
Autores principales: Guo, Zhe, Tang, Zhiwei, Wang, Xinzeng, Deng, Mingliang, Hu, Guangshu, Zhang, Hui
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Molecular Diversity Preservation International (MDPI) 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3673131/
https://www.ncbi.nlm.nih.gov/pubmed/23598502
http://dx.doi.org/10.3390/s130405167
_version_ 1782272216310743040
author Guo, Zhe
Tang, Zhiwei
Wang, Xinzeng
Deng, Mingliang
Hu, Guangshu
Zhang, Hui
author_facet Guo, Zhe
Tang, Zhiwei
Wang, Xinzeng
Deng, Mingliang
Hu, Guangshu
Zhang, Hui
author_sort Guo, Zhe
collection PubMed
description A research prototype CT scanner is currently under development in our lab. One of the key components in this project is the CT detector. This paper describes the design and performance evaluation of the modular CT detector unit for our proposed scanner. It consists of a Photodiode Array Assembly which captures irradiating X-ray photons and converts the energy into electrical current, and a mini Data Acquisition System which performs current integration and converts the analog signal into digital samples. The detector unit can be easily tiled together to form a CT detector. Experiments were conducted to characterize the detector performance both at the single unit level and system level. The noise level, linearity and uniformity of the proposed detector unit were reported and initial imaging studies were also presented which demonstrated the potential application of the proposed detector unit in actual CT scanners.
format Online
Article
Text
id pubmed-3673131
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Molecular Diversity Preservation International (MDPI)
record_format MEDLINE/PubMed
spelling pubmed-36731312013-06-19 Performance Evaluation of a Modular Detector Unit for X-Ray Computed Tomography Guo, Zhe Tang, Zhiwei Wang, Xinzeng Deng, Mingliang Hu, Guangshu Zhang, Hui Sensors (Basel) Article A research prototype CT scanner is currently under development in our lab. One of the key components in this project is the CT detector. This paper describes the design and performance evaluation of the modular CT detector unit for our proposed scanner. It consists of a Photodiode Array Assembly which captures irradiating X-ray photons and converts the energy into electrical current, and a mini Data Acquisition System which performs current integration and converts the analog signal into digital samples. The detector unit can be easily tiled together to form a CT detector. Experiments were conducted to characterize the detector performance both at the single unit level and system level. The noise level, linearity and uniformity of the proposed detector unit were reported and initial imaging studies were also presented which demonstrated the potential application of the proposed detector unit in actual CT scanners. Molecular Diversity Preservation International (MDPI) 2013-04-18 /pmc/articles/PMC3673131/ /pubmed/23598502 http://dx.doi.org/10.3390/s130405167 Text en © 2013 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).
spellingShingle Article
Guo, Zhe
Tang, Zhiwei
Wang, Xinzeng
Deng, Mingliang
Hu, Guangshu
Zhang, Hui
Performance Evaluation of a Modular Detector Unit for X-Ray Computed Tomography
title Performance Evaluation of a Modular Detector Unit for X-Ray Computed Tomography
title_full Performance Evaluation of a Modular Detector Unit for X-Ray Computed Tomography
title_fullStr Performance Evaluation of a Modular Detector Unit for X-Ray Computed Tomography
title_full_unstemmed Performance Evaluation of a Modular Detector Unit for X-Ray Computed Tomography
title_short Performance Evaluation of a Modular Detector Unit for X-Ray Computed Tomography
title_sort performance evaluation of a modular detector unit for x-ray computed tomography
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3673131/
https://www.ncbi.nlm.nih.gov/pubmed/23598502
http://dx.doi.org/10.3390/s130405167
work_keys_str_mv AT guozhe performanceevaluationofamodulardetectorunitforxraycomputedtomography
AT tangzhiwei performanceevaluationofamodulardetectorunitforxraycomputedtomography
AT wangxinzeng performanceevaluationofamodulardetectorunitforxraycomputedtomography
AT dengmingliang performanceevaluationofamodulardetectorunitforxraycomputedtomography
AT huguangshu performanceevaluationofamodulardetectorunitforxraycomputedtomography
AT zhanghui performanceevaluationofamodulardetectorunitforxraycomputedtomography