A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation
A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventr...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society of Cardiology
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3675308/ https://www.ncbi.nlm.nih.gov/pubmed/23755080 http://dx.doi.org/10.4070/kcj.2013.43.5.336 |
_version_ | 1782272508721889280 |
---|---|
author | Choi, Kyu Kim, Jung Hee Kim, Hyo Jin Lee, Se-ok Jang, Eun-young Kim, June Soo |
author_facet | Choi, Kyu Kim, Jung Hee Kim, Hyo Jin Lee, Se-ok Jang, Eun-young Kim, June Soo |
author_sort | Choi, Kyu |
collection | PubMed |
description | A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventricular fibrillation. Chest radiography revealed a significant split in the insulation of the lead allowing the inner wire to protrude. We considered the removal of the failed lead, but the removal of ICD lead is potentially a high risk procedure, so we cut and capped a proximal part of the failed lead and inserted a new lead. This is the first report of a St. Jude Riata® dual coil defibrillator lead failure with clinical and radiologic evidence of a defect in lead insulation in Korea. |
format | Online Article Text |
id | pubmed-3675308 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | The Korean Society of Cardiology |
record_format | MEDLINE/PubMed |
spelling | pubmed-36753082013-06-10 A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation Choi, Kyu Kim, Jung Hee Kim, Hyo Jin Lee, Se-ok Jang, Eun-young Kim, June Soo Korean Circ J Case Report A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventricular fibrillation. Chest radiography revealed a significant split in the insulation of the lead allowing the inner wire to protrude. We considered the removal of the failed lead, but the removal of ICD lead is potentially a high risk procedure, so we cut and capped a proximal part of the failed lead and inserted a new lead. This is the first report of a St. Jude Riata® dual coil defibrillator lead failure with clinical and radiologic evidence of a defect in lead insulation in Korea. The Korean Society of Cardiology 2013-05 2013-05-31 /pmc/articles/PMC3675308/ /pubmed/23755080 http://dx.doi.org/10.4070/kcj.2013.43.5.336 Text en Copyright © 2013 The Korean Society of Cardiology http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Choi, Kyu Kim, Jung Hee Kim, Hyo Jin Lee, Se-ok Jang, Eun-young Kim, June Soo A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation |
title | A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation |
title_full | A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation |
title_fullStr | A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation |
title_full_unstemmed | A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation |
title_short | A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation |
title_sort | case of riata® dual coil defibrillator lead failure in a patient with ventricular fibrillation |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3675308/ https://www.ncbi.nlm.nih.gov/pubmed/23755080 http://dx.doi.org/10.4070/kcj.2013.43.5.336 |
work_keys_str_mv | AT choikyu acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT kimjunghee acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT kimhyojin acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT leeseok acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT jangeunyoung acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT kimjunesoo acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT choikyu caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT kimjunghee caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT kimhyojin caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT leeseok caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT jangeunyoung caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation AT kimjunesoo caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation |