Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution

BACKGROUND: The purpose of this study was to evaluate the impact of Cone Beam CT (CBCT) based setup correction on total dose distributions in fractionated frameless stereotactic radiation therapy of intracranial lesions. METHODS: Ten patients with intracranial lesions treated with 30 Gy in 6 fractio...

Descripción completa

Detalles Bibliográficos
Autores principales: Haertl, Petra Maria, Loeschel, Rainer, Repp, Natalia, Pohl, Fabian, Koelbl, Oliver, Dobler, Barbara
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3707816/
https://www.ncbi.nlm.nih.gov/pubmed/23800172
http://dx.doi.org/10.1186/1748-717X-8-153
_version_ 1782276540238659584
author Haertl, Petra Maria
Loeschel, Rainer
Repp, Natalia
Pohl, Fabian
Koelbl, Oliver
Dobler, Barbara
author_facet Haertl, Petra Maria
Loeschel, Rainer
Repp, Natalia
Pohl, Fabian
Koelbl, Oliver
Dobler, Barbara
author_sort Haertl, Petra Maria
collection PubMed
description BACKGROUND: The purpose of this study was to evaluate the impact of Cone Beam CT (CBCT) based setup correction on total dose distributions in fractionated frameless stereotactic radiation therapy of intracranial lesions. METHODS: Ten patients with intracranial lesions treated with 30 Gy in 6 fractions were included in this study. Treatment planning was performed with Oncentra® for a SynergyS® (Elekta Ltd, Crawley, UK) linear accelerator with XVI® Cone Beam CT, and HexaPOD™ couch top. Patients were immobilized by thermoplastic masks (BrainLab, Reuther). After initial patient setup with respect to lasers, a CBCT study was acquired and registered to the planning CT (PL-CT) study. Patient positioning was corrected according to the correction values (translational, rotational) calculated by the XVI® system. Afterwards a second CBCT study was acquired and registered to the PL-CT to confirm the accuracy of the corrections. An in-house developed software was used for rigid transformation of the PL-CT to the CBCT geometry, and dose calculations for each fraction were performed on the transformed CT. The total dose distribution was achieved by back-transformation and summation of the dose distributions of each fraction. Dose distributions based on PL-CT, CBCT (laser set-up), and final CBCT were compared to assess the influence of setup inaccuracies. RESULTS: The mean displacement vector, calculated over all treatments, was reduced from (4.3 ± 1.3) mm for laser based setup to (0.5 ± 0.2) mm if CBCT corrections were applied. The mean rotational errors around the medial-lateral, superior-inferior, anterior-posterior axis were reduced from (−0.1 ± 1.4)°, (0.1 ± 1.2)° and (−0.2 ± 1.0)°, to (0.04 ± 0.4)°, (0.01 ± 0.4)° and (0.02 ± 0.3)°. As a consequence the mean deviation between planned and delivered dose in the planning target volume (PTV) could be reduced from 12.3% to 0.4% for D(95) and from 5.9% to 0.1% for D(av). Maximum deviation was reduced from 31.8% to 0.8% for D(95), and from 20.4% to 0.1% for D(av). CONCLUSION: Real dose distributions differ substantially from planned dose distributions, if setup is performed according to lasers only. Thermoplasic masks combined with a daily CBCT enabled a sufficient accuracy in dose distribution.
format Online
Article
Text
id pubmed-3707816
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-37078162013-07-11 Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution Haertl, Petra Maria Loeschel, Rainer Repp, Natalia Pohl, Fabian Koelbl, Oliver Dobler, Barbara Radiat Oncol Methodology BACKGROUND: The purpose of this study was to evaluate the impact of Cone Beam CT (CBCT) based setup correction on total dose distributions in fractionated frameless stereotactic radiation therapy of intracranial lesions. METHODS: Ten patients with intracranial lesions treated with 30 Gy in 6 fractions were included in this study. Treatment planning was performed with Oncentra® for a SynergyS® (Elekta Ltd, Crawley, UK) linear accelerator with XVI® Cone Beam CT, and HexaPOD™ couch top. Patients were immobilized by thermoplastic masks (BrainLab, Reuther). After initial patient setup with respect to lasers, a CBCT study was acquired and registered to the planning CT (PL-CT) study. Patient positioning was corrected according to the correction values (translational, rotational) calculated by the XVI® system. Afterwards a second CBCT study was acquired and registered to the PL-CT to confirm the accuracy of the corrections. An in-house developed software was used for rigid transformation of the PL-CT to the CBCT geometry, and dose calculations for each fraction were performed on the transformed CT. The total dose distribution was achieved by back-transformation and summation of the dose distributions of each fraction. Dose distributions based on PL-CT, CBCT (laser set-up), and final CBCT were compared to assess the influence of setup inaccuracies. RESULTS: The mean displacement vector, calculated over all treatments, was reduced from (4.3 ± 1.3) mm for laser based setup to (0.5 ± 0.2) mm if CBCT corrections were applied. The mean rotational errors around the medial-lateral, superior-inferior, anterior-posterior axis were reduced from (−0.1 ± 1.4)°, (0.1 ± 1.2)° and (−0.2 ± 1.0)°, to (0.04 ± 0.4)°, (0.01 ± 0.4)° and (0.02 ± 0.3)°. As a consequence the mean deviation between planned and delivered dose in the planning target volume (PTV) could be reduced from 12.3% to 0.4% for D(95) and from 5.9% to 0.1% for D(av). Maximum deviation was reduced from 31.8% to 0.8% for D(95), and from 20.4% to 0.1% for D(av). CONCLUSION: Real dose distributions differ substantially from planned dose distributions, if setup is performed according to lasers only. Thermoplasic masks combined with a daily CBCT enabled a sufficient accuracy in dose distribution. BioMed Central 2013-06-25 /pmc/articles/PMC3707816/ /pubmed/23800172 http://dx.doi.org/10.1186/1748-717X-8-153 Text en Copyright © 2013 Haertl et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Methodology
Haertl, Petra Maria
Loeschel, Rainer
Repp, Natalia
Pohl, Fabian
Koelbl, Oliver
Dobler, Barbara
Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution
title Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution
title_full Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution
title_fullStr Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution
title_full_unstemmed Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution
title_short Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution
title_sort frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam ct based setup correction on dose distribution
topic Methodology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3707816/
https://www.ncbi.nlm.nih.gov/pubmed/23800172
http://dx.doi.org/10.1186/1748-717X-8-153
work_keys_str_mv AT haertlpetramaria framelessfractionatedstereotacticradiationtherapyofintracraniallesionsimpactofconebeamctbasedsetupcorrectionondosedistribution
AT loeschelrainer framelessfractionatedstereotacticradiationtherapyofintracraniallesionsimpactofconebeamctbasedsetupcorrectionondosedistribution
AT reppnatalia framelessfractionatedstereotacticradiationtherapyofintracraniallesionsimpactofconebeamctbasedsetupcorrectionondosedistribution
AT pohlfabian framelessfractionatedstereotacticradiationtherapyofintracraniallesionsimpactofconebeamctbasedsetupcorrectionondosedistribution
AT koelbloliver framelessfractionatedstereotacticradiationtherapyofintracraniallesionsimpactofconebeamctbasedsetupcorrectionondosedistribution
AT doblerbarbara framelessfractionatedstereotacticradiationtherapyofintracraniallesionsimpactofconebeamctbasedsetupcorrectionondosedistribution