A simple set of validation steps identifies and removes false results in a sandwich enzyme-linked immunosorbent assay caused by anti-animal IgG antibodies in plasma from arthritis patients
Rheumatoid arthritis (RA) and spondyloarthritis (SpA) are chronic diseases characterized by activation of the immune system and production of antibodies. Thus, rheumatoid factor, anti-animal IgG antibodies and heterophilic antibodies in plasma samples from arthritis patients can interfere with immun...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing AG
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3695686/ https://www.ncbi.nlm.nih.gov/pubmed/23875127 http://dx.doi.org/10.1186/2193-1801-2-263 |
_version_ | 1782274999240884224 |
---|---|
author | Kragstrup, Tue W Vorup-Jensen, Thomas Deleuran, Bent Hvid, Malene |
author_facet | Kragstrup, Tue W Vorup-Jensen, Thomas Deleuran, Bent Hvid, Malene |
author_sort | Kragstrup, Tue W |
collection | PubMed |
description | Rheumatoid arthritis (RA) and spondyloarthritis (SpA) are chronic diseases characterized by activation of the immune system and production of antibodies. Thus, rheumatoid factor, anti-animal IgG antibodies and heterophilic antibodies in plasma samples from arthritis patients can interfere with immunoassays such as sandwich enzyme-linked immunosorbent assay (ELISA) systems often used in arthritis research. However, standard methodologies on how to test for false results caused by these antibodies are lacking. The objective of this study was to design a simple set of steps to validate a sandwich ELISA before using it for measuring analytes in plasma from arthritis patients. An interleukin-24 (IL-24) sandwich ELISA system was prepared with a monoclonal mouse capture antibody and a polyclonal goat detection antibody and tested for interference by rheumatoid factor, anti-animal IgG antibodies and heterophilic antibodies. Plasma samples from 23 patients with RA and SpA were used. No differences were found between plasma samples measured in wells coated with anti-IL-24 specific antibody and in wells coated with isotype control antibody (false positive results), and recombinant human IL-24 was not recovered in spiked samples (false negative results). This interference was removed after preincubating the plasma samples from patients with arthritis with goat or bovine IgG, suggesting that anti-animal IgG antibodies found in the plasma of the arthritis patients caused the false results. Additional testing showed that the signal-to-noise ratio could be increased by titration of the capture and detection antibodies and by using the ELAST amplification system. Finally, the calculated concentration of IL-24 was increased in ethylenediaminetetraacetic acid (EDTA) plasma compared to heparin plasma and serum and decreased with repetitive freeze/thaw cycles of the samples illustrating how sample handling could additionally contribute to the variations reported by different laboratories in measurement of the same analyte. This study proposes a simple set of validation steps to evaluate and optimize a sandwich ELISA before using it for measuring analytes in plasma from arthritis patients. Anti-animal IgG antibodies are also present in healthy individuals, suggesting that validation of ELISA systems for measuring non-arthritis samples could also be improved by this simple set of validation steps. |
format | Online Article Text |
id | pubmed-3695686 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Springer International Publishing AG |
record_format | MEDLINE/PubMed |
spelling | pubmed-36956862013-07-18 A simple set of validation steps identifies and removes false results in a sandwich enzyme-linked immunosorbent assay caused by anti-animal IgG antibodies in plasma from arthritis patients Kragstrup, Tue W Vorup-Jensen, Thomas Deleuran, Bent Hvid, Malene Springerplus Research Rheumatoid arthritis (RA) and spondyloarthritis (SpA) are chronic diseases characterized by activation of the immune system and production of antibodies. Thus, rheumatoid factor, anti-animal IgG antibodies and heterophilic antibodies in plasma samples from arthritis patients can interfere with immunoassays such as sandwich enzyme-linked immunosorbent assay (ELISA) systems often used in arthritis research. However, standard methodologies on how to test for false results caused by these antibodies are lacking. The objective of this study was to design a simple set of steps to validate a sandwich ELISA before using it for measuring analytes in plasma from arthritis patients. An interleukin-24 (IL-24) sandwich ELISA system was prepared with a monoclonal mouse capture antibody and a polyclonal goat detection antibody and tested for interference by rheumatoid factor, anti-animal IgG antibodies and heterophilic