Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate
BACKGROUND: There are several factors that can affect the fatigue life of a bone plate, including the mechanical properties of the plate and the complexity of the fracture. The position of the screws can influence construct stiffness, plate strain and cyclic fatigue of the implants. Studies have not...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3704939/ https://www.ncbi.nlm.nih.gov/pubmed/23800317 http://dx.doi.org/10.1186/1746-6148-9-125 |
_version_ | 1782476371379879936 |
---|---|
author | Chao, Peini Conrad, Bryan P Lewis, Daniel D Horodyski, MaryBeth Pozzi, Antonio |
author_facet | Chao, Peini Conrad, Bryan P Lewis, Daniel D Horodyski, MaryBeth Pozzi, Antonio |
author_sort | Chao, Peini |
collection | PubMed |
description | BACKGROUND: There are several factors that can affect the fatigue life of a bone plate, including the mechanical properties of the plate and the complexity of the fracture. The position of the screws can influence construct stiffness, plate strain and cyclic fatigue of the implants. Studies have not investigated these variables in implants utilized for long bone fracture fixation in dogs and cats. The purpose of the present study was to evaluate the effect of plate working length on construct stiffness, gap motion and resistance to cyclic fatigue of dog femora with a simulated fracture gap stabilized using a 12-hole 2.4 mm locking compression plates (LCP). Femora were plated with 12-hole 2.4 mm LCP using 2 screws per fracture segment (long working length group) or with 12-hole 2.4 mm LCP using 5 screws per fracture segment (a short working length group). RESULTS: Construct stiffness did not differ significantly between stabilization techniques. Implant failure did not occur in any of the plated femora during cycling. Mean ± SD yield load at failure in the short plate working length group was significantly higher than in the long plate working length group. CONCLUSION: In a femoral fracture gap model stabilized with a 2.4 mm LCP applied in contact with the bone, plate working length had no effect on stiffness, gap motion and resistance to fatigue. The short plate working length constructs failed at higher loads; however, yield loads for both the short and long plate working length constructs were within physiologic range. |
format | Online Article Text |
id | pubmed-3704939 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-37049392013-07-10 Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate Chao, Peini Conrad, Bryan P Lewis, Daniel D Horodyski, MaryBeth Pozzi, Antonio BMC Vet Res Research Article BACKGROUND: There are several factors that can affect the fatigue life of a bone plate, including the mechanical properties of the plate and the complexity of the fracture. The position of the screws can influence construct stiffness, plate strain and cyclic fatigue of the implants. Studies have not investigated these variables in implants utilized for long bone fracture fixation in dogs and cats. The purpose of the present study was to evaluate the effect of plate working length on construct stiffness, gap motion and resistance to cyclic fatigue of dog femora with a simulated fracture gap stabilized using a 12-hole 2.4 mm locking compression plates (LCP). Femora were plated with 12-hole 2.4 mm LCP using 2 screws per fracture segment (long working length group) or with 12-hole 2.4 mm LCP using 5 screws per fracture segment (a short working length group). RESULTS: Construct stiffness did not differ significantly between stabilization techniques. Implant failure did not occur in any of the plated femora during cycling. Mean ± SD yield load at failure in the short plate working length group was significantly higher than in the long plate working length group. CONCLUSION: In a femoral fracture gap model stabilized with a 2.4 mm LCP applied in contact with the bone, plate working length had no effect on stiffness, gap motion and resistance to fatigue. The short plate working length constructs failed at higher loads; however, yield loads for both the short and long plate working length constructs were within physiologic range. BioMed Central 2013-06-24 /pmc/articles/PMC3704939/ /pubmed/23800317 http://dx.doi.org/10.1186/1746-6148-9-125 Text en Copyright © 2013 Chao et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Chao, Peini Conrad, Bryan P Lewis, Daniel D Horodyski, MaryBeth Pozzi, Antonio Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate |
title | Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate |
title_full | Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate |
title_fullStr | Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate |
title_full_unstemmed | Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate |
title_short | Effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate |
title_sort | effect of plate working length on plate stiffness and cyclic fatigue life in a cadaveric femoral fracture gap model stabilized with a 12-hole 2.4 mm locking compression plate |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3704939/ https://www.ncbi.nlm.nih.gov/pubmed/23800317 http://dx.doi.org/10.1186/1746-6148-9-125 |
work_keys_str_mv | AT chaopeini effectofplateworkinglengthonplatestiffnessandcyclicfatiguelifeinacadavericfemoralfracturegapmodelstabilizedwitha12hole24mmlockingcompressionplate AT conradbryanp effectofplateworkinglengthonplatestiffnessandcyclicfatiguelifeinacadavericfemoralfracturegapmodelstabilizedwitha12hole24mmlockingcompressionplate AT lewisdanield effectofplateworkinglengthonplatestiffnessandcyclicfatiguelifeinacadavericfemoralfracturegapmodelstabilizedwitha12hole24mmlockingcompressionplate AT horodyskimarybeth effectofplateworkinglengthonplatestiffnessandcyclicfatiguelifeinacadavericfemoralfracturegapmodelstabilizedwitha12hole24mmlockingcompressionplate AT pozziantonio effectofplateworkinglengthonplatestiffnessandcyclicfatiguelifeinacadavericfemoralfracturegapmodelstabilizedwitha12hole24mmlockingcompressionplate |