A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation

A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventr...

Descripción completa

Detalles Bibliográficos
Autores principales: Choi, Kyu, Kim, Jung Hee, Kim, Hyo Jin, Lee, Se-ok, Jang, Eun-young, Kim, June Soo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Society of Cardiology 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3675308/
https://www.ncbi.nlm.nih.gov/pubmed/23755080
http://dx.doi.org/10.4070/kcj.2013.43.5.336
_version_ 1782272508721889280
author Choi, Kyu
Kim, Jung Hee
Kim, Hyo Jin
Lee, Se-ok
Jang, Eun-young
Kim, June Soo
author_facet Choi, Kyu
Kim, Jung Hee
Kim, Hyo Jin
Lee, Se-ok
Jang, Eun-young
Kim, June Soo
author_sort Choi, Kyu
collection PubMed
description A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventricular fibrillation. Chest radiography revealed a significant split in the insulation of the lead allowing the inner wire to protrude. We considered the removal of the failed lead, but the removal of ICD lead is potentially a high risk procedure, so we cut and capped a proximal part of the failed lead and inserted a new lead. This is the first report of a St. Jude Riata® dual coil defibrillator lead failure with clinical and radiologic evidence of a defect in lead insulation in Korea.
format Online
Article
Text
id pubmed-3675308
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher The Korean Society of Cardiology
record_format MEDLINE/PubMed
spelling pubmed-36753082013-06-10 A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation Choi, Kyu Kim, Jung Hee Kim, Hyo Jin Lee, Se-ok Jang, Eun-young Kim, June Soo Korean Circ J Case Report A 50-year-old man, who underwent a procedure for an implantable cardioverter defibrillator (ICD), visited the outpatient department of our clinic after suffering multiple ICD shocks. The ICD interrogation revealed recurrent shock due to a high frequency of noise that is sensed by the device as ventricular fibrillation. Chest radiography revealed a significant split in the insulation of the lead allowing the inner wire to protrude. We considered the removal of the failed lead, but the removal of ICD lead is potentially a high risk procedure, so we cut and capped a proximal part of the failed lead and inserted a new lead. This is the first report of a St. Jude Riata® dual coil defibrillator lead failure with clinical and radiologic evidence of a defect in lead insulation in Korea. The Korean Society of Cardiology 2013-05 2013-05-31 /pmc/articles/PMC3675308/ /pubmed/23755080 http://dx.doi.org/10.4070/kcj.2013.43.5.336 Text en Copyright © 2013 The Korean Society of Cardiology http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Choi, Kyu
Kim, Jung Hee
Kim, Hyo Jin
Lee, Se-ok
Jang, Eun-young
Kim, June Soo
A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation
title A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation
title_full A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation
title_fullStr A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation
title_full_unstemmed A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation
title_short A Case of Riata® Dual Coil Defibrillator Lead Failure in a Patient with Ventricular Fibrillation
title_sort case of riata® dual coil defibrillator lead failure in a patient with ventricular fibrillation
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3675308/
https://www.ncbi.nlm.nih.gov/pubmed/23755080
http://dx.doi.org/10.4070/kcj.2013.43.5.336
work_keys_str_mv AT choikyu acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT kimjunghee acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT kimhyojin acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT leeseok acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT jangeunyoung acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT kimjunesoo acaseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT choikyu caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT kimjunghee caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT kimhyojin caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT leeseok caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT jangeunyoung caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation
AT kimjunesoo caseofriatadualcoildefibrillatorleadfailureinapatientwithventricularfibrillation