Highly Sensitive Detection of Malaria Parasitemia in a Malaria-Endemic Setting: Performance of a New Loop-Mediated Isothermal Amplification Kit in a Remote Clinic in Uganda
Background. Current malaria diagnostic tests, including microscopy and antigen-detecting rapid tests, cannot reliably detect low-density infections. Molecular methods such as polymerase chain reaction (PCR) are highly sensitive but remain too complex for field deployment. A new commercial molecular...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3719898/ https://www.ncbi.nlm.nih.gov/pubmed/23633405 http://dx.doi.org/10.1093/infdis/jit184 |
Sumario: | Background. Current malaria diagnostic tests, including microscopy and antigen-detecting rapid tests, cannot reliably detect low-density infections. Molecular methods such as polymerase chain reaction (PCR) are highly sensitive but remain too complex for field deployment. A new commercial molecular assay based on loop-mediated isothermal amplification (LAMP) was assessed for field use. Methods. Malaria LAMP (Eiken Chemical, Japan) was evaluated for samples from 272 outpatients at a rural Ugandan clinic and compared with expert microscopy, nested PCR, and quantitative PCR (qPCR). Two technicians performed the assay after 3 days of training, using 2 alternative blood sample–preparation methods and visual interpretation of results by fluorescence assay. Results. Compared with 3-well nested PCR, the sensitivity of both LAMP and single-well nested PCR was 90%; the microscopy sensitivity was 51%. For samples with a Plasmodium falciparum qPCR titer of ≥2 parasites/µL, LAMP sensitivity was 97.8% (95% confidence interval, 93.7%–99.5%). Most false-negative LAMP results involved samples with parasitemia levels detectable by 3-well nested PCR but very low or undetectable by qPCR. Conclusions. Malaria LAMP in a remote Ugandan clinic achieved sensitivity similar to that of single-well nested PCR in a United Kingdom reference laboratory. LAMP dramatically lowers the detection threshold achievable in malaria-endemic settings, providing a new tool for diagnosis, surveillance, and screening in elimination strategies. |
---|