Effects of diffusional kurtosis imaging parameters on diffusion quantification

Diffusional kurtosis imaging (DKI) is a new technique based on non-Gaussian water diffusion analysis. However, the original DKI protocol (six b values and 30 motion-probing gradient (MPG) directions) requires more than 10 min of scanning time, which is too long for daily clinical use. We aimed to fi...

Descripción completa

Detalles Bibliográficos
Autores principales: Fukunaga, Issei, Hori, Masaaki, Masutani, Yoshitaka, Hamasaki, Nozomi, Sato, Shuji, Suzuki, Yuriko, Kumagai, Fumitaka, Kosuge, Masatsugu, Hoshito, Haruyoshi, Kamagata, Koji, Shimoji, Keigo, Nakanishi, Atsushi, Aoki, Shigeki, Senoo, Atsushi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Japan 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3709076/
https://www.ncbi.nlm.nih.gov/pubmed/23536232
http://dx.doi.org/10.1007/s12194-013-0206-5
_version_ 1782276700634087424
author Fukunaga, Issei
Hori, Masaaki
Masutani, Yoshitaka
Hamasaki, Nozomi
Sato, Shuji
Suzuki, Yuriko
Kumagai, Fumitaka
Kosuge, Masatsugu
Hoshito, Haruyoshi
Kamagata, Koji
Shimoji, Keigo
Nakanishi, Atsushi
Aoki, Shigeki
Senoo, Atsushi
author_facet Fukunaga, Issei
Hori, Masaaki
Masutani, Yoshitaka
Hamasaki, Nozomi
Sato, Shuji
Suzuki, Yuriko
Kumagai, Fumitaka
Kosuge, Masatsugu
Hoshito, Haruyoshi
Kamagata, Koji
Shimoji, Keigo
Nakanishi, Atsushi
Aoki, Shigeki
Senoo, Atsushi
author_sort Fukunaga, Issei
collection PubMed
description Diffusional kurtosis imaging (DKI) is a new technique based on non-Gaussian water diffusion analysis. However, the original DKI protocol (six b values and 30 motion-probing gradient (MPG) directions) requires more than 10 min of scanning time, which is too long for daily clinical use. We aimed to find suitable b value, MPG direction, and diffusion time settings for faster DKI. Four normal healthy subjects participated in the study. All DKI data sets were acquired on a clinical 3T-MRI scanner (Philips Medical Systems) with use of three protocols of 0–7500 s/mm(2) b values, 6–32 MPG directions, and 23–80 ms diffusion time. There was a remarkable difference in the standard deviation (SD) of the mean DK values in the number of MPG directions. The mean DK values were significantly higher in the posterior limb of the internal capsule (p = 0.003, r = 0.924) and thalamus (p = 0.005, r = 0.903), whereas the mean DK values of the cerebrospinal fluid (CSF) (p = 0.001, r = −0.976) were significantly lower when we used a longer diffusion time. Our results indicate that the SD of the mean DK values was higher in 15 MPG directions than in 20 MPG directions and more. Because the mean DK values of the CSF were significantly lower when we used longer diffusion times, we expect longer diffusion times to be useful for DKI. We propose the following imaging parameters for clinical use: 0, 1000, and 2000 s/mm(2) b values; 20 MPG directions; Δ/δ 45.3/13.3 ms.
format Online
Article
Text
id pubmed-3709076
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Springer Japan
record_format MEDLINE/PubMed
spelling pubmed-37090762013-07-15 Effects of diffusional kurtosis imaging parameters on diffusion quantification Fukunaga, Issei Hori, Masaaki Masutani, Yoshitaka Hamasaki, Nozomi Sato, Shuji Suzuki, Yuriko Kumagai, Fumitaka Kosuge, Masatsugu Hoshito, Haruyoshi Kamagata, Koji Shimoji, Keigo Nakanishi, Atsushi Aoki, Shigeki Senoo, Atsushi Radiol Phys Technol Article Diffusional kurtosis imaging (DKI) is a new technique based on non-Gaussian water diffusion analysis. However, the original DKI protocol (six b values and 30 motion-probing gradient (MPG) directions) requires more than 10 min of scanning time, which is too long for daily clinical use. We aimed to find suitable b value, MPG direction, and diffusion time settings for faster DKI. Four normal healthy subjects participated in the study. All DKI data sets were acquired on a clinical 3T-MRI scanner (Philips Medical Systems) with use of three protocols of 0–7500 s/mm(2) b values, 6–32 MPG directions, and 23–80 ms diffusion time. There was a remarkable difference in the standard deviation (SD) of the mean DK values in the number of MPG directions. The mean DK values were significantly higher in the posterior limb of the internal capsule (p = 0.003, r = 0.924) and thalamus (p = 0.005, r = 0.903), whereas the mean DK values of the cerebrospinal fluid (CSF) (p = 0.001, r = −0.976) were significantly lower when we used a longer diffusion time. Our results indicate that the SD of the mean DK values was higher in 15 MPG directions than in 20 MPG directions and more. Because the mean DK values of the CSF were significantly lower when we used longer diffusion times, we expect longer diffusion times to be useful for DKI. We propose the following imaging parameters for clinical use: 0, 1000, and 2000 s/mm(2) b values; 20 MPG directions; Δ/δ 45.3/13.3 ms. Springer Japan 2013-03-28 2013 /pmc/articles/PMC3709076/ /pubmed/23536232 http://dx.doi.org/10.1007/s12194-013-0206-5 Text en © The Author(s) 2013 https://creativecommons.org/licenses/by/2.0/ Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
spellingShingle Article
Fukunaga, Issei
Hori, Masaaki
Masutani, Yoshitaka
Hamasaki, Nozomi
Sato, Shuji
Suzuki, Yuriko
Kumagai, Fumitaka
Kosuge, Masatsugu
Hoshito, Haruyoshi
Kamagata, Koji
Shimoji, Keigo
Nakanishi, Atsushi
Aoki, Shigeki
Senoo, Atsushi
Effects of diffusional kurtosis imaging parameters on diffusion quantification
title Effects of diffusional kurtosis imaging parameters on diffusion quantification
title_full Effects of diffusional kurtosis imaging parameters on diffusion quantification
title_fullStr Effects of diffusional kurtosis imaging parameters on diffusion quantification
title_full_unstemmed Effects of diffusional kurtosis imaging parameters on diffusion quantification
title_short Effects of diffusional kurtosis imaging parameters on diffusion quantification
title_sort effects of diffusional kurtosis imaging parameters on diffusion quantification
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3709076/
https://www.ncbi.nlm.nih.gov/pubmed/23536232
http://dx.doi.org/10.1007/s12194-013-0206-5
work_keys_str_mv AT fukunagaissei effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT horimasaaki effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT masutaniyoshitaka effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT hamasakinozomi effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT satoshuji effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT suzukiyuriko effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT kumagaifumitaka effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT kosugemasatsugu effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT hoshitoharuyoshi effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT kamagatakoji effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT shimojikeigo effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT nakanishiatsushi effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT aokishigeki effectsofdiffusionalkurtosisimagingparametersondiffusionquantification
AT senooatsushi effectsofdiffusionalkurtosisimagingparametersondiffusionquantification