Cargando…
Feasibility of Newly Developed Endoscopic Ultrasound with Zone Sonography Technology for Diagnosis of Pancreatic Diseases
BACKGROUND/AIMS: To confirm the feasibility of using newly developed endoscopic ultrasound (EUS) with Zone sonography™ technology (ZST; Fujifilm Corp.). METHODS: Seventy-five patients with pancreatic disorders were enrolled: 45 with intraductal papillary mucinous neoplasm; 15 with ductal carcinoma;...
Autores principales: | , , , , , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society of Gastroenterology; the Korean Society of Gastrointestinal Endoscopy; the Korean Association for the Study of the Liver; the Korean Society of Neurogastroenterology and Motility; Korean Association for the Study of Intestinal Diseases; Korean College of Helicobacter and Upper Gastrointestinal Research; Korean Pancreatobiliary Association; Korean Society of Gastrointestinal Cancer
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3724040/ https://www.ncbi.nlm.nih.gov/pubmed/23898392 http://dx.doi.org/10.5009/gnl.2013.7.4.486 |
Sumario: | BACKGROUND/AIMS: To confirm the feasibility of using newly developed endoscopic ultrasound (EUS) with Zone sonography™ technology (ZST; Fujifilm Corp.). METHODS: Seventy-five patients with pancreatic disorders were enrolled: 45 with intraductal papillary mucinous neoplasm; 15 with ductal carcinoma; five with neuroendocrine tumors; three with serous cystic neoplasms; and seven with simple cysts. The endoscopes used were EG-530UR2 and EG-530UT2 (Fujifilm Corp.). Two items were evaluated: visualization depth among four frequencies and image quality after automatic adjustment of sound speed (AASS), assessed using a 5-scale Likert scale by two endosonographers blinded to disease status. Because sound speed could be manually controlled, besides AASS, image quality at sound speeds of 1,440 and 1,600 m/sec were also assessed. RESULTS: In all cases, sufficient images were obtained in the range of 3 cm from the EUS probe. Judgments of image quality before AASS were 3.49±0.50, 3.65±0.48, respectively. After AASS, A and B scored 4.36±0.48 and 4.40±0.49 (p<0.0001). There were significant differences in the data before and after AASS and plus 60 m/sec, but no significant difference between the datasets were seen after AASS and at sound speeds manually set for minus 100 m/sec. CONCLUSIONS: EUS with ZST was shown to be feasible in this preliminary experiment. Further evaluation of this novel technology is necessary and awaited. |
---|