Evaluation of magnetic nanoparticle samples made from biocompatible ferucarbotran by time-correlation magnetic particle imaging reconstruction method
BACKGROUND: Molecular imaging using magnetic nanoparticles (MNPs)—magnetic particle imaging (MPI)—has attracted interest for the early diagnosis of cancer and cardiovascular disease. However, because a steep local magnetic field distribution is required to obtain a defined image, sophisticated hardw...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3684524/ https://www.ncbi.nlm.nih.gov/pubmed/23734917 http://dx.doi.org/10.1186/1471-2342-13-15 |
_version_ | 1782273574170525696 |
---|---|
author | Ishihara, Yasutoshi Honma, Takumi Nohara, Satoshi Ito, Yoshio |
author_facet | Ishihara, Yasutoshi Honma, Takumi Nohara, Satoshi Ito, Yoshio |
author_sort | Ishihara, Yasutoshi |
collection | PubMed |
description | BACKGROUND: Molecular imaging using magnetic nanoparticles (MNPs)—magnetic particle imaging (MPI)—has attracted interest for the early diagnosis of cancer and cardiovascular disease. However, because a steep local magnetic field distribution is required to obtain a defined image, sophisticated hardware is required. Therefore, it is desirable to realize excellent image quality even with low-performance hardware. In this study, the spatial resolution of MPI was evaluated using an image reconstruction method based on the correlation information of the magnetization signal in a time domain and by applying MNP samples made from biocompatible ferucarbotran that have adjusted particle diameters. METHODS: The magnetization characteristics and particle diameters of four types of MNP samples made from ferucarbotran were evaluated. A numerical analysis based on our proposed method that calculates the image intensity from correlation information between the magnetization signal generated from MNPs and the system function was attempted, and the obtained image quality was compared with that using the prototype in terms of image resolution and image artifacts. RESULTS: MNP samples obtained by adjusting ferucarbotran showed superior properties to conventional ferucarbotran samples, and numerical analysis showed that the same image quality could be obtained using a gradient magnetic field generator with 0.6 times the performance. However, because image blurring was included theoretically by the proposed method, an algorithm will be required to improve performance. CONCLUSIONS: MNP samples obtained by adjusting ferucarbotran showed magnetizing properties superior to conventional ferucarbotran samples, and by using such samples, comparable image quality (spatial resolution) could be obtained with a lower gradient magnetic field intensity. |
format | Online Article Text |
id | pubmed-3684524 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-36845242013-06-25 Evaluation of magnetic nanoparticle samples made from biocompatible ferucarbotran by time-correlation magnetic particle imaging reconstruction method Ishihara, Yasutoshi Honma, Takumi Nohara, Satoshi Ito, Yoshio BMC Med Imaging Research Article BACKGROUND: Molecular imaging using magnetic nanoparticles (MNPs)—magnetic particle imaging (MPI)—has attracted interest for the early diagnosis of cancer and cardiovascular disease. However, because a steep local magnetic field distribution is required to obtain a defined image, sophisticated hardware is required. Therefore, it is desirable to realize excellent image quality even with low-performance hardware. In this study, the spatial resolution of MPI was evaluated using an image reconstruction method based on the correlation information of the magnetization signal in a time domain and by applying MNP samples made from biocompatible ferucarbotran that have adjusted particle diameters. METHODS: The magnetization characteristics and particle diameters of four types of MNP samples made from ferucarbotran were evaluated. A numerical analysis based on our proposed method that calculates the image intensity from correlation information between the magnetization signal generated from MNPs and the system function was attempted, and the obtained image quality was compared with that using the prototype in terms of image resolution and image artifacts. RESULTS: MNP samples obtained by adjusting ferucarbotran showed superior properties to conventional ferucarbotran samples, and numerical analysis showed that the same image quality could be obtained using a gradient magnetic field generator with 0.6 times the performance. However, because image blurring was included theoretically by the proposed method, an algorithm will be required to improve performance. CONCLUSIONS: MNP samples obtained by adjusting ferucarbotran showed magnetizing properties superior to conventional ferucarbotran samples, and by using such samples, comparable image quality (spatial resolution) could be obtained with a lower gradient magnetic field intensity. BioMed Central 2013-06-05 /pmc/articles/PMC3684524/ /pubmed/23734917 http://dx.doi.org/10.1186/1471-2342-13-15 Text en Copyright © 2013 Ishihara et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Ishihara, Yasutoshi Honma, Takumi Nohara, Satoshi Ito, Yoshio Evaluation of magnetic nanoparticle samples made from biocompatible ferucarbotran by time-correlation magnetic particle imaging reconstruction method |
title | Evaluation of magnetic nanoparticle samples made from biocompatible ferucarbotran by time-correlation magnetic particle imaging reconstruction method |
title_full | Evaluation of magnetic nanoparticle samples made from biocompatible ferucarbotran by time-correlation magnetic particle imaging reconstruction method |
title_fullStr | Evaluation of magnetic nanoparticle samples made from biocompatible ferucarbotran by time-correlation magnetic particle imaging reconstruction method |
title_full_unstemmed | Evaluation of magnetic nanoparticle samples made from biocompatible ferucarbotran by time-correlation magnetic particle imaging reconstruction method |
title_short | Evaluation of magnetic nanoparticle samples made from biocompatible ferucarbotran by time-correlation magnetic particle imaging reconstruction method |
title_sort | evaluation of magnetic nanoparticle samples made from biocompatible ferucarbotran by time-correlation magnetic particle imaging reconstruction method |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3684524/ https://www.ncbi.nlm.nih.gov/pubmed/23734917 http://dx.doi.org/10.1186/1471-2342-13-15 |
work_keys_str_mv | AT ishiharayasutoshi evaluationofmagneticnanoparticlesamplesmadefrombiocompatibleferucarbotranbytimecorrelationmagneticparticleimagingreconstructionmethod AT honmatakumi evaluationofmagneticnanoparticlesamplesmadefrombiocompatibleferucarbotranbytimecorrelationmagneticparticleimagingreconstructionmethod AT noharasatoshi evaluationofmagneticnanoparticlesamplesmadefrombiocompatibleferucarbotranbytimecorrelationmagneticparticleimagingreconstructionmethod AT itoyoshio evaluationofmagneticnanoparticlesamplesmadefrombiocompatibleferucarbotranbytimecorrelationmagneticparticleimagingreconstructionmethod |