Evaluation of ultrasound Tissue Velocity Imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions

BACKGROUND: Tissue Velocity Imaging (TVI) is an ultrasound based technique used for quantitative analysis of the cardiac function and has earlier been evaluated according to myocardial velocities. Recent years several studies have reported applying TVI in the analysis of skeletal muscles. Skeletal t...

Descripción completa

Detalles Bibliográficos
Autores principales: Lindberg, Frida, Mårtensson, Mattias, Grönlund, Christer, Brodin, Lars-Åke
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3680050/
https://www.ncbi.nlm.nih.gov/pubmed/23758876
http://dx.doi.org/10.1186/1471-2342-13-16
_version_ 1782273065380478976
author Lindberg, Frida
Mårtensson, Mattias
Grönlund, Christer
Brodin, Lars-Åke
author_facet Lindberg, Frida
Mårtensson, Mattias
Grönlund, Christer
Brodin, Lars-Åke
author_sort Lindberg, Frida
collection PubMed
description BACKGROUND: Tissue Velocity Imaging (TVI) is an ultrasound based technique used for quantitative analysis of the cardiac function and has earlier been evaluated according to myocardial velocities. Recent years several studies have reported applying TVI in the analysis of skeletal muscles. Skeletal tissue velocities can be very low. In particular, when performing isometric contractions or contractions of low force level the velocities may be much lower compared to the myocardial tissue velocities. METHODS: In this study TVI was evaluated for estimation of tissue velocities below the typical myocardial velocities. An in-house phantom was used to see how different PRF-settings affected the accuracy of the velocity estimations. RESULTS: With phantom peak velocity at 0.03 cm/s the error ranged from 31% up to 313% with the different PRF-settings in this study. For the peak velocities at 0.17 cm/s and 0.26 cm/s there was no difference in error with tested PFR settings, it is kept approximately around 20%. CONCLUSIONS: The results from the present study showed that the PRF setting did not seem to affect the accuracy of the velocity estimation at tissue velocities above 0.17 cm/s. However at lower velocities (0.03 cm/s) the setting was crucial for the accuracy. The PRF should therefore preferable be reduced when the method is applied in low-level muscle contraction.
format Online
Article
Text
id pubmed-3680050
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-36800502013-06-13 Evaluation of ultrasound Tissue Velocity Imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions Lindberg, Frida Mårtensson, Mattias Grönlund, Christer Brodin, Lars-Åke BMC Med Imaging Research Article BACKGROUND: Tissue Velocity Imaging (TVI) is an ultrasound based technique used for quantitative analysis of the cardiac function and has earlier been evaluated according to myocardial velocities. Recent years several studies have reported applying TVI in the analysis of skeletal muscles. Skeletal tissue velocities can be very low. In particular, when performing isometric contractions or contractions of low force level the velocities may be much lower compared to the myocardial tissue velocities. METHODS: In this study TVI was evaluated for estimation of tissue velocities below the typical myocardial velocities. An in-house phantom was used to see how different PRF-settings affected the accuracy of the velocity estimations. RESULTS: With phantom peak velocity at 0.03 cm/s the error ranged from 31% up to 313% with the different PRF-settings in this study. For the peak velocities at 0.17 cm/s and 0.26 cm/s there was no difference in error with tested PFR settings, it is kept approximately around 20%. CONCLUSIONS: The results from the present study showed that the PRF setting did not seem to affect the accuracy of the velocity estimation at tissue velocities above 0.17 cm/s. However at lower velocities (0.03 cm/s) the setting was crucial for the accuracy. The PRF should therefore preferable be reduced when the method is applied in low-level muscle contraction. BioMed Central 2013-06-07 /pmc/articles/PMC3680050/ /pubmed/23758876 http://dx.doi.org/10.1186/1471-2342-13-16 Text en Copyright © 2013 Lindberg et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Lindberg, Frida
Mårtensson, Mattias
Grönlund, Christer
Brodin, Lars-Åke
Evaluation of ultrasound Tissue Velocity Imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions
title Evaluation of ultrasound Tissue Velocity Imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions
title_full Evaluation of ultrasound Tissue Velocity Imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions
title_fullStr Evaluation of ultrasound Tissue Velocity Imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions
title_full_unstemmed Evaluation of ultrasound Tissue Velocity Imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions
title_short Evaluation of ultrasound Tissue Velocity Imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions
title_sort evaluation of ultrasound tissue velocity imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3680050/
https://www.ncbi.nlm.nih.gov/pubmed/23758876
http://dx.doi.org/10.1186/1471-2342-13-16
work_keys_str_mv AT lindbergfrida evaluationofultrasoundtissuevelocityimagingaphantomstudyofvelocityestimationinskeletalmusclelowlevelcontractions
AT martenssonmattias evaluationofultrasoundtissuevelocityimagingaphantomstudyofvelocityestimationinskeletalmusclelowlevelcontractions
AT gronlundchrister evaluationofultrasoundtissuevelocityimagingaphantomstudyofvelocityestimationinskeletalmusclelowlevelcontractions
AT brodinlarsake evaluationofultrasoundtissuevelocityimagingaphantomstudyofvelocityestimationinskeletalmusclelowlevelcontractions