Cement-Implant Interface Contamination: Possible Reason of Inferior Clinical Outcomes for Rough Surface Cemented Stems
BACKGROUND: Shape-closed cemented implants rely on a stronger bond and have displayed inferior clinical outcomes when compared to force-closed designs. Implant contamination such as saline, bone marrow and blood prior to cement application has the potential to affect the cement-implant bond. The con...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Bentham Open
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3722542/ https://www.ncbi.nlm.nih.gov/pubmed/23898352 http://dx.doi.org/10.2174/1874325001307010250 |
_version_ | 1782278206090379264 |
---|---|
author | Wang, Tian Pelletier, Matthew H Bertollo, Nicky Crosky, Alan Walsh, William R |
author_facet | Wang, Tian Pelletier, Matthew H Bertollo, Nicky Crosky, Alan Walsh, William R |
author_sort | Wang, Tian |
collection | PubMed |
description | BACKGROUND: Shape-closed cemented implants rely on a stronger bond and have displayed inferior clinical outcomes when compared to force-closed designs. Implant contamination such as saline, bone marrow and blood prior to cement application has the potential to affect the cement-implant bond. The consequences of implant contamination were investigated in this study. METHODS: Fifty Titanium alloy (Ti-6Al-4V) dowels were separated into ten groups based on surface roughness and contaminant, and then cemented in polyvinyl chloride tubes. Push-out testing was performed at 1mm per minute. The roughness of the dowel surface was measured before and after the testing. The dowel surface and cement mantel were analyzed using a Scanning Electron Microscopy (SEM) to determine the distribution and characteristics of any debris and contaminants on the surface. RESULTS: Contaminants largely decreased stem-cement interfacial shear strength, especially for rough surfaces. Saline produced the greatest decrease, followed by blood. The effect of bone marrow was less pronounced and similar to that of oil. Increasing surface roughness increased the interfacial bonding strength, even with contaminants. There was a non-significant increase in mean bonding strength for smooth surfaces with bone marrow and oil contamination. SEM showed that contaminants influence the interfacial bond by different mechanisms. More debris was found on rough samples following testing. CONCLUSIONS: The results of this study underscore the importance of keeping an implant free from contamination, and suggest if contamination does occur, a saline rinse may further decrease the stability of an implant. The deleterious effects of contamination on rough surface cement bonding were considerable, and indicate that contamination at the time of surgery may, in part, contribute to inferior clinical outcomes for rough surfaced cemented stems. |
format | Online Article Text |
id | pubmed-3722542 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Bentham Open |
record_format | MEDLINE/PubMed |
spelling | pubmed-37225422013-07-29 Cement-Implant Interface Contamination: Possible Reason of Inferior Clinical Outcomes for Rough Surface Cemented Stems Wang, Tian Pelletier, Matthew H Bertollo, Nicky Crosky, Alan Walsh, William R Open Orthop J Article BACKGROUND: Shape-closed cemented implants rely on a stronger bond and have displayed inferior clinical outcomes when compared to force-closed designs. Implant contamination such as saline, bone marrow and blood prior to cement application has the potential to affect the cement-implant bond. The consequences of implant contamination were investigated in this study. METHODS: Fifty Titanium alloy (Ti-6Al-4V) dowels were separated into ten groups based on surface roughness and contaminant, and then cemented in polyvinyl chloride tubes. Push-out testing was performed at 1mm per minute. The roughness of the dowel surface was measured before and after the testing. The dowel surface and cement mantel were analyzed using a Scanning Electron Microscopy (SEM) to determine the distribution and characteristics of any debris and contaminants on the surface. RESULTS: Contaminants largely decreased stem-cement interfacial shear strength, especially for rough surfaces. Saline produced the greatest decrease, followed by blood. The effect of bone marrow was less pronounced and similar to that of oil. Increasing surface roughness increased the interfacial bonding strength, even with contaminants. There was a non-significant increase in mean bonding strength for smooth surfaces with bone marrow and oil contamination. SEM showed that contaminants influence the interfacial bond by different mechanisms. More debris was found on rough samples following testing. CONCLUSIONS: The results of this study underscore the importance of keeping an implant free from contamination, and suggest if contamination does occur, a saline rinse may further decrease the stability of an implant. The deleterious effects of contamination on rough surface cement bonding were considerable, and indicate that contamination at the time of surgery may, in part, contribute to inferior clinical outcomes for rough surfaced cemented stems. Bentham Open 2013-06-28 /pmc/articles/PMC3722542/ /pubmed/23898352 http://dx.doi.org/10.2174/1874325001307010250 Text en © Wang et al.; Licensee Bentham Open. http://creativecommons.org/licenses/by-nc/3.0/ This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited. |
spellingShingle | Article Wang, Tian Pelletier, Matthew H Bertollo, Nicky Crosky, Alan Walsh, William R Cement-Implant Interface Contamination: Possible Reason of Inferior Clinical Outcomes for Rough Surface Cemented Stems |
title | Cement-Implant Interface Contamination: Possible Reason of Inferior Clinical Outcomes for Rough Surface Cemented Stems |
title_full | Cement-Implant Interface Contamination: Possible Reason of Inferior Clinical Outcomes for Rough Surface Cemented Stems |
title_fullStr | Cement-Implant Interface Contamination: Possible Reason of Inferior Clinical Outcomes for Rough Surface Cemented Stems |
title_full_unstemmed | Cement-Implant Interface Contamination: Possible Reason of Inferior Clinical Outcomes for Rough Surface Cemented Stems |
title_short | Cement-Implant Interface Contamination: Possible Reason of Inferior Clinical Outcomes for Rough Surface Cemented Stems |
title_sort | cement-implant interface contamination: possible reason of inferior clinical outcomes for rough surface cemented stems |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3722542/ https://www.ncbi.nlm.nih.gov/pubmed/23898352 http://dx.doi.org/10.2174/1874325001307010250 |
work_keys_str_mv | AT wangtian cementimplantinterfacecontaminationpossiblereasonofinferiorclinicaloutcomesforroughsurfacecementedstems AT pelletiermatthewh cementimplantinterfacecontaminationpossiblereasonofinferiorclinicaloutcomesforroughsurfacecementedstems AT bertollonicky cementimplantinterfacecontaminationpossiblereasonofinferiorclinicaloutcomesforroughsurfacecementedstems AT croskyalan cementimplantinterfacecontaminationpossiblereasonofinferiorclinicaloutcomesforroughsurfacecementedstems AT walshwilliamr cementimplantinterfacecontaminationpossiblereasonofinferiorclinicaloutcomesforroughsurfacecementedstems |