Bisphosphonate-coated external fixation pins appear similar to hydroxyapatite-coated pins in the tibial metaphysis and to uncoated pins in the shaft: A randomized trial

BACKGROUND AND PURPOSE: External fixation pins tend to loosen with time, especially from cancellous bone. A coating that releases a bisphosphonate has been shown to improve the fixation of dental implants in humans. We now tested a bisphosphonate coating on steel pins for external fixation. The prim...

Descripción completa

Detalles Bibliográficos
Autores principales: Toksvig-Larsen, Sören, Aspenberg, Per
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Informa Healthcare 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3715817/
https://www.ncbi.nlm.nih.gov/pubmed/23621808
http://dx.doi.org/10.3109/17453674.2013.797315
_version_ 1782277507885563904
author Toksvig-Larsen, Sören
Aspenberg, Per
author_facet Toksvig-Larsen, Sören
Aspenberg, Per
author_sort Toksvig-Larsen, Sören
collection PubMed
description BACKGROUND AND PURPOSE: External fixation pins tend to loosen with time, especially from cancellous bone. A coating that releases a bisphosphonate has been shown to improve the fixation of dental implants in humans. We now tested a bisphosphonate coating on steel pins for external fixation. The primary hypothesis was that coated pins would be better fixed in the diaphysis. METHODS: 20 patients with medial knee osteoarthritis underwent proximal tibial correction osteotomy with hemicallotasis. They received a pair of pins (Orthofix) in the tibial shaft, one bisphosphonate-coated and one uncoated. Another pair of pins was inserted in the metaphysis, near the joint. This pair was a bisphosphonate-coated pin and an HA-coated pin. All pins were inserted according to a random list. The pins were removed after the osteotomy had healed (8–15 weeks), and extraction torque served as the predetermined main outcome variable. RESULTS: No pins showed clinical signs of loosening. Removal torque for the shaft pins was 6.6 Nm (SD 2.2) for the bisphosphonate and 6.0 Nm (SD 2.5) for the uncoated (difference = 0.5, 95% CI: –0.03 to 1.3). Removal torque for the metaphyseal pins was 4.4 Nm (SD 1.3) for the bisphosphonate-coated and 4.2 Nm (SD 1.6) for the HA-coated (difference = 0.2, 95% CI: –0.5 to 1.0). INTERPRETATION: We could not show any improved cortical fixation, but the metaphyseal findings are striking. In a previous study on 19 patients with a similar layout, HA-coated and uncoated pins were compared. In the metaphysis, all 19 uncoated pins loosened before removal. It was concluded that uncoated pins could not be used in the metaphyseal region. The present results suggest that a bisphosphonate coating enables metaphyseal fixation similar to that of hydroxyapatite coatings, with no difference from uncoated pins in cortical bone.
format Online
Article
Text
id pubmed-3715817
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Informa Healthcare
record_format MEDLINE/PubMed
spelling pubmed-37158172013-07-19 Bisphosphonate-coated external fixation pins appear similar to hydroxyapatite-coated pins in the tibial metaphysis and to uncoated pins in the shaft: A randomized trial Toksvig-Larsen, Sören Aspenberg, Per Acta Orthop Implant Fixation BACKGROUND AND PURPOSE: External fixation pins tend to loosen with time, especially from cancellous bone. A coating that releases a bisphosphonate has been shown to improve the fixation of dental implants in humans. We now tested a bisphosphonate coating on steel pins for external fixation. The primary hypothesis was that coated pins would be better fixed in the diaphysis. METHODS: 20 patients with medial knee osteoarthritis underwent proximal tibial correction osteotomy with hemicallotasis. They received a pair of pins (Orthofix) in the tibial shaft, one bisphosphonate-coated and one uncoated. Another pair of pins was inserted in the metaphysis, near the joint. This pair was a bisphosphonate-coated pin and an HA-coated pin. All pins were inserted according to a random list. The pins were removed after the osteotomy had healed (8–15 weeks), and extraction torque served as the predetermined main outcome variable. RESULTS: No pins showed clinical signs of loosening. Removal torque for the shaft pins was 6.6 Nm (SD 2.2) for the bisphosphonate and 6.0 Nm (SD 2.5) for the uncoated (difference = 0.5, 95% CI: –0.03 to 1.3). Removal torque for the metaphyseal pins was 4.4 Nm (SD 1.3) for the bisphosphonate-coated and 4.2 Nm (SD 1.6) for the HA-coated (difference = 0.2, 95% CI: –0.5 to 1.0). INTERPRETATION: We could not show any improved cortical fixation, but the metaphyseal findings are striking. In a previous study on 19 patients with a similar layout, HA-coated and uncoated pins were compared. In the metaphysis, all 19 uncoated pins loosened before removal. It was concluded that uncoated pins could not be used in the metaphyseal region. The present results suggest that a bisphosphonate coating enables metaphyseal fixation similar to that of hydroxyapatite coatings, with no difference from uncoated pins in cortical bone. Informa Healthcare 2013-06 2013-05-31 /pmc/articles/PMC3715817/ /pubmed/23621808 http://dx.doi.org/10.3109/17453674.2013.797315 Text en Copyright: © Nordic Orthopaedic Federation http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the source is credited.
spellingShingle Implant Fixation
Toksvig-Larsen, Sören
Aspenberg, Per
Bisphosphonate-coated external fixation pins appear similar to hydroxyapatite-coated pins in the tibial metaphysis and to uncoated pins in the shaft: A randomized trial
title Bisphosphonate-coated external fixation pins appear similar to hydroxyapatite-coated pins in the tibial metaphysis and to uncoated pins in the shaft: A randomized trial
title_full Bisphosphonate-coated external fixation pins appear similar to hydroxyapatite-coated pins in the tibial metaphysis and to uncoated pins in the shaft: A randomized trial
title_fullStr Bisphosphonate-coated external fixation pins appear similar to hydroxyapatite-coated pins in the tibial metaphysis and to uncoated pins in the shaft: A randomized trial
title_full_unstemmed Bisphosphonate-coated external fixation pins appear similar to hydroxyapatite-coated pins in the tibial metaphysis and to uncoated pins in the shaft: A randomized trial
title_short Bisphosphonate-coated external fixation pins appear similar to hydroxyapatite-coated pins in the tibial metaphysis and to uncoated pins in the shaft: A randomized trial
title_sort bisphosphonate-coated external fixation pins appear similar to hydroxyapatite-coated pins in the tibial metaphysis and to uncoated pins in the shaft: a randomized trial
topic Implant Fixation
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3715817/
https://www.ncbi.nlm.nih.gov/pubmed/23621808
http://dx.doi.org/10.3109/17453674.2013.797315
work_keys_str_mv AT toksviglarsensoren bisphosphonatecoatedexternalfixationpinsappearsimilartohydroxyapatitecoatedpinsinthetibialmetaphysisandtouncoatedpinsintheshaftarandomizedtrial
AT aspenbergper bisphosphonatecoatedexternalfixationpinsappearsimilartohydroxyapatitecoatedpinsinthetibialmetaphysisandtouncoatedpinsintheshaftarandomizedtrial