Validation of the Omron HEM-7201 upper arm blood pressure monitor, for self-measurement in a high altitude environment, according to the European Society of Hypertension International Protocol revision 2010

Few studies have been conducted on blood pressure monitors and their use at high altitude. This study is the first to evaluate an automated blood pressure device at high altitude following a standard validation protocol. The Omron HEM-7201 upper arm automatic blood pressure monitor was tested for ac...

Descripción completa

Detalles Bibliográficos
Autores principales: Cho, KaWing, Tian, Maoyi, Lan, Yonghao, Zhao, Xingshan, Yan, Lijing L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3709008/
https://www.ncbi.nlm.nih.gov/pubmed/23466876
http://dx.doi.org/10.1038/jhh.2013.4