Cargando…
Poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry
PURPOSE: The aim of this study is to provide data on the controversial issue of whether handheld measurements of intraocular pressure (IOP) are capable of accurately predicting elevated intracranial pressure (ICP) in patients undergoing lumbar puncture (LP). METHODS: All patients over the age of 18...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove Medical Press
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3682855/ https://www.ncbi.nlm.nih.gov/pubmed/23785230 http://dx.doi.org/10.2147/OPTH.S38910 |
_version_ | 1782273409719205888 |
---|---|
author | Golan, Shani Kurtz, Shimon Mezad-Koursh, Daphna Waisbourd, Michael Kesler, Anat Halpern, Pinchas |
author_facet | Golan, Shani Kurtz, Shimon Mezad-Koursh, Daphna Waisbourd, Michael Kesler, Anat Halpern, Pinchas |
author_sort | Golan, Shani |
collection | PubMed |
description | PURPOSE: The aim of this study is to provide data on the controversial issue of whether handheld measurements of intraocular pressure (IOP) are capable of accurately predicting elevated intracranial pressure (ICP) in patients undergoing lumbar puncture (LP). METHODS: All patients over the age of 18 years who underwent an LP in the emergency or neurological departments at the Tel Aviv Medical Center for any reason between October 2007 and July 2010 were eligible to participate in this prospective observational pilot study. IOP was measured with the Tono-Pen XL while patients were in the supine position before undergoing LP. ICP was measured in the lateral recumbent position. ICP and bilateral IOP were measured, and the mean and maximum values of IOP were calculated. The association between ICP and each one of the four IOP measures was evaluated by the Pearson correlation coefficient. RESULTS: Twenty-four patients (mean age 37.8 ± 15.8 years, ten males and 14 females) were enrolled. The reasons for their requiring an LP were headache (19/24 patients), evaluation for hemiparesis (2/24), cognitive deterioration (1/24), and seizures (2/24). Nine had elevated mean opening pressure (>20 cm H(2)O), six had an elevated mean IOP (>20 mmHg), and four of these six also had an elevated opening pressure. There was no significant correlation between the ICP measurements and any of the IOP measurements. CONCLUSION: Handheld ocular tonometry has poor sensitivity and specificity for the prediction of increased ICP and is not an effective tool for screening for ICP in the ED or in the neurology department. |
format | Online Article Text |
id | pubmed-3682855 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Dove Medical Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-36828552013-06-19 Poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry Golan, Shani Kurtz, Shimon Mezad-Koursh, Daphna Waisbourd, Michael Kesler, Anat Halpern, Pinchas Clin Ophthalmol Original Research PURPOSE: The aim of this study is to provide data on the controversial issue of whether handheld measurements of intraocular pressure (IOP) are capable of accurately predicting elevated intracranial pressure (ICP) in patients undergoing lumbar puncture (LP). METHODS: All patients over the age of 18 years who underwent an LP in the emergency or neurological departments at the Tel Aviv Medical Center for any reason between October 2007 and July 2010 were eligible to participate in this prospective observational pilot study. IOP was measured with the Tono-Pen XL while patients were in the supine position before undergoing LP. ICP was measured in the lateral recumbent position. ICP and bilateral IOP were measured, and the mean and maximum values of IOP were calculated. The association between ICP and each one of the four IOP measures was evaluated by the Pearson correlation coefficient. RESULTS: Twenty-four patients (mean age 37.8 ± 15.8 years, ten males and 14 females) were enrolled. The reasons for their requiring an LP were headache (19/24 patients), evaluation for hemiparesis (2/24), cognitive deterioration (1/24), and seizures (2/24). Nine had elevated mean opening pressure (>20 cm H(2)O), six had an elevated mean IOP (>20 mmHg), and four of these six also had an elevated opening pressure. There was no significant correlation between the ICP measurements and any of the IOP measurements. CONCLUSION: Handheld ocular tonometry has poor sensitivity and specificity for the prediction of increased ICP and is not an effective tool for screening for ICP in the ED or in the neurology department. Dove Medical Press 2013 2013-06-10 /pmc/articles/PMC3682855/ /pubmed/23785230 http://dx.doi.org/10.2147/OPTH.S38910 Text en © 2013 Golan et al, publisher and licensee Dove Medical Press Ltd This is an Open Access article which permits unrestricted noncommercial use, provided the original work is properly cited. |
spellingShingle | Original Research Golan, Shani Kurtz, Shimon Mezad-Koursh, Daphna Waisbourd, Michael Kesler, Anat Halpern, Pinchas Poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry |
title | Poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry |
title_full | Poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry |
title_fullStr | Poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry |
title_full_unstemmed | Poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry |
title_short | Poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry |
title_sort | poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3682855/ https://www.ncbi.nlm.nih.gov/pubmed/23785230 http://dx.doi.org/10.2147/OPTH.S38910 |
work_keys_str_mv | AT golanshani poorcorrelationbetweenintracranialpressureandintraocularpressurebyhandheldtonometry AT kurtzshimon poorcorrelationbetweenintracranialpressureandintraocularpressurebyhandheldtonometry AT mezadkourshdaphna poorcorrelationbetweenintracranialpressureandintraocularpressurebyhandheldtonometry AT waisbourdmichael poorcorrelationbetweenintracranialpressureandintraocularpressurebyhandheldtonometry AT kesleranat poorcorrelationbetweenintracranialpressureandintraocularpressurebyhandheldtonometry AT halpernpinchas poorcorrelationbetweenintracranialpressureandintraocularpressurebyhandheldtonometry |