antibodies. Plasma samples from 23 patients with RA and SpA were used. No differences were found between plasma samples measured in wells coated with anti-IL-24 specific antibody and in wells coated with isotype control antibody (false positive results), and recombinant human IL-24 was not recovered in spiked samples (false negative results). This interference was removed after preincubating the plasma samples from patients with arthritis with goat or bovine IgG, suggesting that anti-animal IgG antibodies found in the plasma of the arthritis patients caused the false results. Additional testing showed that the signal-to-noise ratio could be increased by titration of the capture and detection antibodies and by using the ELAST amplification system. Finally, the calculated concentration of IL-24 was increased in ethylenediaminetetraacetic acid (EDTA) plasma compared to heparin plasma and serum and decreased with repetitive freeze/thaw cycles of the samples illustrating how sample handling could additionally contribute to the variations reported by different laboratories in measurement of the same analyte. This study proposes a simple set of validation steps to evaluate and optimize a sandwich ELISA before using it for measuring analytes in plasma from arthritis patients. Anti-animal IgG antibodies are also present in healthy individuals, suggesting that validation of ELISA systems for measuring non-arthritis samples could also be improved by this simple set of validation steps. Springer International Publishing AG 2013-06-15 /pmc/articles/PMC3695686/ /pubmed/23875127 http://dx.doi.org/10.1186/2193-1801-2-263 Text en © Kragstrup et al.; licensee Springer. 2013 This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Kragstrup, Tue W Vorup-Jensen, Thomas Deleuran, Bent Hvid, Malene A simple set of validation steps identifies and removes false results in a sandwich enzyme-linked immunosorbent assay caused by anti-animal IgG antibodies in plasma from arthritis patients |
title | A simple set of validation steps identifies and removes false results in a sandwich enzyme-linked immunosorbent assay caused by anti-animal IgG antibodies in plasma from arthritis patients |
title_full | A simple set of validation steps identifies and removes false results in a sandwich enzyme-linked immunosorbent assay caused by anti-animal IgG antibodies in plasma from arthritis patients |
title_fullStr | A simple set of validation steps identifies and removes false results in a sandwich enzyme-linked immunosorbent assay caused by anti-animal IgG antibodies in plasma from arthritis patients |
title_full_unstemmed | A simple set of validation steps identifies and removes false results in a sandwich enzyme-linked immunosorbent assay caused by anti-animal IgG antibodies in plasma from arthritis patients |
title_short | A simple set of validation steps identifies and removes false results in a sandwich enzyme-linked immunosorbent assay caused by anti-animal IgG antibodies in plasma from arthritis patients |
title_sort | simple set of validation steps identifies and removes false results in a sandwich enzyme-linked immunosorbent assay caused by anti-animal igg antibodies in plasma from arthritis patients |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3695686/ https://www.ncbi.nlm.nih.gov/pubmed/23875127 http://dx.doi.org/10.1186/2193-1801-2-263 |
work_keys_str_mv | AT kragstruptuew asimplesetofvalidationstepsidentifiesandremovesfalseresultsinasandwichenzymelinkedimmunosorbentassaycausedbyantianimaliggantibodiesinplasmafromarthritispatients AT vorupjensenthomas asimplesetofvalidationstepsidentifiesandremovesfalseresultsinasandwichenzymelinkedimmunosorbentassaycausedbyantianimaliggantibodiesinplasmafromarthritispatients AT deleuranbent asimplesetofvalidationstepsidentifiesandremovesfalseresultsinasandwichenzymelinkedimmunosorbentassaycausedbyantianimaliggantibodiesinplasmafromarthritispatients AT hvidmalene asimplesetofvalidationstepsidentifiesandremovesfalseresultsinasandwichenzymelinkedimmunosorbentassaycausedbyantianimaliggantibodiesinplasmafromarthritispatients AT kragstruptuew simplesetofvalidationstepsidentifiesandremovesfalseresultsinasandwichenzymelinkedimmunosorbentassaycausedbyantianimaliggantibodiesinplasmafromarthritispatients AT vorupjensenthomas simplesetofvalidationstepsidentifiesandremovesfalseresultsinasandwichenzymelinkedimmunosorbentassaycausedbyantianimaliggantibodiesinplasmafromarthritispatients AT deleuranbent simplesetofvalidationstepsidentifiesandremovesfalseresultsinasandwichenzymelinkedimmunosorbentassaycausedbyantianimaliggantibodiesinplasmafromarthritispatients AT hvidmalene simplesetofvalidationstepsidentifiesandremovesfalseresultsinasandwichenzymelinkedimmunosorbentassaycausedbyantianimaliggantibodiesinplasmafromarthritispatients |